National Electrical Manufacturers Association NU-4 performance evaluation of the PET component of the NanoPET/CT preclinical PET/CT scanner.

Istvan Szanda1, Jane Mackewn, Gergely Patay, Peter Major, Kavitha Sunassee, Gregory E Mullen, Gabor Nemeth, York Haemisch, Philip J Blower, Paul K Marsden.   

Abstract

UNLABELLED: The NanoPET/CT represents the latest generation of commercial preclinical PET/CT systems. This article presents a performance evaluation of the PET component of the system according to the National Electrical Manufacturers Association (NEMA) NU-4 2008 standard.
METHODS: The NanoPET/CT consists of 12 lutetium yttrium orthosilicate:cerium modular detectors forming 1 ring, with 9.5-cm axial coverage and a 16-cm animal port. Each detector crystal is 1.12 × 1.12 × 13 mm, and 1 module contains 81 × 39 of these crystals. An optical light guide transmits the scintillation light to the flat-panel multianode position-sensitive photomultiplier tubes. Analog-to-digital converter cards and a field-programmable gate array-based data-collecting card provide the readout. Spatial resolution, sensitivity, counting rate capabilities, and image quality were evaluated in accordance with the NEMA NU-4 standard. Energy and temporal resolution measurements and a mouse imaging study were performed in addition to the standard.
RESULTS: Energy resolution was 19% at 511 keV. The spatial resolution, measured as full width at half maximum on single-slice rebinning/filtered backprojection-reconstructed images, approached 1 mm on the axis and remained below 2.5 mm in the central 5-cm transaxial region both in the axial center and at one-quarter field of view. The maximum absolute sensitivity for a point source at the center of the field of view was 7.7%. The maximum noise equivalent counting rates were 430 kcps at 36 MBq and 130 kcps at 27 MBq for the mouse- and rat-sized phantoms, respectively. The uniformity and recovery coefficients were measured with the image-quality phantom, giving good-quality images. In a mouse study with an (18)F-labeled thyroid-specific tracer, the 2 lobes of the thyroid were clearly distinguishable, despite the small size of this organ. The flexible readout system allowed experiments to be performed in an efficient manner, and the system remained stable throughout.
CONCLUSION: The large number of detector crystals, arranged with a fine pitch, results in excellent spatial resolution, which is the best reported for currently available commercial systems. The absolute sensitivity is high over the field of view. Combined with the excellent image quality, these features make the NanoPET/CT a powerful tool for preclinical research.

Entities:  

Mesh:

Year:  2011        PMID: 21969357     DOI: 10.2967/jnumed.111.088260

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  30 in total

Review 1.  Preclinical imaging: an essential ally in modern biosciences.

Authors:  Lídia Cunha; Ildiko Horvath; Sara Ferreira; Joana Lemos; Pedro Costa; Domingos Vieira; Dániel S Veres; Krisztián Szigeti; Teresa Summavielle; Domokos Máthé; Luís F Metello
Journal:  Mol Diagn Ther       Date:  2014-04       Impact factor: 4.074

2.  Preclinical positron emission tomography scanner based on a monolithic annulus of scintillator: initial design study.

Authors:  Alexander V Stolin; Peter F Martone; Gangadhar Jaliparthi; Raymond R Raylman
Journal:  J Med Imaging (Bellingham)       Date:  2017-01-05

3.  Molecular Imaging of PDE10A Knockout Mice with a Novel PET Radiotracer: [(11)C]T-773.

Authors:  Miklós Tóth; Jenny Häggkvist; Vladimir Stepanov; Akihiro Takano; Ryuji Nakao; Nahid Amini; Shotaro Miura; Haruhide Kimura; Takahiko Taniguchi; Balázs Gulyás; Christer Halldin
Journal:  Mol Imaging Biol       Date:  2015-08       Impact factor: 3.488

4.  Estimation of subject coregistration errors during multimodal preclinical imaging using separate instruments: origins and avoidance of artifacts.

Authors:  Jean-Philippe Dillenseger; Christian Goetz; Amira Sayeh; Chris Healy; Isabelle Duluc; Jean-Noël Freund; André Constantinesco; Gaëlle Aubertin-Kirch; Philippe Choquet
Journal:  J Med Imaging (Bellingham)       Date:  2017-08-22

5.  A Prototype High-Resolution Small-Animal PET Scanner Dedicated to Mouse Brain Imaging.

Authors:  Yongfeng Yang; Julien Bec; Jian Zhou; Mengxi Zhang; Martin S Judenhofer; Xiaowei Bai; Kun Di; Yibao Wu; Mercedes Rodriguez; Purushottam Dokhale; Kanai S Shah; Richard Farrell; Jinyi Qi; Simon R Cherry
Journal:  J Nucl Med       Date:  2016-03-24       Impact factor: 10.057

6.  Characterization of Large-Area SiPM Array for PET Applications.

Authors:  Junwei Du; Yongfeng Yang; Xiaowei Bai; Martin S Judenhofer; Eric Berg; Kun Di; Steve Buckley; Carl Jackson; Simon R Cherry
Journal:  IEEE Trans Nucl Sci       Date:  2016-02-15       Impact factor: 1.679

7.  Sodium [¹⁸F]fluoride PET/CT in myocardial infarction.

Authors:  Jeong Hee Han; Sue Yeon Lim; Min Su Lee; Won Woo Lee
Journal:  Mol Imaging Biol       Date:  2015-04       Impact factor: 3.488

8.  NEMA NU-4 performance evaluation of PETbox4, a high sensitivity dedicated PET preclinical tomograph.

Authors:  Z Gu; R Taschereau; N T Vu; H Wang; D L Prout; R W Silverman; B Bai; D B Stout; M E Phelps; A F Chatziioannou
Journal:  Phys Med Biol       Date:  2013-05-10       Impact factor: 3.609

9.  Performance evaluation of the MOLECUBES β-CUBE-a high spatial resolution and high sensitivity small animal PET scanner utilizing monolithic LYSO scintillation detectors.

Authors:  Srilalan Krishnamoorthy; Eric Blankemeyer; Pieter Mollet; Suleman Surti; Roel Van Holen; Joel S Karp
Journal:  Phys Med Biol       Date:  2018-07-27       Impact factor: 3.609

10.  Variations in PET/MRI Operations: Results from an International Survey Among 39 Active Sites.

Authors:  Wolfgang Peter Fendler; Johannes Czernin; Ken Herrmann; Thomas Beyer
Journal:  J Nucl Med       Date:  2016-08-11       Impact factor: 10.057

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.