NEMA NU-4 performance evaluation of PETbox4, a high sensitivity dedicated PET preclinical tomograph.

Z Gu1, R Taschereau, N T Vu, H Wang, D L Prout, R W Silverman, B Bai, D B Stout, M E Phelps, A F Chatziioannou.   

Abstract

PETbox4 is a new, fully tomographic bench top PET scanner dedicated to high sensitivity and high resolution imaging of mice. This manuscript characterizes the performance of the prototype system using the National Electrical Manufacturers Association NU 4-2008 standards, including studies of sensitivity, spatial resolution, energy resolution, scatter fraction, count-rate performance and image quality. The PETbox4 performance is also compared with the performance of PETbox, a previous generation limited angle tomography system. PETbox4 consists of four opposing flat-panel type detectors arranged in a box-like geometry. Each panel is made by a 24 × 50 pixelated array of 1.82 × 1.82 × 7 mm bismuth germanate scintillation crystals with a crystal pitch of 1.90 mm. Each of these scintillation arrays is coupled to two Hamamatsu H8500 photomultiplier tubes via a glass light guide. Volumetric images for a 45 × 45 × 95 mm field of view (FOV) are reconstructed with a maximum likelihood expectation maximization algorithm incorporating a system model based on a parameterized detector response. With an energy window of 150-650 keV, the peak absolute sensitivity is approximately 18% at the center of FOV. The measured crystal energy resolution ranges from 13.5% to 48.3% full width at half maximum (FWHM), with a mean of 18.0%. The intrinsic detector spatial resolution is 1.5 mm FWHM in both transverse and axial directions. The reconstructed image spatial resolution for different locations in the FOV ranges from 1.32 to 1.93 mm, with an average of 1.46 mm. The peak noise equivalent count rate for the mouse-sized phantom is 35 kcps for a total activity of 1.5 MBq (40 µCi) and the scatter fraction is 28%. The standard deviation in the uniform region of the image quality phantom is 5.7%. The recovery coefficients range from 0.10 to 0.93. In comparison to the first generation two panel PETbox system, PETbox4 achieves substantial improvements on sensitivity and spatial resolution. The overall performance demonstrates that the PETbox4 scanner is suitable for producing high quality images for molecular imaging based biomedical research.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23666034      PMCID: PMC4009355          DOI: 10.1088/0031-9155/58/11/3791

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  48 in total

Review 1.  Molecular imaging.

Authors:  R Weissleder; U Mahmood
Journal:  Radiology       Date:  2001-05       Impact factor: 11.105

Review 2.  Molecular imaging of small animals with dedicated PET tomographs.

Authors:  Arion F Chatziioannou
Journal:  Eur J Nucl Med Mol Imaging       Date:  2001-11-22       Impact factor: 9.236

Review 3.  The biological application of small animal PET imaging.

Authors:  R Myers
Journal:  Nucl Med Biol       Date:  2001-07       Impact factor: 2.408

4.  MicroPET II: design, development and initial performance of an improved microPET scanner for small-animal imaging.

Authors:  Yuan-Chuan Tai; Arion F Chatziioannou; Yongfeng Yang; Robert W Silverman; Ken Meadors; Stefan Siegel; Danny F Newport; Jennifer R Stickel; Simon R Cherry
Journal:  Phys Med Biol       Date:  2003-06-07       Impact factor: 3.609

5.  NEMA NU-04-based performance characteristics of the LabPET-8™ small animal PET scanner.

Authors:  Rameshwar Prasad; Osman Ratib; Habib Zaidi
Journal:  Phys Med Biol       Date:  2011-09-23       Impact factor: 3.609

6.  National Electrical Manufacturers Association NU-4 performance evaluation of the PET component of the NanoPET/CT preclinical PET/CT scanner.

Authors:  Istvan Szanda; Jane Mackewn; Gergely Patay; Peter Major; Kavitha Sunassee; Gregory E Mullen; Gabor Nemeth; York Haemisch; Philip J Blower; Paul K Marsden
Journal:  J Nucl Med       Date:  2011-10-03       Impact factor: 10.057

7.  Fast gradient-based methods for Bayesian reconstruction of transmission and emission PET images.

Authors:  E U Mumcuoglu; R Leahy; S R Cherry; Z Zhou
Journal:  IEEE Trans Med Imaging       Date:  1994       Impact factor: 10.048

8.  Performance evaluation of PETbox: a low cost bench top preclinical PET scanner.

Authors:  Hui Zhang; Qinan Bao; Nam T Vu; Robert W Silverman; Richard Taschereau; Brittany N Berry-Pusey; Ali Douraghy; Fernando R Rannou; David B Stout; Arion F Chatziioannou
Journal:  Mol Imaging Biol       Date:  2011-10       Impact factor: 3.488

9.  The design and physical characteristics of a small animal positron emission tomograph.

Authors:  P M Bloomfield; S Rajeswaran; T J Spinks; S P Hume; R Myers; S Ashworth; K M Clifford; W F Jones; L G Byars; J Young
Journal:  Phys Med Biol       Date:  1995-06       Impact factor: 3.609

10.  Monte Carlo simulations of absorbed dose in a mouse phantom from 18-fluorine compounds.

Authors:  Richard Taschereau; Arion F Chatziioannou
Journal:  Med Phys       Date:  2007-03       Impact factor: 4.071

View more
  14 in total

1.  Evaluation of Matrix9 silicon photomultiplier array for small-animal PET.

Authors:  Junwei Du; Jeffrey P Schmall; Yongfeng Yang; Kun Di; Emilie Roncali; Gregory S Mitchell; Steve Buckley; Carl Jackson; Simon R Cherry
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

2.  Preclinical positron emission tomography scanner based on a monolithic annulus of scintillator: initial design study.

Authors:  Alexander V Stolin; Peter F Martone; Gangadhar Jaliparthi; Raymond R Raylman
Journal:  J Med Imaging (Bellingham)       Date:  2017-01-05

3.  A DOI Detector With Crystal Scatter Identification Capability for High Sensitivity and High Spatial Resolution PET Imaging.

Authors:  Z Gu; D L Prout; R W Silverman; H Herman; A Dooraghi; A F Chatziioannou
Journal:  IEEE Trans Nucl Sci       Date:  2015-06       Impact factor: 1.679

4.  A deformable atlas of the laboratory mouse.

Authors:  Hongkai Wang; David B Stout; Arion F Chatziioannou
Journal:  Mol Imaging Biol       Date:  2015-02       Impact factor: 3.488

5.  Optimization of the Energy Window for PETbox4, a Preclinical PET Tomograph With a Small Inner Diameter.

Authors:  Z Gu; Q Bao; R Taschereau; H Wang; B Bai; A F Chatziioannou
Journal:  IEEE Trans Nucl Sci       Date:  2014-06-01       Impact factor: 1.679

6.  Performance evaluation of dual-ended readout PET detectors based on BGO arrays with different reflector arrangements.

Authors:  Junwei Du; Qian Wang; Chih-Chieh Liu; Jinyi Qi; Simon R Cherry
Journal:  Phys Med Biol       Date:  2021-10-19       Impact factor: 4.174

7.  Performance Evaluation of G8, a High-Sensitivity Benchtop Preclinical PET/CT Tomograph.

Authors:  Zheng Gu; Richard Taschereau; Nam T Vu; David L Prout; Robert W Silverman; Jason T Lee; Arion F Chatziioannou
Journal:  J Nucl Med       Date:  2018-06-14       Impact factor: 10.057

8.  TandemPET- A High Resolution, Small Animal, Virtual Pinhole-Based PET Scanner: Initial Design Study.

Authors:  Raymond R Raylman; Alexander V Stolin; Peter F Martone; Mark F Smith
Journal:  IEEE Trans Nucl Sci       Date:  2015-10-29       Impact factor: 1.679

9.  H2RSPET: a 0.5 mm resolution high-sensitivity small-animal PET scanner, a simulation study.

Authors:  Youfang Lai; Qian Wang; Shiwei Zhou; Zhaoheng Xie; Jinyi Qi; Simon R Cherry; Mingwu Jin; Yujie Chi; Junwei Du
Journal:  Phys Med Biol       Date:  2021-03-09       Impact factor: 3.609

10.  A digital phoswich detector using time-over-threshold for depth of interaction in PET.

Authors:  David L Prout; Zheng Gu; Max Shustef; Arion F Chatziioannou
Journal:  Phys Med Biol       Date:  2020-12-15       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.