Literature DB >> 11927420

A comparative analysis of five methods of determining corneal refractive power in eyes that have undergone myopic laser in situ keratomileusis.

Abdelmonem M Hamed1, Li Wang, Manjula Misra, Douglas D Koch.   

Abstract

OBJECTIVE: To evaluate the accuracy of computerized videokeratography, keratometry, and the Gaussian optics formula for measuring corneal refractive power in patients after myopic laser in situ keratomileusis (LASIK).
DESIGN: Noncomparative case series. PARTICIPANTS: One hundred eyes of 63 patients (mean age, 45.0 +/- 10.9 [standard deviation] years) who underwent LASIK were included in the study.
METHODS: Using the clinical history method as the standard, we evaluated the accuracy of values of corneal refractive power derived from computerized videokeratography (the EffRP value of the EyeSys Corneal Analysis System, which averages corneal refractive power over the central 3 mm), keratometry (K), the Gaussian optics formula (GauRP), and values of EffRP and keratometry as modified according to the amount of LASIK-induced refractive change. MAIN OUTCOME MEASURES: Correlation of measured corneal power values to those obtained using clinical history method (HisRP).
RESULTS: Although the values for HisRP were significantly correlated with postoperative EffRP and K values and with GauRP, postoperative EffRP and K values were higher than HisRP (0.87 +/- 0.68 diopters [D] and 1.16 +/- 1.10 D, respectively), and GauRP were lower than HisRP (0.44 +/- 0.66 D) (P < 0.001 for all three comparisons). The differences between HisRP and both postoperative EffRP and K values increased significantly with the amount of myopic correction. The most accurate results were obtained by modifying the postoperative values of EffRP according to the amount of LASIK-induced refractive change; 70% of these values were within +/- 0.5 D and 94% within +/- 1 D of HisRP values.
CONCLUSIONS: Using the clinical history method as the standard, we found that the most accurate method for determining corneal refractive power in post-LASIK eyes was to adjust the postoperative corneal measurement according to the amount of LASIK-induced refractive change.

Entities:  

Mesh:

Year:  2002        PMID: 11927420     DOI: 10.1016/s0161-6420(01)01001-6

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  22 in total

1.  Comparison of intraocular lens power calculation methods in eyes that have undergone laser-assisted in-situ keratomileusis.

Authors:  Li Wang; Marc A Booth; Douglas D Koch
Journal:  Trans Am Ophthalmol Soc       Date:  2004

2.  Correcting the corneal power measurements after myopic laser in situ keratomileusis.

Authors:  Lei Liu; Jieyu Dong; Xinyu Li
Journal:  J Huazhong Univ Sci Technolog Med Sci       Date:  2006

3.  Orbscan II-assisted intraocular lens power calculation for cataract surgery following myopic laser in situ keratomileusis (an American Ophthalmological Society thesis).

Authors:  Henry Gelender
Journal:  Trans Am Ophthalmol Soc       Date:  2006

4.  Predictability of Sirius dual-scanning corneal tomography in the measurement of corneal power after photorefractive surgery.

Authors:  Sameh M Fouda; Haitham Y Al-Nashar; Basem M Ibrahim; Ashraf Bor'i
Journal:  Int Ophthalmol       Date:  2015-05-19       Impact factor: 2.031

5.  Comparison of the accuracy of intraocular lens power calculations for cataract surgery in eyes after phototherapeutic keratectomy.

Authors:  Yukari Yaguchi; Kazuno Negishi; Megumi Saiki; Hidemasa Torii; Kazuo Tsubota
Journal:  Jpn J Ophthalmol       Date:  2016-06-08       Impact factor: 2.447

6.  Corneal power measurement with Fourier-domain optical coherence tomography.

Authors:  Maolong Tang; Alex Chen; Yan Li; David Huang
Journal:  J Cataract Refract Surg       Date:  2010-12       Impact factor: 3.351

7.  Comparison of the accuracy of intraocular lens power calculation formulas for eyes after corneal refractive surgery.

Authors:  Xiao-Zhen Wang; Rui Cui; Xu-Dong Song; Bo Yun; Jin Qian; Ning Ding
Journal:  Ann Transl Med       Date:  2020-07

8.  Intraocular lens power calculation after corneal refractive surgery.

Authors:  Vahid Feiz
Journal:  Middle East Afr J Ophthalmol       Date:  2010-01

9.  Comparison of Newer Intraocular Lens Power Calculation Methods for Eyes after Corneal Refractive Surgery.

Authors:  Li Wang; Maolong Tang; David Huang; Mitchell P Weikert; Douglas D Koch
Journal:  Ophthalmology       Date:  2015-10-14       Impact factor: 12.079

10.  Scheimpflug analysis of corneal power changes after hyperopic small incision lenticule extraction.

Authors:  Dan Fu; Jianmin Shang; Xiaoyu Zhang; Lingling Niu; Tian Han; Xingtao Zhou
Journal:  BMC Ophthalmol       Date:  2021-07-20       Impact factor: 2.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.