Literature DB >> 10334441

Pacemaker lead extraction with the laser sheath: results of the pacing lead extraction with the excimer sheath (PLEXES) trial.

B L Wilkoff1, C L Byrd, C J Love, D L Hayes, T D Sellers, R Schaerf, V Parsonnet, L M Epstein, R A Sorrentino, C Reiser.   

Abstract

OBJECTIVES: The purpose of this study was to evaluate the safety and effectiveness of pacemaker lead extraction with the excimer sheath in comparison to nonlaser lead extraction.
BACKGROUND: Fibrotic attachments that develop between chronically implanted pacemaker leads and to the venous, valvular and cardiac structures are the major obstacles to safe and consistent lead extraction. Locking stylets and telescoping sheaths produce a technically demanding but effective technique of mechanically disrupting the fibrosis. However, ultraviolet excimer laser light dissolves instead of tearing the tissue attachments.
METHODS: A randomized trial of lead extraction was conducted in 301 patients with 465 chronically implanted pacemaker leads. The laser group patients had the leads removed with identical tools as the nonlaser group with the exception that the inner telescoping sheath was replaced with the 12-F excimer laser sheath. Success for both groups was defined as complete lead removal with the randomized therapy without complications.
RESULTS: Complete lead removal rate was 94% in the laser group and 64% in the nonlaser group (p = 0.001). Failed nonlaser extraction was completed with the laser tools 88% of the time. The mean time to achieve a successful lead extraction was significantly reduced for patients randomized to the laser tools, 10.1 +/- 11.5 min compared with 12.9 +/- 19.2 min for patients randomized to nonlaser techniques (p < 0.04). Potentially life-threatening complications occurred in none of the nonlaser and three of the laser patients, including one death (p = NS).
CONCLUSIONS: Laser-assisted pacemaker lead extraction has significant clinical advantages over extraction without laser tools and is associated with significant risks.

Entities:  

Mesh:

Year:  1999        PMID: 10334441     DOI: 10.1016/s0735-1097(99)00074-1

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  72 in total

1.  Exchange of pacing or defibrillator leads following laser sheath extraction of non-functional leads in patients with ipsilateral obstructed venous access.

Authors:  F A Bracke; L M van Gelder; N Sreeram; A Meijer
Journal:  Heart       Date:  2000-06       Impact factor: 5.994

Review 2.  Pacemaker lead complications: when is extraction appropriate and what can we learn from published data?

Authors:  F A Bracke; A Meijer; L M van Gelder
Journal:  Heart       Date:  2001-03       Impact factor: 5.994

3.  Cardiac lead extraction with a novel locking stylet.

Authors:  C Kennergren; R H Schaerf; T D Sellers; B L Wilkoff; C L Byrd; G F Tyres; S Coe; C W Coates; C Reiser
Journal:  J Interv Card Electrophysiol       Date:  2000-12       Impact factor: 1.900

4.  Determinants of procedural outcome of chronically implanted pacemaker and defibrillator leads using the Excimer laser sheath.

Authors:  C A Rinaldi; J Bostock; N Patel; C A Bucknall
Journal:  Heart       Date:  2002-02       Impact factor: 5.994

5.  Replacement of extracted permanent pacemaker or defibrillator leads by cannulation of veins using the femoral "drag through" technique.

Authors:  R J Schilling
Journal:  Heart       Date:  2002-03       Impact factor: 5.994

Review 6.  Advances in device lead extraction.

Authors:  C L Byrd
Journal:  Curr Cardiol Rep       Date:  2001-07       Impact factor: 2.931

7.  Blunt atrial transseptal puncture using excimer laser in swine.

Authors:  Abdalla A Elagha; Ann H Kim; Ozgur Kocaturk; Robert J Lederman
Journal:  Catheter Cardiovasc Interv       Date:  2007-10-01       Impact factor: 2.692

Review 8.  [ICD lead defects: diagnosis and therapeutical options].

Authors:  Christian G Wollmann; Dirk Böcker; Andreas Löher; Hans H Scheld; Günter Breithardt; Rainer Gradaus
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2009-02-11

9.  Re-evaluation of transvenous lead extraction with modified standard technique: a prospective study in 229 patients.

Authors:  Xian-Ming Chu; Xue-Bin Li; Ping Zhang; Long Wang; Ding Li; Jiang-Bo Duan; Bing Li; Ji-Hong Guo
Journal:  J Huazhong Univ Sci Technolog Med Sci       Date:  2013-10-20

10.  Implantable cardioverter defibrillator lead-related methicillin resistant Staphylococcus aureus endocarditis: Importance of heightened awareness.

Authors:  Obiora F Anusionwu; Cheri Smith; Alan Cheng
Journal:  World J Cardiol       Date:  2012-07-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.