Evaluation of the Sinmed Mastercouch as replacement for a standard couch.

Sofie Gillis1, Samuel Bral, Carlos De Wagter, Cristina Derie, Leen Paelinck, Koen Van Vaerenbergh, Marc Coghe, Gert De Meerleer, Wilfried De Neve.   

Abstract

BACKGROUND AND
PURPOSE: Introducing into practice a new carbon-fibre couch necessitates its evaluation with regard to the present clinical situation.
MATERIALS AND METHODS: In this study, a geometric and dosimetric evaluation has been made for the Sinmed Mastercouch as a replacement for the Elekta C-arm couch with Mylar-tennis racket combination. Geometrically, feasible gantry angles with regard to collision and beam intersection were discriminated as function of isocentric table rotation for 10 treatment isocentres. Dosimetrically, the build-up distortion and attenuation by the aforementioned tabletops and a carbon-fibre tabletop of an Elekta Precise simulator was measured. Finally, the clinical implications of these influences were assessed for a 3-field prostate treatment in three configurations: Mastercouch, C-arm couch and no-intersection situation.
RESULTS: With regard to collision, the largest advantages are observed for the Mastercouch with the Omega-shaped treatment module compared to the C-arm couch for isocentres located in the upper part of the body, thanks to its shape and the absence of any metal. Dosimetrically, one has to take into account the build-up loss and attenuation by beam intersection with Mastercouch and the carbon-fibre edges of the tennis racket (C-arm couch). The clinical relevance of these dosimetric aspects depends on the dose delivered by the compromised beams.

Entities:  

Mesh:

Year:  2005        PMID: 15922469     DOI: 10.1016/j.radonc.2005.04.013

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  4 in total

1.  A validation of carbon fiber imaging couch top modeling in two radiation therapy treatment planning systems: Philips Pinnacle3 and BrainLAB iPlan RT Dose.

Authors:  Christopher F Njeh; Jason Parker; Joseph Spurgin; Elizabeth Rhoe
Journal:  Radiat Oncol       Date:  2012-11-09       Impact factor: 3.481

2.  Impact of 6MV photon beam attenuation by carbon fiber couch and immobilization devices in IMRT planning and dose delivery.

Authors:  R K Munjal; P S Negi; A G Babu; S N Sinha; A K Anand; T Kataria
Journal:  J Med Phys       Date:  2006-04

3.  Increased beam attenuation and surface dose by different couch inserts of treatment tables used in megavoltage radiotherapy.

Authors:  Jan K H Seppälä; Jarmo A J Kulmala
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

4.  Evaluating and modeling of photon beam attenuation by a standard treatment couch.

Authors:  Zhihui Hu; Jianrong Dai; Liang Li; Yin Cao; Guishan Fu
Journal:  J Appl Clin Med Phys       Date:  2011-07-12       Impact factor: 2.102

  4 in total