MRI with intraoral orthodontic appliance-a comparative in vitro and in vivo study of image artefacts at 1.5 T.

C Zachriat1, P Asbach2, K I Blankenstein3, I Peroz1, F H Blankenstein1.   

Abstract

OBJECTIVES: We investigated artefacts caused from orthodontic appliances at 1.5-T MRI of the head and neck region and whether the image quality can be improved utilizing the artefact-minimizing sequence WARP.
METHODS: In vitro tests were performed by phantom measurements of different orthodontic devices applying different types of MR sequences [echoplanar imaging (EPI), turbo spin echo (TSE) and TSE-WARP, gradient echo (GRE)]. Two independent readers determined after calibration the level of artefacts. Subsequently, the interobserver agreement was calculated. The measurement of artefacts was based on the American Society for Testing Materials Standard F 2119-07. For in vivo imaging, one test person was scanned with an inserted multibracket appliance. The level of artefacts for 27 target regions was evaluated.
RESULTS: In vitro: ceramic brackets and ferromagnetic steel brackets produced artefact radii up to 1.12 and 7.40 cm, respectively. WARP reduced these artefacts by an average of 32.7%. The Bland-Altman-Plot indicated that maximum measurement differences of 3 mm have to be expected with two calibrated observers. In vivo: the EPI sequence for brain imaging was not analysable. The TSE sequence of the brain did not demonstrate artefacts except for the nasal cavity. Conversely, the TSE sequence of the cervical spine revealed severe artefacts in the midface region. The GRE sequence appeared to be more susceptible to artefacts than did the TSE sequence.
CONCLUSIONS: In vitro measurements allow an estimation of the in vivo artefact size. Orthodontic appliances may often remain intraorally when performing MRI. WARP showed a more significant effect in vitro than in vivo.

Entities:  

Keywords:  MRI; artefacts; brackets; orthodontic appliances; phantoms

Mesh:

Year:  2015        PMID: 25734243      PMCID: PMC4628401          DOI: 10.1259/dmfr.20140416

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  22 in total

1.  Coronary arterial stents: safety and artifacts during MR imaging.

Authors:  J Hug; E Nagel; A Bornstedt; B Schnackenburg; H Oswald; E Fleck
Journal:  Radiology       Date:  2000-09       Impact factor: 11.105

2.  Magnetic resonance imaging diagnosis of the temporomandibular joint in patients with orthodontic appliances.

Authors:  Yoshie Okano; Mitsuaki Yamashiro; Takashi Kaneda; Kazutaka Kasai
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2003-02

3.  A complete distortion correction for MR images: I. Gradient warp correction.

Authors:  Simon J Doran; Liz Charles-Edwards; Stefan A Reinsberg; Martin O Leach
Journal:  Phys Med Biol       Date:  2005-03-16       Impact factor: 3.609

4.  Minimizing artifacts caused by metallic implants at MR imaging: experimental and clinical studies.

Authors:  J S Suh; E K Jeong; K H Shin; J H Cho; J B Na; D H Kim; C D Han
Journal:  AJR Am J Roentgenol       Date:  1998-11       Impact factor: 3.959

5.  Predictability of magnetic susceptibility artifacts from metallic orthodontic appliances in magnetic resonance imaging.

Authors:  F Blankenstein; B T Truong; A Thomas; N Thieme; C Zachriat
Journal:  J Orofac Orthop       Date:  2014-11-26       Impact factor: 1.938

6.  Dental material artifacts on MR images.

Authors:  D B Hinshaw; B A Holshouser; H I Engstrom; A H Tjan; E L Christiansen; W F Catelli
Journal:  Radiology       Date:  1988-03       Impact factor: 11.105

7.  Effect of orthodontic brackets and different wires on radiofrequency heating and magnetic field interactions during 3-T MRI.

Authors:  S Görgülü; S Ayyildiz; K Kamburoglu; S Gökçe; T Ozen
Journal:  Dentomaxillofac Radiol       Date:  2013-11-20       Impact factor: 2.419

8.  A new carbide finishing bur for bracket debonding.

Authors:  R J Radlanski
Journal:  J Orofac Orthop       Date:  2001-07       Impact factor: 1.938

9.  New MR imaging methods for metallic implants in the knee: artifact correction and clinical impact.

Authors:  Christina A Chen; Weitian Chen; Stuart B Goodman; Brian A Hargreaves; Kevin M Koch; Wenmaio Lu; Anja C Brau; Christine E Draper; Scott L Delp; Garry E Gold
Journal:  J Magn Reson Imaging       Date:  2011-05       Impact factor: 4.813

10.  Debonding forces of three different customized bases of a lingual bracket system.

Authors:  Jang-Won Sung; Tae-Yub Kwon; Hee-Moon Kyung
Journal:  Korean J Orthod       Date:  2013-10-25       Impact factor: 1.372

View more
  7 in total

1.  [Metal objects of the head and neck region in magnetic resonance imaging : Survey among radiologists].

Authors:  Ulrike Kielburg; Felix H Blankenstein
Journal:  Radiologe       Date:  2019-10       Impact factor: 0.635

2.  Magnetic permeability as a predictor of the artefact size caused by orthodontic appliances at 1.5 T magnetic resonance imaging.

Authors:  Felix H Blankenstein; Patrick Asbach; Florian Beuer; Johannes Glienke; Stefan Mayer; Christine Zachriat
Journal:  Clin Oral Investig       Date:  2016-03-17       Impact factor: 3.573

3.  MRI compatibility of orthodontic brackets and wires: systematic review article.

Authors:  Adrienn Dobai; Fanni Dembrovszky; Tamás Vízkelety; Péter Barsi; Fanni Juhász; Csaba Dobó-Nagy
Journal:  BMC Oral Health       Date:  2022-07-19       Impact factor: 3.747

4.  Quantitative evaluation of artefact reduction from metallic dental materials in short tau inversion recovery imaging: efficacy of syngo WARP at 3.0 tesla.

Authors:  Lan Thi Xuan Tran; Junichiro Sakamoto; Ami Kuribayashi; Hiroshi Watanabe; Hiroshi Tomisato; Tohru Kurabayashi
Journal:  Dentomaxillofac Radiol       Date:  2019-07-05       Impact factor: 2.419

5.  Microleakage beneath orthodontic brackets in high field magnetic resonance imaging (MRI) AT 1.5 & 3 Tesla.

Authors:  Esra Bolat Gümüş; Samed Şatir; Alper Kuştarci
Journal:  Dentomaxillofac Radiol       Date:  2022-02-09       Impact factor: 3.525

Review 6.  Unwanted effects due to interactions between dental materials and magnetic resonance imaging: a review of the literature.

Authors:  Sherin Jose Chockattu; Deepak Byathnal Suryakant; Sophia Thakur
Journal:  Restor Dent Endod       Date:  2018-08-30

7.  Evaluation of magnetic resonance imaging artifacts caused by fixed orthodontic CAD/CAM retainers-an in vitro study.

Authors:  Christoph Roser; Tim Hilgenfeld; Sinan Sen; Tobias Badrow; Sebastian Zingler; Sabine Heiland; Martin Bendszus; Christopher J Lux; Alexander Juerchott
Journal:  Clin Oral Investig       Date:  2020-08-12       Impact factor: 3.573

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.