Paweł Stefańczyk1, Dorota Nowosielecka1, Anna Polewczyk2,3, Łukasz Tułecki4, Konrad Tomków4, Wojciech Jacheć5, Ewa Lewicka6, Andrzej Tomaszewski7, Andrzej Kutarski7.
Abstract
BACKGROUND: Transvenous lead extraction (TLE) in patients with implantable cardioverter defibrillator (ICD) and cardiac resynchronization therapy (CRT) devices is considered as more risky. The aim of this study was to assess the safety and effectiveness of TLE in patients with infected CRT systems.Entities:
Keywords: cardiac resynchronization therapy; infectious indications; safety and effectiveness; transvenous lead extraction
Mesh:
Substances:
Year: 2022 PMID: 35627340 PMCID: PMC9141355 DOI: 10.3390/ijerph19105803
Source DB: PubMed Journal: Int J Environ Res Public Health ISSN: 1660-4601 Impact factor: 4.614
Figure 1Annual number of transvenous lead extraction procedures procedures, taking into account the type of devices.
Indications for initial implantation of CIED.
| Number of Patients | SSS | II–III Degree of a-v Block | HF | HF with LBBB | |
|---|---|---|---|---|---|
| AAI | 54 | 54 | 0 | 0 | 0 |
| VVI | 129 | 29 | 100 | 0 | 0 |
| DDD | 556 | 236 | 320 | 0 | 0 |
| VDD | 17 | 0 | 17 | 0 | 0 |
| CRT-P | 38 | 0 | 6 | 0 | 32 |
| ICDVR | 122 | 0 | 20 | 102 | 0 |
| ICDDR | 110 | 0 | 0 | 110 | 0 |
| CRT-D | 112 | 0 | 0 | 12 | 100 |
HF—heart failure, LBBB—left bundle branch block, SSS—sick sinus syndrome.
Clinical characteristics of patients with different types of CIED system.
| Pacemakers All (AAI, VVI, DDD, VDD) | ICDs All (VVI, DDD) | CRT-P | CRT-D | All CRT Systems (CRT-P Plus CRT-D) | |
|---|---|---|---|---|---|
| Group 1 | Group 2 | Group 3 | Group 4 | Group 5 | |
| Mean ± sd | Mean ± sd | Mean ± sd | Mean ± sd | Mean ± sd | |
| Chi2 test, | 1 vs. 2 | 1 vs. 3 | 1 vs. 4 | 1 vs. 5 | |
| Patient’s age during TLE | 69.96 ± 14.17 | 64.26 ± 12.63 | 70.26 ± 9.57 | 67.42 ± 10.23 | 68.14 ± 10.11 |
| Patient’s age during first system implantation | 61.10 ± 15.82 | 59.62 ± 12.87 | 62.92 ± 9.966 | 62.42 ± 10.55 | 62.55 ± 10.37 |
| Sex (% of female patients) | 272 (35.98) | 32 (13.79) | 10 (26.32) | 18 (16.07) | 28 (18.67) |
| Etiology other than IHD | 403 (53.31) | 152 (65.51) | 18 (47.37) | 47 (41.96) | 65 (43.33) |
| NYHA III or IV class | 64 (8.466) | 65 (28.02) | 15 (39.47) | 50 (44.64) | 65 (43.33) |
| LVEF (%) | 54.69 (10.25) | 37.19 (14.19) | 36.21 (13.75) | 30.77 (12.37) | 32.15 (12.91) |
| Renal failure moderate(creatinine >1.3—≤2.2 mg%) | 136 (17.10) | 61 (26.29) | 9 (23.68) | 29 (25.89) | 38 (25.33) |
| Renal failure severe or hemodialysis(creatinine ≥ 2.3 mg%) | 42 (5.556) | 22 (9.48) | 4 (10.54) | 10 (8.93) | 14 (9.33) |
| Renal failure (all),creatinine ≥ 1.3 mg% | 178 (22.66) | 83 (35.77) | 13 (34.22) | 39 (34.82) | 52 (34.66) |
| Diabetes | 158 (20.90) | 68 (29.31) | 13 (34.22) | 35 (31.25) | 48 (32.00) |
| Carlson’s index (points) | 4.86 ± 3.559 | 5.78±4.01 | 6.40 ± 4.175 | 6.01 ± 3.85 | 6.11 ± 3.92 |
| TLE indications—more exact division of infective indications | |||||
| Lead related infective endocarditis certain (with pocket infection or without) | 375 (49. 60) | 124 (53.45) | 23 (60.53) | 61 (54.46) | 84 (56.00) |
| Lead related infective endocarditis probable (with pocket infection or without) | 144 (19.05) | 38 (16.38) | 7 (18.42) | 24 (21.42) | 31 (20.67) |
| Local (isolated) pocket infection | 237 (31.35) | 70 (30.17) | 8 (21.05) | 27 (24.11) | 35 (23.33) |
TLE—transvenous lead extraction, CIED—cardiac implantable electric devices, AAI—pacemaker with one atrial lead, VVI—pacemaker with one ventricular lead, DDD—dual chamber pacemaker, VDD—pacemaker with one ventricular lead, ICD—implantable cardioverter defibrillator, CRTP—cardiac resynchronization therapy pacemaker, CRTD—cardiac resynchronization therapy defibrillator, N—number, sd—standard deviation, IHD—ischemic heart disease, NYHA—New York Heart Association functional class, LVEF—left ventricle ejection fraction.
System and history of pacing, TLE procedure and potential risk factors for major TLE complications and technical problems.
| Pacemakers All (AAI, VVI, DDD, VDD) | ICDs All (VVI, DDD) | CRT-P | CRT-D | All CRT Systems (CRT-P Plus CRT-D) | |
|---|---|---|---|---|---|
| Group 1 | Group 2 | Group 3 | Group 4 | Group 5 | |
| Mean ± sd | Mean ± sd | Mean ± sd | Mean ± sd | Mean ± sd | |
| Chi2 test,“U” Mann–Whitney test | 1 vs. 2 | 1 vs. 3 | 1 vs. 42 vs. 4 | 1 vs. 52 vs. 5 | |
| System and history of pacing | |||||
| Presence of abandoned lead before TLE | 144 (19.05) | 20 (8.26) | 3 (790) | 21 (18.75) | 24 (16.00) |
| Number of leads in the heart before TLE | 2.04 ± 0.69 | 1.61 ± 0.65 | 2.92 ± 0.63 | 3.13 ± 0.64 | 3.08 ± 0.64 |
| 4 and > 4 in the heart before TLE | 34 (4.50) | 4 (1.72) | 4 (10.53) | 22 (19.64) | 26 (17.33) |
| Number of procedures before lead extraction | 2.21 ± 1.25 | 1.74 ± 1.00 | 2.22 ± 1.27 | 2.39 ± 1.50 | 2.34 ± 1.44 |
| Time since last CIED procedure (any) (months) | 38.72 ± 35.19 | 29.81 ± 21.99 | 31.28 ± 25.27 | 20.70 ± 18.93 | 23.46 ± 21.39 |
| Potential risk factors for major TLE complications and procedure complexity | |||||
| Number of extracted leads in one patient | 1.96 ± 0.64 | 1.57 ± 0.60 | 2.76 ± 0.71 | 3.06 ± 0.76 | 2.99 ± 0.76 |
| Three or more leads were extracted | 85 (11.25) | 9 (3.88) | 28 (73.68) | 96 (85.72) | 124 (82.67) |
| Utilized approach other than lead venous entry | 43 (5.69) | 3 (1.24) | 2 (5.26) | 2 (1.79) | 4 (2.67) |
| Extraction of abandoned lead(s) (any) | 137 (18.12) | 19 (8.19) | 2 (5.26) | 21 (18.75) | 23 (15.33) |
| Oldest extracted lead body dwell time in the patient | 106.1 ± 72.73 | 55.73 ± 45.96 | 87.16 ± 50.18 | 61.53 ± 42.43 | 68.02 ± 45.73 |
| Average extracted lead dwell time in the patient (months) | 97.42 ± 63.86 | 51.45 ± 36.82 | 68.64 ± 35.96 | 50.17 ± 31.01 | 54.85 ± 33.20 |
| Average lead duration in analyzed group (months) | 98.93 ± 69.71 | 55.26 ± 43.30 | 68.12 ± 46.87 | 51.48 ± 38.85 | 55.43 ± 41.45 |
| Cumulative dwell time of extracted lead (in years) in the patient | 16.25 ± 13.13 | 7.265 ± 6.776 | 16.13 ± 10.13 | 13.29 ± 9.472 | 14.01 ± 9.688 |
| Number of patients with extracted lead(s) with passive fixation | 556 (73.54) | 78 (33.62) | 34 (89.47) | 106 (94.64) | 140 (93.33) |
| PADIT score [points] | 4.723 ± 2.744 | 6.135 ± 2.208 | 4.816 ± 1.625 | 8.836 ± 1.738 | 7.581 ± 2.359 |
| SAFeTY-TLE calculator of risk of MC TLE [points] | 7.29 ± 4.47 | 3.95 ± 3.20 | 7.13 ± 4.34 | 6.03 ± 3.97 | 6.31 ± 4.08 |
| SAFeTY-TLE calculator of risk of MC TLE [%] | 2 49 ± 4.13 | 0.88 ± 1.86 | 2.50 ± 4.87 | 1.58 ± 2.18 | 1.82 ± 3.11 |
TLE—transvenous lead extraction, CIED—cardiac implantable electric devices, AAI—pacemaker with one atrial lead, VVI—pacemaker with one ventricular lead, DDD—dual chamber pacemaker, VDD—pacemaker with one ventricular lead, ICD—implantable cardioverter defibrillator, CRTP—cardiac resynchronization therapy pacemaker, CRTD—cardiac resynchronization therapy defibrillator, MC—major, N—number, sd—standard deviation, MC—major complications.
TLE complexity in compared groups of patients with different CIED systems.
| Pacemakers All (AAI, VVI, DDD, VDD) | ICDs All (VVI, DDD) | CRT-P | CRT-D | All CRT Systems (CRT-P Plus CRT-D) | |
|---|---|---|---|---|---|
| Mean ± sd/N (%) | Mean ± sd/N (%) | Mean ± sd/N (%) | Mean ± sd/N (%) | Mean ± sd/N (%) | |
| Chi2 test, | 1 vs. 2 | 1 vs. 3 | 1 vs. 4 | 1 vs. 5 | |
| TLE complexity | |||||
| Procedure duration (skin to skin) [minutes] | 48.45 ± 25.51 | 42.13 ± 20.94 | 49.37 ± 17.82 | 52.65 ± 23.39 | 51.82 ± 22.10 |
| Procedure duration (sheath to sheath) [minutes] | 16.59 ± 24.04 | 10.79 ± 19.09 | 17.47 ± 16.90 | 22.04 ± 23.16 | 20.88 ± 21.78 |
| Average time of single lead extraction [minutes] | 8.11 ± 9.95 | 6.41 ± 8.93 | 6.90 ± 7.34 | 6.99 ± 6.78 | 6.97 ± 6.90 |
| Technical problems during TLE (any) | 144 (19.05) | 20 (8.62) | 12 (31.58) | 17 (15.18) | 29 (19.33) |
| Necessity to change venous approach | 53 (7.02) | 4 (1.72) | 2 (5.26) | 3 (2.68) | 5 (3.33) |
| Mutual lead to lead connection with strong scar | 50 (6.61) | 5 (2.16) | 4 (10.53) | 9 (8.04) | 13 (8.67) |
| Break of extracted lead | 58 (7.67) | 3 (1.29) | 5 (13.16) | 3 (2.68) | 8 (5.33) |
| Byrd dilator collapse/detorsion | 20 (2.65) | 5 (2.16) | 1 (2.63) | 4 (3.57) | 5 (3.33) |
| Block in venous lead entry region | 38 (5.03) | 9 (3.88) | 3 (7.90) | 9 (8.04) | 12 (8.00) |
| Two or more technical problems | 21 (2.78) | 3 (1.29) | 2 (5.26) | 5 (4.46) | 7 (4.67) |
| Utility of additional tools | |||||
| Evolution (old and new) or TighRail | 6 (0.79) | 2 (0.86) | 2 (5.26) | 2 (1.79) | 4 (2.67) |
| Metal sheath | 37 (4.89) | 9 (3.88) | 3 (7.89) | 9 (8.04) | 12 (8.00) |
| Lasso catheter/snare | 25 (3.31) | 1 (0.43) | 1 (2.63) | 3 (2.68) | 4 (2.67) |
| Basket catheter | 15 (1.98) | 1 (0.43) | 0 (0.00) | 0 (0.00) | 0 (0.00) |
| Temporary pacing during procedure | 204 (26.99) | 19 (8.19) | 12 (31.58) | 37 (33.04) | 49 (32.67) |
TLE—transvenous lead extraction, CIED—cardiac implantable electric devices, AAI—pacemaker with one atrial lead, VVI—pacemaker with one ventricular lead, DDD—dual chamber pacemaker, VDD—pacemaker with one ventricular lead, ICD—implantable cardioverter defibrillator, CRTP—cardiac resynchronization therapy pacemaker, CRTD—cardiac resynchronization therapy defibrillator, N—number, sd—standard deviation.
TLE efficacy and complications in compared groups of patients with different CIED systems.
| Pacemakers All (AAI, VVI, DDD, VDD) | ICDs All (VVI, DDD) | CRT-P | CRT-D | All CRT Systems (CRT-P Plus CRT-D) | |
|---|---|---|---|---|---|
| Group 1 | Group 2 | Group 3 | Group 4 | Group 5 | |
| N (%) | N (%) | N (%) | N (%) | N (%) | |
| Chi2 test, | 1 vs. 2 | 1 vs. 3 | 1 vs. 4 | 1 vs. 5 | |
| TLE efficacy and complications | |||||
| Major complications (any) | 20 (2.65) | 1 (0.43) | 0 (0.00) | 0 (0.00) | 0 (0.00) |
| Hemopericardium | 10 (1.32) | 1 (0.43) | 0 (0.00) | 0 (0.00) | 0 (0.00) |
| Hemothorax | 1 (0.13) | 0 (0.00) | 0 (0.00) | 0 (0.00) | 0 (0.00) |
| Tricuspid valve damage during TLE (severe) | 6 (0.79) | 0 (0.00) | 0 (0.00) | 0 (0.00) | 0 (0.00) |
| Rescue cardiac surgery | 9 (1.19) | 1 (0.43) | 0 (0.00) | 0 (0.00) | 0 (0.00) |
| Death procedure-related (intra-, post-procedural) | 1 (0.13) | 0 (0.00) | 0 (0.00) | 0 (0.00) | 0 (0.00) |
| Death indication-related (intra-, post-procedural) | 4 (0.53) | 0 (0.00) | 0 (0.00) | 0 (0.00) | 0 (0.00) |
| Partial radiological success (remaining tip or <4 cm lead fragment) | 40 (5.29) | 3 (1.29) | 3 (7.90) | 1 (0.89) | 4 (2.67) |
| Full clinical success | 710 (93.92) | 228 (98.28) | 34 (89.47) | 111 (99.11) | 145 (96.67) |
| Full procedural success | 706 (93.39) | 228 (98.28) | 34 (89.47) | 111 (99.11) | 145 (96.67) |
TLE—transvenous lead extraction, CIED—cardiac implantable electric devices, AAI—pacemaker with one atrial lead, VVI—pacemaker with one ventricular lead, DDD—dual chamber pacemaker, VDD—pacemaker with one ventricular lead, ICD—implantable cardioverter defibrillator, CRTP—cardiac resynchronization therapy pacemaker, CRTD—cardiac resynchronization therapy defibrillator, N—number, MN—methodically noncomparable.
Prognosis in short-, mean- and long-term follow-up in compared groups of patients.
| Pacemakers All (AAI, VVI, DDD, VDD) | ICDs All (VVI, DDD) | CRT-P | CRT-D | All CRT Systems (CRT-P Plus CRT-D) | |
|---|---|---|---|---|---|
| Group 1 | Group 2 | Group 3 | Group 4 | Group 5 | |
| N (%) | N (%) | N (%) | N (%) | N (%) | |
| Chi2 test, | 1 vs. 2 | 1 vs. 3 | 1 vs. 42 vs. 4 | 1 vs. 52 vs. 5 | |
| Prognosis in short-, mean- and long-term follow-up | |||||
| Alive during | 409 (54.10) | 110 (47.41) | 13 (34.21) | 41 (36.61) | 54 (36.00) |
| 48 h mortality | 6 (0.79) | 1 (0.43) | 0 (0.00) | 0 (0.00) | 0 (0.00) |
| 1 month mortality after TLE; 2–30 days n (% of patients with follow-up longer than 2 days) | 19/750 | 5/231 (2.17) | 2/38 (5.26) | 5/112 (4.46) | 7/150 (4.67) |
| 1 year mortality after TLE (31–365 days); n (% of patients with follow-up longer than 30 days) | 63/722 (8.73) | 31/222 (13.96) | 8/36 (22.22) | 24/105 (22.86) | 32/141 (22.70) |
| 3 year mortality after TLE (366–1095 days); n (% of patients with follow-up longer than 365 days) | 70/644 | 33/180 (18.33) | 8/28 (28.57) | 24/76 (31.58) | 32/104 (30.77) |
| Death late > 3 years after TLE (after 1095 days); n (% of patients with follow-up longer than 1095 days) | 189/530 | 52/127 (40.94) | 7/18 (38.89) | 18/47 (38.30) | 25/65 (38.46) |
| All deaths | 347 (45.90) | 122 (52.59) | 25 (65.79) | 71 (63.39) | 96 (64.00) |
TLE—transvenous lead extraction, CIED—cardiac implantable electric devices, AAI—pacemaker with one atrial lead, VVI—pacemaker with one ventricular lead, DDD—dual chamber pacemaker, VDD—pacemaker with one ventricular lead, ICD—implantable cardioverter defibrillator, CRTP—cardiac resynchronization therapy pacemaker, CRTD—cardiac resynchronization therapy defibrillator, N—number.
Figure 2Kaplan–Meier survival curves of patients with the different types of infected CIED.