Christopher F Njeh1, Jason Parker, Joseph Spurgin, Elizabeth Rhoe.
Abstract
BACKGROUND: Carbon fiber (CF) is now the material of choice for radiation therapy couch tops. Initial designs included side metal bars for rigidity; however, with the advent of IGRT, involving on board imaging, new thicker CF couch tops without metal bars have been developed. The new design allows for excellent imaging at the expense of potentially unacceptable dose attenuation and perturbation.Entities:
Mesh:
Substances:
Year: 2012 PMID: 23140425 PMCID: PMC3549905 DOI: 10.1186/1748-717X-7-190
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Figure 1Photographs of the BrainLAB imaging couch top (ICT), extension and frameless extension.
Figure 2Printout from Pinnacle showing the water phantom and the modeled couch (the CF outer shell and the foam inner core) as well as the 2 mm skin ring around the water phantom.
Modeling of the BrainLAB Imaging couch top (ICT) in the Philips Pinnacle treatment planning system using different combinations of carbon fibers (CF) and foam densities (F) for 10 cm × 10 cm field size
| | | | |||||||
| 180 | 89.3 | 90.5 | 88.4 | 89.1 | 88.5 | 87.7 | 86.8 | 88.7 | 86.4 |
| 170 | 89.0 | 89.8 | 88.1 | 88.7 | 88.1 | 87.3 | 86.5 | 88.4 | 86.0 |
| 160 | 87.9 | 89.4 | 87.2 | 87.8 | 87.2 | 86.4 | 85.5 | 87.5 | 85.1 |
| 150 | 85.9 | 87.3 | 85.6 | 86.2 | 85.5 | 84.6 | 83.7 | 85.9 | 83.4 |
| 140 | 82.4 | 84.6 | 82.1 | 82.8 | 82.2 | 81.1 | 80.1 | 82.5 | 79.7 |
| 130 | 76.7 | 80.1 | 76.9 | 77.7 | 77.0 | 75.8 | 74.6 | 77.3 | 74.1 |
| 120 | 67.5 | 72.3 | 68.2 | 69.2 | 68.3 | 67.0 | 65.6 | 68.7 | 65.1 |
| 110 | 58.0 | 61.4 | 57.4 | 58.3 | 58.0 | 57.3 | 56.0 | 57.9 | 55.8 |
| 100 | 64.5 | 63.9 | 63.9 | 63.9 | 63.9 | 63.9 | 63.9 | 63.9 | 64.1 |
Modeling of the BrainLAB Imaging couch top (ICT) in the Philips Pinnacle treatment planning system using different combinations of carbon fibers (CF) and foam densities (F) for 10 cm × 10 cm field size
| | | | |||||||
| 180 | 89.3 | 1.31 | -1.04 | -0.26 | -0.93 | -1.82 | -2.83 | -0.71 | -3.28 |
| 170 | 89.0 | 0.94 | -0.97 | -0.30 | -0.97 | -1.87 | -2.77 | -0.64 | -3.33 |
| 160 | 87.9 | 1.72 | -0.79 | -0.10 | -0.79 | -1.70 | -2.72 | -0.44 | -3.18 |
| 150 | 85.9 | 1.68 | -0.30 | 0.40 | -0.41 | -1.46 | -2.51 | 0.05 | -2.86 |
| 140 | 82.4 | 2.67 | -0.36 | 0.49 | -0.24 | -1.58 | -2.79 | 0.12 | -3.28 |
| 130 | 76.7 | 4.40 | 0.23 | 1.27 | 0.36 | -1.21 | -2.77 | 0.75 | -3.42 |
| 120 | 67.5 | 7.12 | 1.05 | 2.53 | 1.19 | -0.73 | -2.81 | 1.79 | -3.55 |
| 110 | 58.0 | 5.83 | -1.06 | 0.49 | -0.03 | -1.23 | -3.47 | -0.20 | -3.82 |
| 100 | 64.5 | -0.96 | -0.96 | -0.96 | -0.96 | -0.96 | -0.96 | -0.96 | -0.65 |
| Sum | | 24.72 | -4.20 | 3.56 | -2.77 | -12.56 | -23.63 | 0.23 | -27.36 |
| 0.0059 | 0.07 | 0.27 | 0.18 | 0.00007 | 0.000005 | 0.99 | 0.00002 | ||
Note that the “measured dose (cGy)” is the dose measured as described in phantom dose measurement through the imaging couch top.
Modeling of the BrainLAB imaging couch top (ICT) in the Philips Pinnacle treatment planning system using different combinations of carbon fibers (CF) and foam densities (F) for 5 cm × 5 cm field size
| | | | |||||||
| 180 | 81.2 | 82.7 | 80.4 | 81.1 | 80.6 | 79.7 | 78.8 | 80.8 | 78.6 |
| 170 | 80.8 | 82.3 | 80 | 80.6 | 80.1 | 79.3 | 78.4 | 80.3 | 78.1 |
| 160 | 79.7 | 81.4 | 78.9 | 79.6 | 79 | 78.2 | 77.3 | 79.2 | 77 |
| 150 | 77.5 | 79.6 | 77.2 | 77.9 | 77.3 | 76.3 | 75.4 | 77.6 | 75.1 |
| 140 | 74.0 | 76.7 | 73.7 | 74.5 | 73.9 | 72.8 | 71.7 | 74.1 | 71.4 |
| 130 | 68.3 | 71.9 | 68.4 | 69.2 | 68.5 | 67.4 | 66.2 | 68.8 | 65.8 |
| 120 | 59.2 | 63.9 | 59.9 | 60.8 | 60 | 58.7 | 57.4 | 60.3 | 57 |
| 110 | 50.0 | 53.3 | 49.4 | 50.3 | 50 | 49.4 | 48.2 | 49.8 | 47.8 |
| 100 | 56.0 | 55.4 | 55.4 | 55.4 | 55.4 | 55.4 | 55.4 | 55.4 | 55.4 |
Modeling of the BrainLAB imaging couch top (ICT) in the Philips Pinnacle treatment planning system using different combinations of carbon fibers (CF) and foam densities (F) for 5 cm × 5 cm field size
| | | | |||||||
| 180 | 81.2 | 1.84 | -0.99 | -0.13 | -0.75 | -1.86 | -2.96 | -0.50 | -3.21 |
| 170 | 80.8 | 1.80 | -1.04 | -0.30 | -0.92 | -1.91 | -3.02 | -0.67 | -3.39 |
| 160 | 79.7 | 2.19 | -0.94 | -0.07 | -0.82 | -1.82 | -2.95 | -0.57 | -3.33 |
| 150 | 77.5 | 2.71 | -0.39 | 0.51 | -0.26 | -1.55 | -2.71 | 0.13 | -3.10 |
| 140 | 74.0 | 3.66 | -0.40 | 0.68 | -0.13 | -1.62 | -3.10 | 0.14 | -3.51 |
| 130 | 68.3 | 5.24 | 0.12 | 1.29 | 0.26 | -1.35 | -3.10 | 0.70 | -3.69 |
| 120 | 59.2 | 7.90 | 1.14 | 2.66 | 1.31 | -0.88 | -3.08 | 1.82 | -3.75 |
| 110 | 50.0 | 6.51 | -1.28 | 0.52 | -0.08 | -1.28 | -3.68 | -0.48 | -4.48 |
| 100 | 56.0 | -1.15 | -1.15 | -1.15 | -1.15 | -1.15 | -1.15 | -1.15 | -1.15 |
| sum | | 30.69 | -4.95 | 4.01 | -2.54 | -13.42 | -25.77 | -0.59 | -29.62 |
| 0.002 | 0.047 | 0.23 | 0.21 | 0.00004 | 0.000005 | 0.76 | 0.000005 | ||
Note that the “measured dose (cGy)” is the dose measured as described in section “Phantom dose measurement” through the imaging couch top.
Validation of BrainLAB imaging couch top (ICT) modeled in iPlan RT Dose treatment planning system (TPS); comparing measured dose with TPS computed dose
| Angle | Measured dose (cGy) | No ICT | With ICT | No ICT | With ICT |
| 180 | 89.3 | 92.45 | 89.34 | 3.49 | 0.01 |
| 170 | 89.0 | 92.16 | 89.00 | 3.59 | 0.04 |
| 160 | 87.9 | 91.24 | 87.92 | 3.81 | 0.03 |
| 150 | 85.9 | 89.34 | 85.92 | 4.06 | 0.07 |
| 140 | 82.4 | 86.63 | 82.44 | 5.14 | 0.05 |
| 130 | 76.7 | 82.01 | 77.59 | 6.89 | 1.13 |
| 120 | 67.5 | 74.27 | 67.32 | 10.04 | -0.26 |
| 110 | 58.0 | 63.17 | 58.55 | 8.89 | 0.92 |
| 100 | 64.5 | 65.42 | 65.43 | 1.40 | 1.42 |
| 180 | 81.2 | 84.09 | 80.48 | 3.55 | 0.90 |
| 170 | 80.8 | 83.76 | 80.14 | 3.61 | 0.87 |
| 160 | 79.7 | 82.74 | 79.03 | 3.88 | 0.78 |
| 150 | 77.5 | 80.67 | 76.76 | 4.09 | 0.96 |
| 140 | 74.0 | 77.79 | 73.26 | 5.13 | 0.99 |
| 130 | 68.3 | 72.91 | 67.74 | 6.72 | 0.85 |
| 120 | 59.2 | 64.71 | 58.56 | 9.27 | 1.12 |
| 110 | 50.0 | 53.7 | 49.78 | 7.31 | 0.52 |
| 100 | 56.0 | 55.74 | 55.74 | -0.55 | 0.55 |
Note that the “measured dose (cGy)” is the dose measured as described in phantom dose measurement through the imaging couch top.
Pinnacle and iPlan RT Dose TPS predicted dose attenuation of the BrainLAB imaging couch top (ICT) compared to previously reported values for 6 MV photon beam for 10 cm × 10 cm field sizes and 5 cm × 5 cm field sizes
| | ||||||
|---|---|---|---|---|---|---|
| 180 | 3.4 | 1.99 | 3.4 | 4.3 | 2.30 | 4.9 |
| 170 | 3.4 | 1.56 | 3.5 | 4.3 | 2.43 | 5.0 |
| 160 | 3.6 | 2.13 | 3.7 | 4.5 | 2.70 | 5.2 |
| 150 | 3.8 | 1.60 | 4.3 | 4.8 | 2.51 | 5.6 |
| 140 | 4.8 | 2.48 | 4.9 | 5.8 | 3.39 | 6.5 |
| 130 | 5.4 | 3.50 | 6.2 | 7.1 | 4.31 | 7.7 |
| 120 | 9.4 | 4.98 | 8.3 | 9.5 | 5.63 | 10.0 |
| 110 | 7.3 | 5.70 | 8.0 | 7.3 | 6.57 | 9.4 |
| 100 | 0.0 | 0.00 | -0.7 | 0.0 | 0.00 | 0.4 |
Pinnacle TPS determined maximum skin dose as a percentage of the Dmax dose for a 10 cm × 10 cm direct incidence
| | ||||
|---|---|---|---|---|
| 180 | 61.4 | 97.4 | 53.3 | 86.5 |
| 170 | 60.8 | 96.7 | 52.6 | 86.5 |
| 160 | 59.1 | 97.4 | 51.0 | 87.4 |
| 150 | 56.1 | 100.5 | 48.3 | 89.5 |
| 140 | 59.0 | 105.9 | 49.2 | 93.4 |
| 130 | 67.9 | 113.9 | 56.2 | 98.5 |
| 120 | 77.8 | 123.2 | 66.1 | 109.0 |
| 110 | 82.6 | 120.6 | 73.6 | 110.8 |
| 100 | 121.6 | 122.3 | 66.2 | 67.4 |