Literature DB >> 26478600

A DOI Detector With Crystal Scatter Identification Capability for High Sensitivity and High Spatial Resolution PET Imaging.

Z Gu1, D L Prout1, R W Silverman1, H Herman1, A Dooraghi1, A F Chatziioannou1.   

Abstract

A new phoswich detector is being developed at the Crump Institute, aiming to provide improvements in sensitivity, and spatial resolution for PET. The detector configuration is comprised of two layers of pixelated scintillator crystal arrays, a glass light guide and a light detector. The annihilation photon entrance (top) layer is a 48 × 48 array of 1.01 × 1.01 × 7 mm3 LYSO crystals. The bottom layer is a 32 × 32 array of 1.55 × 1.55 × 9 mm3 BGO crystals. A tapered, multiple-element glass lightguide is used to couple the exit end of the BGO crystal array (52 × 52 mm2) to the photosensitive area of the Position Sensitive Photomultiplier Tube (46 × 46 mm2), allowing the creation of flat panel detectors without gaps between the detector modules. Both simulations and measurements were performed to evaluate the characteristics and benefits of the proposed design. The GATE Monte Carlo simulation indicated that the total fraction of the cross layer crystal scatter (CLCS) events in singles detection mode for this detector geometry is 13.2%. The large majority of these CLCS events (10.1% out of 13.2%) deposit most of their energy in a scintillator layer other than the layer of first interaction. Identification of those CLCS events for rejection or correction may lead to improvements in data quality and imaging performance. Physical measurements with the prototype detector showed that the LYSO, BGO and CLCS events were successfully identified using the delayed charge integration (DCI) technique, with more than 95% of the LYSO and BGO crystal elements clearly resolved. The measured peak-to-valley ratios (PVR) in the flood histograms were 3.5 for LYSO and 2.0 for BGO. For LYSO, the energy resolution ranged from 9.7% to 37.0% full width at half maximum (FWHM), with a mean of 13.4 ± 4.8%. For BGO the energy resolution ranged from 16.0% to 33.9% FWHM, with a mean of 18.6 ± 3.2%. In conclusion, these results demonstrate that the proposed detector is feasible and can potentially lead to a high spatial resolution, high sensitivity and DOI PET system.

Entities:  

Keywords:  BGO; DOI; GATE; LYSO; PET; contrast; crystal scatter; detector; phoswich

Year:  2015        PMID: 26478600      PMCID: PMC4608445          DOI: 10.1109/TNS.2015.2408333

Source DB:  PubMed          Journal:  IEEE Trans Nucl Sci        ISSN: 0018-9499            Impact factor:   1.679


  25 in total

Review 1.  Molecular imaging.

Authors:  R Weissleder; U Mahmood
Journal:  Radiology       Date:  2001-05       Impact factor: 11.105

Review 2.  Molecular imaging of small animals with dedicated PET tomographs.

Authors:  Arion F Chatziioannou
Journal:  Eur J Nucl Med Mol Imaging       Date:  2001-11-22       Impact factor: 9.236

3.  High-resolution PET detector design: modelling components of intrinsic spatial resolution.

Authors:  Jennifer R Stickel; Simon R Cherry
Journal:  Phys Med Biol       Date:  2005-01-21       Impact factor: 3.609

4.  Depth of interaction resolution measurements for a high resolution PET detector using position sensitive avalanche photodiodes.

Authors:  Yongfeng Yang; Purushottam A Dokhale; Robert W Silverman; Kanai S Shah; Mickel A McClish; Richard Farrell; Gerald Entine; Simon R Cherry
Journal:  Phys Med Biol       Date:  2006-04-10       Impact factor: 3.609

Review 5.  The 2006 Henry N. Wagner Lecture: Of mice and men (and positrons)--advances in PET imaging technology.

Authors:  Simon R Cherry
Journal:  J Nucl Med       Date:  2006-11       Impact factor: 10.057

6.  Depth of interaction decoding of a continuous crystal detector module.

Authors:  T Ling; T K Lewellen; R S Miyaoka
Journal:  Phys Med Biol       Date:  2007-03-29       Impact factor: 3.609

7.  Study of a high-resolution, 3D positioning cadmium zinc telluride detector for PET.

Authors:  Y Gu; J L Matteson; R T Skelton; A C Deal; E A Stephan; F Duttweiler; T M Gasaway; C S Levin
Journal:  Phys Med Biol       Date:  2011-02-18       Impact factor: 3.609

8.  Performance evaluation of the GE healthcare eXplore VISTA dual-ring small-animal PET scanner.

Authors:  Yuchuan Wang; Jurgen Seidel; Benjamin M W Tsui; Juan J Vaquero; Martin G Pomper
Journal:  J Nucl Med       Date:  2006-11       Impact factor: 10.057

Review 9.  Recent developments in PET detector technology.

Authors:  Tom K Lewellen
Journal:  Phys Med Biol       Date:  2008-08-11       Impact factor: 3.609

10.  Micro insert: a prototype full-ring PET device for improving the image resolution of a small-animal PET scanner.

Authors:  Heyu Wu; Debashish Pal; Tae Yong Song; Joseph A O'Sullivan; Yuan-Chuan Tai
Journal:  J Nucl Med       Date:  2008-09-15       Impact factor: 10.057

View more
  3 in total

1.  Preclinical positron emission tomography scanner based on a monolithic annulus of scintillator: initial design study.

Authors:  Alexander V Stolin; Peter F Martone; Gangadhar Jaliparthi; Raymond R Raylman
Journal:  J Med Imaging (Bellingham)       Date:  2017-01-05

2.  A layered single-side readout depth of interaction time-of-flight-PET detector.

Authors:  L Bläckberg; S Sajedi; G El Fakhri; H Sabet
Journal:  Phys Med Biol       Date:  2021-02-11       Impact factor: 3.609

3.  A digital phoswich detector using time-over-threshold for depth of interaction in PET.

Authors:  David L Prout; Zheng Gu; Max Shustef; Arion F Chatziioannou
Journal:  Phys Med Biol       Date:  2020-12-15       Impact factor: 3.609

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.