Artifacts in magnetic resonance imaging and computed tomography caused by dental materials.

Thomas Klinke1, Amro Daboul, Juliane Maron, Tomasz Gredes, Ralf Puls, Ahmad Jaghsi, Reiner Biffar.   

Abstract

BACKGROUND: Artifacts caused by dental restorations, such as dental crowns, dental fillings and orthodontic appliances, are a common problem in MRI and CT scans of the head and neck. The aim of this in-vitro study was to identify and evaluate the artifacts produced by different dental restoration materials in CT and MRI images.
METHODS: Test samples of 44 materials (Metal and Non-Metal) commonly used in dental restorations were fabricated and embedded with reference specimens in gelatin moulds. MRI imaging of 1.5T and CT scan were performed on the samples and evaluated in two dimensions. Artifact size and distortions were measured using a digital image analysis software.
RESULTS: In MRI, 13 out of 44 materials produced artifacts, while in CT 41 out of 44 materials showed artifacts. Artifacts produced in both MRI and CT images were categorized according to the size of the artifact. SIGNIFICANCE: Metal based restoration materials had strong influence on CT and less artifacts in MRI images. Rare earth elements such as Ytterbium trifluoride found in composites caused artifacts in both MRI and CT. Recognizing these findings would help dental materials manufacturers and developers to produce materials which can cause less artifacts in MRI and CT images.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22384071      PMCID: PMC3285178          DOI: 10.1371/journal.pone.0031766

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Today, various kinds of materials are used in the dental treatment, materials such as metal alloys, composites, acrylics, porcelain and ceramics are used as filling materials and in dental prosthesis like crowns, dental bridges and dentures. The properties and specifications of these materials are well defined and studied. However, their influence and effect on image quality with computed tomography (CT) and magnetic resonance imaging (MRI) is not covered sufficiently in the literature. In CT imaging, Streak artifacts are a common problem. The presence of high attenuation metal objects in the field of view such as dental restorations, orthodontic bands, surgical plates and pins can cause this type of artifacts. That is because the metal materials highly attenuate the x-ray beam resulting in incorrect high attenuation values of objects behind the metal. However, in MRI, images are created using a combination of strong uniform magnetic field and radio frequency pulses. All substances when placed in a magnetic field are magnetized at various degrees depending on their magnetic susceptibility. The variations in the magnetic field strength that occur on the interface between the dental material and the adjacent tissues will cause magnetic field distortions and signal loss which will generate an artifact in the image. The artifact severity will vary depending on the shape, position, orientation and number of objects in the image, sequence type used and sequence parameters [1]–[5]. MRI and CT image quality can be dramatically degraded by artifacts caused by dental materials, which limits their usefulness as diagnostic tools. Artifacts might obscure a pathology (e.g. Tumors, Inflammatory tissues) or obscure the anatomy of the area examined and make it difficult to locate anatomical structure for surgical procedures [2], [6]. Many studies have investigated the artifacts generated by metals used in medicine and dentistry on MRI and CT [1]–[5], [7]–[17]. However, the effect of non-metal based materials on MRI and CT image quality was not covered sufficiently. Hinshaw et al. [13] discussed artifacts that were caused by some materials commonly used in dental restorations, mainly stainless steel materials, such as orthodontic bands, braces, metal pins and posts.Fache et al. [11] evaluated a variety of dental materials and the mechanism of artifact production in MRI. Their study analyzed the composition and magnetic susceptibility of tested dental materials and compared the results with the artifacts produced by each material. It was concluded that the size of an artifact is related to the magnetic permeability of the dental material. New et al. [18] investigated the deflection force of clips and the intensity of dental amalgam in MRI artifacts. Starcukova et al. [19] showed that MR imaging without artifacts is possible even close to dental devices (amalgam, precious alloys and titanium) only if they are made of materials with low magnetic susceptibility. However, they mentioned that not all dental materials in the current use meet this criterion of low magnetic susceptibility. Although the previously mentioned studies have described the effects of metallic objects on MRI interpretation, few have addressed the effect of non-metal based materials on MRI or CT image quality. The purpose of this study was to identify and evaluate the artifacts produced by different dental materials (Metal and Non-Metal) in CT and MRI images.

Materials and Methods

Samples of 44 different materials (metal and non-metal), which are commonly used in dental practice, were included in this study (N = 44). The samples were divided into 4 groups; group 1: Composites (N = 31), group 2: temporary filling materials (N = 4), group 3: Dental ceramics (N = 5) and group 4: Metal alloys (N = 4). The test specimens were wedge-shaped with the dimensions of 16.0 mm in length by 9.0 mm in width, 3.5 mm high at the rear and 1 mm high at the tip (Figure 1a). Additionally, acrylic round reference specimens with an 11.0 mm diameter by 3.0 mm height were made (Figure 1b). We selected the wedge shape to have different material thickness from the tip and rear of each wedge, which will allow comparing the effect of artifacts produced by the test specimens of varying thickness on the adjacent acrylic round reference specimens.
Figure 1

Dimensions of (A) wedge-shaped specimens and (B) reference.

Both, test specimens and reference specimens were embedded in gelatin (RUF Lebensmittelwerk KG, Quakenbrueck, Germany). The gelatin was mixed with water, cooked at 80°C for 2 minutes, poured in a mould about 60 mm from the base. After the first layer of gelatin had gelled, the tested samples and reference samples were placed in the middle of the mould. A second layer of gelatin was poured to entirely fill the mould. The embedded samples were placed in a CT (Sensation 16, Siemens, Erlangen Germany) and 1.5T MRI (Symphony 1.5T, Siemens, Erlangen Germany) respectively. A spiral CT technique was used with 4 mm thickness, the parameters used were: 250 mA, 120 kV, 0.75 collimation and re-constructive thickness of 1 mm. In the MRI, images were taken in the axial and sagittal planes. A T1-weighted spin-echo sequence (TR: 650 ms, TE: 40 ms, TSE bandwidth: 139 Hz/Pixel) and T2-weighted turbo spin-echo sequence (TR: 3000 ms, TE: 90 ms, TSE bandwidth: 130 Hz/Pixel). Images were taken with the following parameters: thickness 3.0 mm, matrix size 512×512 pixel; field of view (FOV) 250×250 mm. Area of interest (AOI) was selected, artifact size and distortions in MRI were measured using Dicom image processing software (Osirix v 3.8 32bit). Lengths of artifacts produced in the CT were measured using an analyzing software (GE Advantage Workstation AW 4.207). In both softwares, all measurements were performed with the help of multi-planar reconstruction technique (MPR).

Results

In both MRI and CT, the thickness of the wedge-shaped samples had no significant influence on artifacts produced. Artifacts produced in MRI and CT were categorized according to the size and direction of the artifact. the artifact categories (Table 1) were in a similar manner to the classifications of Hinshaw et al. [13] and Behr et al. [4]. Artifacts produced by each group of materials were categorized: Composites (Table 2), Temporary filling materials (Table 3), Ceramics (Table 4) and Metal alloys (Table 5).
Table 1

MRI and CT artifact categories.

Category0123
MRINo artifactsArtifact less than 30.0 mm2 Artifact 30.0 mm2 to 60.0 mm2 Artifact more than 60.0 mm2
CTNo artifactsStreak artifact <15.0 mmStreak artifact 15.0–30.0 mmStreak artifact >30.0 mm
Table 2

MRI and CT artifacts produced by composites (Group 1).

MaterialCT artifact categoryMRI T1 artifact categoryMRI T1 artifact Shape changeMRI T2 artifact categoryMRI T2 artifact Shape change
Adaptic®100
X-flow™100
Quixfil™100
Core X™200
Dyract® Xtra200
Ceram·X Duo200
Esthet•X®300
Dyract® flow200
Spectrum®TPH®300
X-tra fil200
Grandio100
Admira300
Twinky Star200
Arabesk®200
Tetric EvoCeram®31−/+−1−/(+)
Tetric Evoflow®31−/+−1−/(+)
Adamant®31−/+−1−/(+)
InTen-S®21−/+−1−/(+)
Tetric® Flow31−/+−1−/+
Tetric Ceram®31−/+−1−/(+)
Compoglass® F31−/+−1−/(+)
Heliomolar®21−/+−1−/(+)
Helio Progress®000
Filtek™100
3 M™ Z100™ MP100
Nanosit™300
Synergy D6100
Solidex000
Henry Schein®200
Charisma®200
Revolcin® Flow100

−/+ Test specimen not recognizable, adjacent reference specimen recognizable.

−/+− Test specimen not recognizable, adjacent reference specimen partially recognizable.

−/(+) Test specimen not recognizable, adjacent reference specimen recognizable but changed size or shape.

Table 3

MRI and CT artifacts produced by temporary filling materials (Group 2).

MaterialCT artifact categoryMRI T1 artifact categoryMRI T1 artifact Shape changeMRI T2 artifact categoryMRI T2 artifact Shape change
Clip000
Guttapercha200
Ketac™200
Cavit™ G21−/+−1−/+

−/+ test specimen not recognizable, adjacent reference specimen recognizable.

−/+− test specimen not recognizable, adjacent reference specimen partially recognizable.

Table 4

MRI and CT artifacts produced by Cermics (Group 3).

MaterialCT artifact categoryMRI T1 artifact categoryMRI T1 artifact Shape changeMRI T2 artifact categoryMRI T2 artifact Shape change
Zirconium dioxide300
IPS Empress®100
Vita Omega 900200
Cergo200
Duceragold100
Table 5

MRI and CT artifacts produced by Metal alloys (Group 4).

MaterialCT artifact categoryMRI T1 artifact categoryMRI T1 artifact Shape changeMRI T2 artifact categoryMRI T2 artifact Shape change
Amalcap® Plus31−/+1−/+
Remanium® Star31−/+1−/+
Degunorm®31−/+1−/+
Kavo Everest® Titan31−/+−1−/+−

−/+ test specimen not recognizable, adjacent reference specimen recognizable.

−/+− test specimen not recognizable, adjacent reference specimen partially recognizable.

−/+ Test specimen not recognizable, adjacent reference specimen recognizable. −/+− Test specimen not recognizable, adjacent reference specimen partially recognizable. −/(+) Test specimen not recognizable, adjacent reference specimen recognizable but changed size or shape. −/+ test specimen not recognizable, adjacent reference specimen recognizable. −/+− test specimen not recognizable, adjacent reference specimen partially recognizable. −/+ test specimen not recognizable, adjacent reference specimen recognizable. −/+− test specimen not recognizable, adjacent reference specimen partially recognizable. In MR images, there were no significant differences in artifact area extent between T1 and T2. In group 1 (Composites) 23 materials showed no artifacts (category 0), while 8 materials were assigned to category 1 (Figure 2). In group 2 (Temporary filling materials), 3 materials showed no artifacts, while only one material (Cavit) was classified in category 1. In group 3 (Ceramics) all 5 materials showed no artifacts. In group 4 (metal alloys) 4 materials were classified in category 1, while 2 materials (Remanium Star and Degunorm) showed object projections in an area far from the object itself(Figure 3).
Figure 2

MRI artifact produced by Tetric EvoCeram in (A) T1 and (B) T2.

Figure 3

MRI artifact produced by Remanium in (A) T1 and (B) T2.

In CT, group 1 (Composites) 2 materials (Helio Progress and Solidex) showed no streak artifacts (Category 1), 8 materials showed artifacts less than 15.0 mm (Category 2), 11 materials showed streak artifacts between 15.0 mm to 30.0 mm (Category 3) and 10 materials showed artifacts larger than 30.0 mm (Category 4). In group 2 (Temporary filling materials) 1 material (Clip) showed no artifacts (category 1) and 3 materials showed streak artifacts between 15.0 mm to 30.0 mm (Group 3). In group 3 (Ceramics), 2 materials (IPS Empress and Duceragold) showed artifacts less than 15.0 mm (Category 2), 2 materials (Cergo and Vita Omega 900) showed artifacts between 15.0 mm to 30.0 mm (Category 3) and 1 material (Zirconium dioxide) showed artifacts larger than 30.0 mm (Category 4) (Figure 4). In group 4 (Metal alloys), all 4 materials showed artifacts larger than 30.0 mm (Category 4)(Figure 5).
Figure 4

Streak artifact produced by Zirconium dioxide in CT.

Figure 5

Streak artifact produced by Amalcap in CT.

Discussion

Various metal restorations and prosthesis may limit the usefulness of CT and MRI imaging, either by degrading the quality of the image or causing disturbances in the image, both of which will complicate the image interpretation and subsequently the diagnosis. In CT, metal based materials can cause severe artifacts [20], because attenuation data of the CT scan are distorted by the high density of metallic objects, leading to inconsistencies which prevents adequate calculation of the projection data, commonly resulting in a star burst artifact which consists of radiating lines originating from the metal alloy. In MRI, the presence of ferromagnetic metals in some of the dental materials causes magnetic field inhomogeneity [21], where metal-based materials create their own magnetic field and dramatically alter precession frequencies of protons in the adjacent tissues. Tissues adjacent to ferromagnetic components become influenced by the induced magnetic field of the metal, therefore, they either fail to precess or do so at a different frequency, hence they do not generate a useful signal. However, in this study, it was shown that not only metal based dental materials causes susceptibility artifacts, but also Non-Metal materials can cause artifacts and disturbances in the CT and MRI images. The ceramic frame material (Zirconium dioxide) led surprisingly to the same effect of metal-based materials and disturbed the delineation of anatomic structures in the CT images. Furthermore, Ingredients like Ytterbium trifluoride, Ferric oxide and Lanthanum oxide, which can be found in composites as coloring agents caused image disturbances in CT and MRI. In MRI, this is due to the fact that these materials contain some ferromagnetic metal ingredients. According to Eggers et al. [10], even small amounts of a ferromagnetic substance can cause an extensive signal void in the image. Diamagnetic materials such as gold and paramagnetic materials like titanium are used extensively in oral restorations and are less likely to create artifacts in MRI. However, in this study they generated artifacts and distortions in both MRI and CT images. That is because the alloys contained traces of others ferromagnetic metals such as iron. Furthermore, paramagnetic materials that are used as additives in dental materials and prosthetic appliances could cause artifacts. In a study done by Bartels et al. [22] paramagnetic substances in vascular stents caused artifacts on MRI, the authors explained that the generated artifacts depended on the MRI sequence used and the material size and thickness. Camacho et al [23] investigated MRI artifacts caused by radiofrequency eddy currents. They concluded that the resulted substantial signal intensity artifact, in addition to any susceptibility effect, also depends on the shape, orientation, and material of the object under investigation. As mentioned earlier, the magnitude of susceptibility artifacts in MRI is also related to the type of imaging sequence used. Gradient echo (GRE) sequences are sensitive to the presence of metal, where intravoxel dephasing is the predominant cause of signal loss, resulting in a dark or black area around the metal on the processed images. [24] Shortening the echo time (TE) and decreasing voxel size can be used to reduce the degree of intravoxel dephasing seen on GRE acquisition. [25] On the other hand, Spin-echo (SE) sequences have a 180° RF-pulse that refocuses the spins at the echo time and thereby diminishes the phase shifts in the voxel which are caused by local static magnetic field gradients. The refocusing 180° RF-pulse makes the SE sequence less sensitive to susceptibility effects [26]. In this study, we did not try artifact reduction with short TE, and the sequence protocol in both scans (T1 with TE 40 ms, T2 with TE 90 ms) showed similar artifact sizes and shapes for most of the material tested. Considering the results of this study, the use of ceramic materials like Zirconium dioxide in crowns, bridges or permanent fillings has to be critically analyzed in the case of CT or MR imaging of the lower mid face. Further more, dental materials developers would benefit from the results of this study to produce materials that cause less artifacts and distortions in CT and MRI images.
  25 in total

1.  Artefacts in magnetic resonance imaging caused by dental material.

Authors:  Georg Eggers; Marcus Rieker; Bodo Kress; Jochen Fiebach; Hartmut Dickhaus; Stefan Hassfeld
Journal:  MAGMA       Date:  2005-03-22       Impact factor: 2.310

2.  Experimental and computational analyses of the effects of slice distortion from a metallic sphere in an MRI phantom.

Authors:  Tim A J Hopper; Branimir Vasilić; James M Pope; Catherine E Jones; Charles L Epstein; Hee Kwon Song; Felix W Wehrli
Journal:  Magn Reson Imaging       Date:  2006-09-07       Impact factor: 2.546

3.  Magnetic susceptibility and electrical conductivity of metallic dental materials and their impact on MR imaging artifacts.

Authors:  Jana Starcuková; Zenon Starcuk; Hana Hubálková; Igor Linetskiy
Journal:  Dent Mater       Date:  2007-09-19       Impact factor: 5.304

Review 4.  Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT.

Authors:  Mi-Jung Lee; Sungjun Kim; Sung-Ah Lee; Ho-Taek Song; Yong-Min Huh; Dae-Hong Kim; Seung Hwan Han; Jin-Suck Suh
Journal:  Radiographics       Date:  2007 May-Jun       Impact factor: 5.333

5.  [Artifacts from dental metal alloys in magnetic resonance imaging].

Authors:  Zheng Cao; Li-li Chen; Xiang-yang Gong
Journal:  Zhonghua Yi Xue Za Zhi       Date:  2008-07-08

6.  Metallic artifacts in MRI caused by dental alloys and magnetic keeper.

Authors:  Darline Destine; Hiroshi Mizutani; Yoshimasa Igarashi
Journal:  Nihon Hotetsu Shika Gakkai Zasshi       Date:  2008-04

7.  Nonsusceptibility artifacts due to metallic objects in MR imaging.

Authors:  C R Camacho; D B Plewes; R M Henkelman
Journal:  J Magn Reson Imaging       Date:  1995 Jan-Feb       Impact factor: 4.813

Review 8.  Dental alloys and magnetic resonance imaging.

Authors:  Hana Hubálková; Paola La Serna; Igor Linetskiy; Tat'jana Dostálová
Journal:  Int Dent J       Date:  2006-06       Impact factor: 2.512

9.  Correlation between magnetic resonance imaging disturbances and the magnetic susceptibility of dental materials.

Authors:  O Beuf; M Lissac; Y Crémillieux; A Briguet
Journal:  Dent Mater       Date:  1994-07       Impact factor: 5.304

10.  Tumor volume assessment by 18F-FDG PET/CT in patients with oral cavity cancer with dental artifacts on CT or MR images.

Authors:  Chung-Hwan Baek; Man Ki Chung; Young-Ik Son; Joon Young Choi; Hyung-Jin Kim; Yoo Jeong Yim; Young Hyeh Ko; Jeesun Choi; Jae Keun Cho; Han-Sin Jeong
Journal:  J Nucl Med       Date:  2008-08-14       Impact factor: 10.057

View more
  33 in total

1.  Induced magnetic moment in stainless steel components of orthodontic appliances in 1.5 T MRI scanners.

Authors:  Zhiyue J Wang; Nancy K Rollins; Hui Liang; Yong Jong Park
Journal:  Med Phys       Date:  2015-10       Impact factor: 4.071

2.  Initial clinical results of simultaneous 18F-FDG PET/MRI in comparison to 18F-FDG PET/CT in patients with head and neck cancer.

Authors:  K Kubiessa; S Purz; M Gawlitza; A Kühn; J Fuchs; K G Steinhoff; A Boehm; O Sabri; R Kluge; T Kahn; P Stumpp
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-11-29       Impact factor: 9.236

3.  Influence of pulse sequence parameters at 1.5 T and 3.0 T on MRI artefacts produced by metal-ceramic restorations.

Authors:  A R G Cortes; R Abdala-Junior; M Weber; E S Arita; J L Ackerman
Journal:  Dentomaxillofac Radiol       Date:  2015-06-18       Impact factor: 2.419

4.  Influence of the artefact reduction algorithm of Picasso Trio CBCT system on the diagnosis of vertical root fractures in teeth with metal posts.

Authors:  I S Q Bezerra; F S Neves; T V Vasconcelos; G M B Ambrosano; D Q Freitas
Journal:  Dentomaxillofac Radiol       Date:  2015-03-12       Impact factor: 2.419

5.  MRI with intraoral orthodontic appliance-a comparative in vitro and in vivo study of image artefacts at 1.5 T.

Authors:  C Zachriat; P Asbach; K I Blankenstein; I Peroz; F H Blankenstein
Journal:  Dentomaxillofac Radiol       Date:  2015-03-03       Impact factor: 2.419

6.  PETRA, MSVAT-SPACE and SEMAC sequences for metal artefact reduction in dental MR imaging.

Authors:  Tim Hilgenfeld; Marcel Prager; Alexander Heil; Franz Sebastian Schwindling; Mathias Nittka; David Grodzki; Peter Rammelsberg; Martin Bendszus; Sabine Heiland
Journal:  Eur Radiol       Date:  2017-07-11       Impact factor: 5.315

Review 7.  MRI in Dentistry- A Future Towards Radiation Free Imaging - Systematic Review.

Authors:  Lav Kumar Niraj; Basavaraj Patthi; Ashish Singla; Ritu Gupta; Irfan Ali; Kuldeep Dhama; Jishnu Krishna Kumar; Monika Prasad
Journal:  J Clin Diagn Res       Date:  2016-10-01

Review 8.  Should the orthodontic brackets always be removed prior to magnetic resonance imaging (MRI)?

Authors:  Arash Poorsattar-Bejeh Mir; Manouchehr Rahmati-Kamel
Journal:  J Oral Biol Craniofac Res       Date:  2015-09-15

9.  Assessment of impacted and partially impacted lower third molars with panoramic radiography compared to MRI-a proof of principle study.

Authors:  Barbara Kirnbauer; Norbert Jakse; Petra Rugani; Michael Schwaiger; Marton Magyar
Journal:  Dentomaxillofac Radiol       Date:  2018-02-13       Impact factor: 2.419

10.  Metal artefact reduction from dental hardware in carotid CT angiography using iterative reconstructions.

Authors:  Fabian Morsbach; Moritz Wurnig; Daniel M Kunz; Andreas Krauss; Bernhard Schmidt; Spyros S Kollias; Hatem Alkadhi
Journal:  Eur Radiol       Date:  2013-05-19       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.