Dose perturbations by two carbon fiber treatment couches and the ability of a commercial treatment planning system to predict these effects.

L H Gerig1, M Niedbala, B J Nyiri.   

Abstract

PURPOSE: To measure the effect of the treatment couch on dose distributions and to investigate the ability of a modern planning system to accurately model these effects.
METHODS: This work measured the dose perturbation at depth and in the dose buildup region when one of two treatment couches, CIVCO (formerly MED-TEC) or Medical Intelligence, was placed between a photon beam source (6, 10, and 18 MV) and the phantom. Beam attenuation was measured in the center of a cylindrical acrylic phantom with a Farmer type ion chamber at multiple gantry angles. Dose buildup was measured in Solid Water with plane parallel ion chambers (NACP-02 and PTW Markus) with the beam normal to both the phantom and couch surfaces. The effective point of measurement method as described [M. R. McEwen et al. "The effective point of measurement of ionization chambers and the build-up anomaly in MV x-ray beams," Med. Phys. 35(3), 950-958 (2008)] was employed to calculate dose in the buildup region. Both experiments were modeled in XiO. Images of the treatment couches were merged with images of the phantoms such that they were included as part of the "patient" image. Dose distributions calculated with superposition and fast superposition algorithms were compared to measurement.
RESULTS: The two treatment couches have different radiological signatures and dissimilar water equivalent thicknesses (4.2 vs 6.3 mm.) Maximum attenuation was 7%. Both couches caused significant loss of skin sparing, the worst case showing an increase in surface dose from 17% (no couch) to 88% (with couch). The TPS accurately predicted the surface dose (+/-3%) and the attenuation at depth when the phantom was in contact with the couch. For the open beam the TPS was less successful in the buildup region.
CONCLUSIONS: The treatment couch is not radio-transparent. Its presence between the patient and beam source significantly alters dose in the patient. For the most part, a modern treatment planning system can adequately predict the altered dose distribution.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20175495     DOI: 10.1118/1.3271364

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  13 in total

1.  Effects of Siemens TT-D carbon fiber table top on beam attenuation, and build up region of 6 MV photon beam.

Authors:  Asma Sheykhoo; Sara Abdollahi; Mohammad Hadi Hadizadeh Yazdi; Mahdi Ghorbani; Mohammad Mohammadi
Journal:  Rep Pract Oncol Radiother       Date:  2016-10-17

2.  Verification of the dose attenuation of a newly developed vacuum cushion for intensity-modulated radiation therapy of prostate cancer.

Authors:  Toru Takakura; Yoshiyuki Ito; Akinori Higashikawa; Tomohiro Nishiyama; Takashi Sakamoto
Journal:  Radiol Phys Technol       Date:  2016-06-03

3.  The clinical impact of the couch top and rails on IMRT and arc therapy.

Authors:  Kiley B Pulliam; Rebecca M Howell; David Followill; Dershan Luo; R Allen White; Stephen F Kry
Journal:  Phys Med Biol       Date:  2011-11-04       Impact factor: 3.609

4.  Reduction of Potential Risk for Skin Toxicity in Megavoltage Radiotherapy Using a Novel Rigid Couch.

Authors:  Mikoto Tamura; Hajime Monzen; Kenji Matsumoto; Masahiko Okumura; Hiroshi Doi; Yasumasa Nishimura
Journal:  In Vivo       Date:  2018 May-Jun       Impact factor: 2.155

5.  A validation of carbon fiber imaging couch top modeling in two radiation therapy treatment planning systems: Philips Pinnacle3 and BrainLAB iPlan RT Dose.

Authors:  Christopher F Njeh; Jason Parker; Joseph Spurgin; Elizabeth Rhoe
Journal:  Radiat Oncol       Date:  2012-11-09       Impact factor: 3.481

6.  Increased beam attenuation and surface dose by different couch inserts of treatment tables used in megavoltage radiotherapy.

Authors:  Jan K H Seppälä; Jarmo A J Kulmala
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

7.  Quantification and comparison the dosimetric impact of two treatment couch model in VMAT.

Authors:  Ruohui Zhang; Yulan Gao; Wenwen Bai
Journal:  J Appl Clin Med Phys       Date:  2017-11-02       Impact factor: 2.102

8.  An Update of Couch Effect on the Attenuation of Megavoltage Radiotherapy Beam and the Variation of Absorbed Dose in the Build-up Region.

Authors:  T Sedaghatian; M Momennezhad; S H Rasta; Y Makhdoomi; S Abdollahian
Journal:  J Biomed Phys Eng       Date:  2017-09-01

9.  Evaluating and modeling of photon beam attenuation by a standard treatment couch.

Authors:  Zhihui Hu; Jianrong Dai; Liang Li; Yin Cao; Guishan Fu
Journal:  J Appl Clin Med Phys       Date:  2011-07-12       Impact factor: 2.102

10.  Radiation dosimetry effect evaluation of a carbon fiber couch on novel uRT-linac 506c accelerator.

Authors:  Dazhen Jiang; Zhen Cao; Yongchang Wei; Tingting Cao; Jiuling Shen; Conghua Xie; Yunfeng Zhou; Hui Liu; Jun Zhang
Journal:  Sci Rep       Date:  2021-06-29       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.