The design and physical characteristics of a small animal positron emission tomograph.

P M Bloomfield1, S Rajeswaran, T J Spinks, S P Hume, R Myers, S Ashworth, K M Clifford, W F Jones, L G Byars, J Young.   

Abstract

A small diameter positron emission tomography, designed specifically for small animal studies, was constructed from existing, commercially available, bismuth germanate (BGO) detectors and electronics. The scanner consists of 16 BGO detector blocks arranged to give a tomograph with a diameter of 115 mm and an axial field of view (FOV) of 50 mm. Each block is cut to produce eight (axial) by seven (radial) individual detector elements. The absence of interplane septa enables the acquisition of 3D data sets consisting of 64 sinograms. A 2D data set of 15 sinograms, consisting of eight direct and seven adjacent cross planes, can be extracted from the 3D data set. Images are reconstructed from the 2D sinograms using a conventional filtered backprojection algorithm. Two methods of normalization were investigated, based on either a rotating 68Ge rod source, or a uniform 68Ge plane source, with a uniform cylindrical 18F phantom. Attenuation of the emitted photons was estimated using a rotating 68Ge rod source. The transaxial resolution of the tomograph was measured as 2.3 mm full width at half maximum (FWHM) and 5.6 mm full width at tenth maximum (FWTM) at the centre of the FOV, degrading to 6.6 mm (radial) and 4.4 mm (tangential) FWHM and 10.4 mm (radial) and 14.4 mm (tangential) FWTM at 40.0 mm from the centre of the FOV. The axial slice width was 4.3 mm FWHM, 10.3 mm FWTM at the centre of the transaxial field of view and 4.4 mm FWHM, 10.6 mm FWTM at 20.0 mm from the centre of the FOV. A scatter fraction of 31.0% was measured at 250-850 keV, for an 18F line source centred in a 60 mm diameter, water-filled phantom, reducing to 20.4% and 13.8% as the lower energy discrimination was increased to 380 keV and 450 keV, respectively. The count rate performance was measured using a noise equivalent count rate method, and the linearity of the dead time correction was confirmed over the count rates encountered during routine scanning. In 2D mode, the absolute sensitivity of the tomograph was measured as 9948 counts s-1 MBq-1 at 250-850 keV, 8284 counts s-1 MBq-1 at 380-850 keV and 6280 counts s-1 MBq-1 at 450-850 keV.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7659733     DOI: 10.1088/0031-9155/40/6/010

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  20 in total

Review 1.  Ultra-high-resolution imaging of small animals: implications for preclinical and research studies.

Authors:  D A Weber; M Ivanovic
Journal:  J Nucl Cardiol       Date:  1999 May-Jun       Impact factor: 5.952

2.  Performance evaluation of microPET: a high-resolution lutetium oxyorthosilicate PET scanner for animal imaging.

Authors:  A F Chatziioannou; S R Cherry; Y Shao; R W Silverman; K Meadors; T H Farquhar; M Pedarsani; M E Phelps
Journal:  J Nucl Med       Date:  1999-07       Impact factor: 10.057

Review 3.  Imaging transgene expression with radionuclide imaging technologies.

Authors:  S S Gambhir; H R Herschman; S R Cherry; J R Barrio; N Satyamurthy; T Toyokuni; M E Phelps; S M Larson; J Balatoni; R Finn; M Sadelain; J Tjuvajev; R Blasberg
Journal:  Neoplasia       Date:  2000 Jan-Apr       Impact factor: 5.715

Review 4.  Small animal PET: aspects of performance assessment.

Authors:  Simone Weber; Andreas Bauer
Journal:  Eur J Nucl Med Mol Imaging       Date:  2004-10-02       Impact factor: 9.236

Review 5.  Instrumentation for molecular imaging in preclinical research: Micro-PET and Micro-SPECT.

Authors:  Arion F Chatziioannou
Journal:  Proc Am Thorac Soc       Date:  2005

6.  Experimental characterization and system simulations of depth of interaction PET detectors using 0.5 mm and 0.7 mm LSO arrays.

Authors:  Sara St James; Yongfeng Yang; Yibao Wu; Richard Farrell; Purushottam Dokhale; Kanai S Shah; Simon R Cherry
Journal:  Phys Med Biol       Date:  2009-06-30       Impact factor: 3.609

Review 7.  The imaging science of positron emission tomography.

Authors:  T Jones
Journal:  Eur J Nucl Med       Date:  1996-07

8.  Lack of permanent nigrostriatal dopamine deficit following 6-hydroxydopamine injection into the rat striatum. Short communication.

Authors:  J Opacka-Juffry; S Ashworth; A M Sullivan; R B Banati; S B Blunt
Journal:  J Neural Transm (Vienna)       Date:  1996       Impact factor: 3.575

9.  Preclinical positron emission tomography scanner based on a monolithic annulus of scintillator: initial design study.

Authors:  Alexander V Stolin; Peter F Martone; Gangadhar Jaliparthi; Raymond R Raylman
Journal:  J Med Imaging (Bellingham)       Date:  2017-01-05

10.  Performance evaluation of PETbox: a low cost bench top preclinical PET scanner.

Authors:  Hui Zhang; Qinan Bao; Nam T Vu; Robert W Silverman; Richard Taschereau; Brittany N Berry-Pusey; Ali Douraghy; Fernando R Rannou; David B Stout; Arion F Chatziioannou
Journal:  Mol Imaging Biol       Date:  2011-10       Impact factor: 3.488

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.