| Literature DB >> 22384071 |
Thomas Klinke1, Amro Daboul, Juliane Maron, Tomasz Gredes, Ralf Puls, Ahmad Jaghsi, Reiner Biffar.
Abstract
BACKGROUND: Artifacts caused by dental restorations, such as dental crowns, dental fillings and orthodontic appliances, are a common problem in MRI and CT scans of the head and neck. The aim of this in-vitro study was to identify and evaluate the artifacts produced by different dental restoration materials in CT and MRI images.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22384071 PMCID: PMC3285178 DOI: 10.1371/journal.pone.0031766
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Figure 1Dimensions of (A) wedge-shaped specimens and (B) reference.
MRI and CT artifact categories.
| Category | 0 | 1 | 2 | 3 |
| MRI | No artifacts | Artifact less than 30.0 mm2 | Artifact 30.0 mm2 to 60.0 mm2 | Artifact more than 60.0 mm2 |
| CT | No artifacts | Streak artifact <15.0 mm | Streak artifact 15.0–30.0 mm | Streak artifact >30.0 mm |
MRI and CT artifacts produced by composites (Group 1).
| Material | CT artifact category | MRI T1 artifact category | MRI T1 artifact Shape change | MRI T2 artifact category | MRI T2 artifact Shape change |
| Adaptic® | 1 | 0 | 0 | ||
| X-flow™ | 1 | 0 | 0 | ||
| Quixfil™ | 1 | 0 | 0 | ||
| Core X™ | 2 | 0 | 0 | ||
| Dyract® Xtra | 2 | 0 | 0 | ||
| Ceram·X Duo | 2 | 0 | 0 | ||
| Esthet•X® | 3 | 0 | 0 | ||
| Dyract® flow | 2 | 0 | 0 | ||
| Spectrum®TPH® | 3 | 0 | 0 | ||
| X-tra fil | 2 | 0 | 0 | ||
| Grandio | 1 | 0 | 0 | ||
| Admira | 3 | 0 | 0 | ||
| Twinky Star | 2 | 0 | 0 | ||
| Arabesk® | 2 | 0 | 0 | ||
| Tetric EvoCeram® | 3 | 1 | −/+− | 1 | −/(+) |
| Tetric Evoflow® | 3 | 1 | −/+− | 1 | −/(+) |
| Adamant® | 3 | 1 | −/+− | 1 | −/(+) |
| InTen-S® | 2 | 1 | −/+− | 1 | −/(+) |
| Tetric® Flow | 3 | 1 | −/+− | 1 | −/+ |
| Tetric Ceram® | 3 | 1 | −/+− | 1 | −/(+) |
| Compoglass® F | 3 | 1 | −/+− | 1 | −/(+) |
| Heliomolar® | 2 | 1 | −/+− | 1 | −/(+) |
| Helio Progress® | 0 | 0 | 0 | ||
| Filtek™ | 1 | 0 | 0 | ||
| 3 M™ Z100™ MP | 1 | 0 | 0 | ||
| Nanosit™ | 3 | 0 | 0 | ||
| Synergy D6 | 1 | 0 | 0 | ||
| Solidex | 0 | 0 | 0 | ||
| Henry Schein® | 2 | 0 | 0 | ||
| Charisma® | 2 | 0 | 0 | ||
| Revolcin® Flow | 1 | 0 | 0 |
−/+ Test specimen not recognizable, adjacent reference specimen recognizable.
−/+− Test specimen not recognizable, adjacent reference specimen partially recognizable.
−/(+) Test specimen not recognizable, adjacent reference specimen recognizable but changed size or shape.
MRI and CT artifacts produced by temporary filling materials (Group 2).
| Material | CT artifact category | MRI T1 artifact category | MRI T1 artifact Shape change | MRI T2 artifact category | MRI T2 artifact Shape change |
| Clip | 0 | 0 | 0 | ||
| Guttapercha | 2 | 0 | 0 | ||
| Ketac™ | 2 | 0 | 0 | ||
| Cavit™ G | 2 | 1 | −/+− | 1 | −/+ |
−/+ test specimen not recognizable, adjacent reference specimen recognizable.
−/+− test specimen not recognizable, adjacent reference specimen partially recognizable.
MRI and CT artifacts produced by Cermics (Group 3).
| Material | CT artifact category | MRI T1 artifact category | MRI T1 artifact Shape change | MRI T2 artifact category | MRI T2 artifact Shape change |
| Zirconium dioxide | 3 | 0 | 0 | ||
| IPS Empress® | 1 | 0 | 0 | ||
| Vita Omega 900 | 2 | 0 | 0 | ||
| Cergo | 2 | 0 | 0 | ||
| Duceragold | 1 | 0 | 0 |
MRI and CT artifacts produced by Metal alloys (Group 4).
| Material | CT artifact category | MRI T1 artifact category | MRI T1 artifact Shape change | MRI T2 artifact category | MRI T2 artifact Shape change |
| Amalcap® Plus | 3 | 1 | −/+ | 1 | −/+ |
| Remanium® Star | 3 | 1 | −/+ | 1 | −/+ |
| Degunorm® | 3 | 1 | −/+ | 1 | −/+ |
| Kavo Everest® Titan | 3 | 1 | −/+− | 1 | −/+− |
−/+ test specimen not recognizable, adjacent reference specimen recognizable.
−/+− test specimen not recognizable, adjacent reference specimen partially recognizable.
Figure 2MRI artifact produced by Tetric EvoCeram in (A) T1 and (B) T2.
Figure 3MRI artifact produced by Remanium in (A) T1 and (B) T2.
Figure 4Streak artifact produced by Zirconium dioxide in CT.
Figure 5Streak artifact produced by Amalcap in CT.