Optical coherence tomography accurately measures corneal power change from laser refractive surgery.

Ryan P McNabb1, Sina Farsiu2, Sandra S Stinnett3, Joseph A Izatt2, Anthony N Kuo4.   

Abstract

PURPOSE: To determine the ability of motion-corrected optical coherence tomography (OCT) to measure the corneal refractive power change due to LASIK.
DESIGN: Evaluation of a diagnostic test or technology in a cohort.
SUBJECTS: A total of 70 eyes from 37 subjects undergoing LASIK were measured preoperatively. A total of 39 eyes from 22 subjects were measured postoperatively and completed the study.
METHODS: Consecutive patients undergoing LASIK at the Duke Eye Center who consented to participate were imaged with Placido-ring topography, Scheimpflug photography, and OCT on the day of their surgery. Patients were then reimaged with the same imaging systems at the postoperative month 3 visit. Change in preoperative to postoperative corneal refractive power as measured by each of the imaging modalities was compared with the preoperative to postoperative change in manifest refraction (MRx) using the t test with generalized estimating equations. MAIN OUTCOME MEASURES: Corneal refractive power change due to LASIK as measured by Placido-ring topography, Scheimpflug photography, and OCT compared with the MRx change vertexed to the corneal plane. The change in MRx should correspond to the change in the corneal refractive power from LASIK and was considered the reference measurement.
RESULTS: In 22 individuals (39 eyes) returning after LASIK, we found no significant difference between the clinically measured pre- to post-LASIK change in MRx and both Scheimpflug photography (P = 0.714) and OCT (P = 0.216). In contrast, keratometry values from Placido-ring topography were found to be significantly different from the measured refractive change (P < 0.001). In addition, of the 3 imaging modalities, OCT recorded the smallest mean absolute difference from the reference measurement with the least amount of variability.
CONCLUSIONS: Motion-corrected OCT more accurately measures the change in corneal refractive power due to laser refractive surgery than other currently available clinical devices. By offering accurate corneal refractive power measurements in normal and surgically modified subjects, OCT offers a compelling alternative to current clinical devices for determining corneal refractive power.
Copyright © 2015 American Academy of Ophthalmology. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2014        PMID: 25487424      PMCID: PMC4372493          DOI: 10.1016/j.ophtha.2014.10.003

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  36 in total

1.  Simultaneous swept source optical coherence tomography of the anterior segment and retina using coherence revival.

Authors:  Al-Hafeez Dhalla; Derek Nankivil; Theresa Bustamante; Anthony Kuo; Joseph A Izatt
Journal:  Opt Lett       Date:  2012-06-01       Impact factor: 3.776

2.  In vivo human retinal imaging by Fourier domain optical coherence tomography.

Authors:  Maciej Wojtkowski; Rainer Leitgeb; Andrzej Kowalczyk; Tomasz Bajraszewski; Adolf F Fercher
Journal:  J Biomed Opt       Date:  2002-07       Impact factor: 3.170

3.  Intraocular lens calculations after refractive surgery.

Authors:  Robert A Latkany; Amit R Chokshi; Mark G Speaker; Jodi Abramson; Barrie D Soloway; Guopei Yu
Journal:  J Cataract Refract Surg       Date:  2005-03       Impact factor: 3.351

4.  Anterior segment biometry with the Pentacam: comprehensive assessment of repeatability of automated measurements.

Authors:  Hema Shankar; Deepa Taranath; Chandramalar T Santhirathelagan; Konrad Pesudovs
Journal:  J Cataract Refract Surg       Date:  2008-01       Impact factor: 3.351

5.  Imaging the ocular anterior segment with real-time, full-range Fourier-domain optical coherence tomography.

Authors:  Marinko V Sarunic; Sanjay Asrani; Joseph A Izatt
Journal:  Arch Ophthalmol       Date:  2008-04

6.  Reproducibility of normal corneal power measurements with a keratometer, photokeratoscope, and video imaging system.

Authors:  S B Hannush; S L Crawford; G O Waring; M C Gemmill; M J Lynn; A Nizam
Journal:  Arch Ophthalmol       Date:  1990-04

7.  On the calculation of power from curvature of the cornea.

Authors:  T Olsen
Journal:  Br J Ophthalmol       Date:  1986-02       Impact factor: 4.638

8.  Nomogram-based intraocular lens power adjustment after myopic photorefractive keratectomy and LASIK: a new approach.

Authors:  Vahid Feiz; Majid Moshirfar; Mark J Mannis; Charles D Reilly; Francisco Garcia-Ferrer; Jeffrey J Caspar; Michele C Lim
Journal:  Ophthalmology       Date:  2005-08       Impact factor: 12.079

9.  Comparison of intraocular lens power calculation methods in eyes that have undergone LASIK.

Authors:  Li Wang; Marc A Booth; Douglas D Koch
Journal:  Ophthalmology       Date:  2004-10       Impact factor: 12.079

10.  Distributed scanning volumetric SDOCT for motion corrected corneal biometry.

Authors:  Ryan P McNabb; Francesco Larocca; Sina Farsiu; Anthony N Kuo; Joseph A Izatt
Journal:  Biomed Opt Express       Date:  2012-08-10       Impact factor: 3.732

View more
  13 in total

1.  Wide-field whole eye OCT system with demonstration of quantitative retinal curvature estimation.

Authors:  Ryan P McNabb; James Polans; Brenton Keller; Moseph Jackson-Atogi; Charlene L James; Robin R Vann; Joseph A Izatt; Anthony N Kuo
Journal:  Biomed Opt Express       Date:  2018-12-21       Impact factor: 3.732

2.  Golden angle based scanning for robust corneal topography with OCT.

Authors:  Joerg Wagner; David Goldblum; Philippe C Cattin
Journal:  Biomed Opt Express       Date:  2017-01-03       Impact factor: 3.732

3.  Review of intraoperative optical coherence tomography: technology and applications [Invited].

Authors:  Oscar M Carrasco-Zevallos; Christian Viehland; Brenton Keller; Mark Draelos; Anthony N Kuo; Cynthia A Toth; Joseph A Izatt
Journal:  Biomed Opt Express       Date:  2017-02-21       Impact factor: 3.732

4.  Segmentation guided registration of wide field-of-view retinal optical coherence tomography volumes.

Authors:  José Lezama; Dibyendu Mukherjee; Ryan P McNabb; Guillermo Sapiro; Anthony N Kuo; Sina Farsiu
Journal:  Biomed Opt Express       Date:  2016-11-01       Impact factor: 3.732

5.  Comprehensive review of surgical microscopes: technology development and medical applications.

Authors:  Ling Ma; Baowei Fei
Journal:  J Biomed Opt       Date:  2021-01       Impact factor: 3.170

6.  Effects of posterior corneal astigmatism on the accuracy of AcrySof toric intraocular lens astigmatism correction.

Authors:  Bin Zhang; Jing-Xue Ma; Dan-Yan Liu; Cong-Rong Guo; Ying-Hua Du; Xiu-Jin Guo; Yue-Xian Cui
Journal:  Int J Ophthalmol       Date:  2016-09-18       Impact factor: 1.779

7.  Novel microscope-integrated stereoscopic heads-up display for intrasurgical optical coherence tomography.

Authors:  Liangbo Shen; Oscar Carrasco-Zevallos; Brenton Keller; Christian Viehland; Gar Waterman; Paul S Hahn; Anthony N Kuo; Cynthia A Toth; Joseph A Izatt
Journal:  Biomed Opt Express       Date:  2016-04-06       Impact factor: 3.732

8.  Scheimpflug analysis of corneal power changes after hyperopic small incision lenticule extraction.

Authors:  Dan Fu; Jianmin Shang; Xiaoyu Zhang; Lingling Niu; Tian Han; Xingtao Zhou
Journal:  BMC Ophthalmol       Date:  2021-07-20       Impact factor: 2.209

9.  Accuracy of OCT-derived net corneal astigmatism measurement.

Authors:  Clara Llorens-Quintana; Elias Pavlatos; Omkar Thaware; Seema Gupta; Daniel Gradin; Denzil Romfh; Yan Li; David Huang
Journal:  J Cataract Refract Surg       Date:  2022-03-01       Impact factor: 3.528

10.  Longitudinal Evaluation of Cornea With Swept-Source Optical Coherence Tomography and Scheimpflug Imaging Before and After Lasik.

Authors:  Tommy C Y Chan; Sayantan Biswas; Marco Yu; Vishal Jhanji
Journal:  Medicine (Baltimore)       Date:  2015-07       Impact factor: 1.889

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.