Jan K H Seppälä1, Jarmo A J Kulmala.
Abstract
The use of solid carbon fiber table materials in radiotherapy has become more common with the implementation of image-guided radiotherapy (IGRT), since the solid materials give less imaging artifacts than the so-called tennis racket couchtops. The downside of the solid carbon fiber couch inserts is that they increase the beam attenuation, resulting in increased surface doses and inaccuracies in determine the dose in the patient. The purpose of this study was to evaluate the interaction of 6 and 15 MV photons with eight different couch inserts. The presented results enable direct comparison of the attenuation properties of the studied couchtops. With a direct posterior beam the maximum attenuations reach 3.6% and 2.4% with 6 and 15 MV, respectively. The measured maximum attenuation by a couchtop with an oblique gantry angle was 10.8% and 7.4% at 6 and 15 MV energies, respectively. The skin-sparing effect was decreased substantially with every couchtop. The highest increases in surface doses were recorded to be four- and threefold, as compared to the direct posterior open field surface doses of 6 and 15 MV, respectively. In conclusion, the carbon fiber tabletops decrease the skin-sparing effect of megavoltage photon energies. The increased beam attenuation and skin doses should be taken into account in the process of treatment planning.Entities:
Mesh:
Substances:
Year: 2011 PMID: 22089010 PMCID: PMC5718753 DOI: 10.1120/jacmp.v12i4.3554
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Couch inserts studied: 1) BrainLAB imaging couch top, 2) Qfix kVue Standard, 3) Qfix kVue DoseMax, 4) MEDTEC model MT‐IL 3303, 5) universal sandwich panel, 6) Varian grid insert, 7) DIGNITY AirPlate, 8) Varian Exact IGRT couchtop.
Figure 2Measurement setup for couchtop beam attenuation (a) and surface dose (b) measurements. With attenuation measurements, the cylindrical ionization chamber was in the middle of the phantom. With surface dose measurements, the parallel‐plane chamber was facing the couchtop.
Figure 3Beam attenuations of eight different couch inserts measured with energies of 6 (a) and 15 MV (b) with various gantry angles. The measured dose difference (%) is with respect to the dose measured with a gantry angle of 0°.
The direct posterior (gantry angle 180°), maximum, and average beam attenuations (in percentage) measured with eight different couch inserts with energies of 6 and 15 MV. The average beam attenuation was calculated between gantry angles of 100°–180°. The couch inserts are in descending order with respect to attenuation with a direct posterior 6 MV beam.
|
|
| |||||
|---|---|---|---|---|---|---|
|
|
|
|
|
|
| |
| BrainLAB | 3.6 | 8.7 | 5.0 | 2.4 | 5.9 | 3.3 |
| Qfix kVue Standard | 2.1 | 5.2 | 2.9 | 1.4 | 3.3 | 1.9 |
| MEDTEC | 1.9 | 7.1 | 2.9 | 1.5 | 5.0 | 2.1 |
| Varian Exact IGRT | 1.9 | 4.7 | 2.4 | 1.3 | 3.1 | 1.6 |
| DIGNITY AirPlate | 1.9 | 3.6 | 2.3 | 1.2 | 2.4 | 1.5 |
| Universal Sandwich Panel | 1.5 | 7.3 | 2.8 | 1.0 | 5.0 | 2.0 |
| Qfix kVue DoseMax | 1.3 | 8.1 | 2.2 | 1.0 | 5.4 | 1.4 |
| Varian Grid Insert | 0.3 | 10.8 | 1.3 | 0.2 | 7.4 | 0.9 |
Measured surface doses relative to the dose maximum (in percentage) with different couch inserts in the beam with energies of 6 and 15 MV. Measurements were performed with and open fields with a direct posterior beam (gantry angle 180°). The corrected surface dose takes into account the perturbation effect of the used plane‐parallel ionization chamber. The couch inserts are in descending order with respect to relative surface dose.
|
|
| |||
|---|---|---|---|---|
|
|
|
|
| |
| Measured Surface Dose | 44.3 | 53.4 | 27.6 | 39.1 |
| Corrected Surface Dose | 35.2 | 44.3 | 21.0 | 32.6 |
| Dose at Dmax | 100.0 | 100.0 | 100.0 | 100.0 |
| BrainLAB | 98.6 | 99.4 | 84.5 | 89.9 |
| Varian Exact IGRT | 90.8 | 94.0 | 69.6 | 77.7 |
| Universal Sandwich Panel | 90.2 | 93.2 | 69.2 | 76.9 |
| MEDTEC | 90.1 | 93.4 | 69.1 | 77.3 |
| Qfix kVue Standard | 88.5 | 92.1 | 66.9 | 75.2 |
| DIGNITY AirPlate | 86.0 | 89.8 | 66.0 | 74.3 |
| Qfix kVue DoseMax | 75.1 | 80.9 | 52.8 | 63.0 |
| Varian Grid Insert | 61.2 | 69.0 | 41.0 | 51.7 |