Accuracy and reproducibility of biometry using partial coherence interferometry.

Rafe Connors1, Peter Boseman, Randall J Olson.   

Abstract

PURPOSE: To determine whether partial coherence interferometry is comparable to standard contact ultrasonic biometry in accuracy, reproducibility, and ease of use.
SETTING: Academic university practice.
METHODS: In this prospective study, simultaneous biometry was performed in 91 consecutive patients (111 eyes) using partial coherence interferometry (IOLMaster, Zeiss Humphrey Systems) and ultrasonic biometry (I(3), Innovative Imaging, Inc.). The differences between the predicted spherical equivalent in diopters (D) and the actual results with both techniques were compared.
RESULTS: The IOLMaster was significantly better in the mean absolute error (0.533 D +/- 0.589 [SD] versus 0.757 +/- 0.723 D; P = .012) and the percentage of eyes within +/-0.5 D (61.2% versus 42.3%; P = .003) and +/-1.0 D (87.4% versus 77.5%; P =.05) of the predicted refraction.
CONCLUSION: The IOLMaster was more accurate and reproducible than the contact ultrasonographic technique.

Entities:  

Mesh:

Year:  2002        PMID: 11821202     DOI: 10.1016/s0886-3350(01)01179-8

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  20 in total

1.  IOLMaster biometry: refractive results of 100 consecutive cases.

Authors:  H Eleftheriadis
Journal:  Br J Ophthalmol       Date:  2003-08       Impact factor: 4.638

2.  Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation.

Authors:  Mirjana Bjeloš Rončević; Mladen Bušić; Ivan Cima; Biljana Kuzmanović Elabjer; Damir Bosnar; Daliborka Miletić
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-09-18       Impact factor: 3.117

3.  Light-adjustable lens: development of in vitro nomograms.

Authors:  Daniel M Schwartz; Christian A Sandstedt; Shiao H Chang; Julie A Kornfield; Robert H Grubbs
Journal:  Trans Am Ophthalmol Soc       Date:  2004

4.  Which keratometer is most reliable for correcting astigmatism with toric intraocular lenses?

Authors:  Minwook Chang; Su-Yeon Kang; Hyo Myung Kim
Journal:  Korean J Ophthalmol       Date:  2012-01-14

5.  Light-adjustable lens.

Authors:  Daniel M Schwartz
Journal:  Trans Am Ophthalmol Soc       Date:  2003

6.  Axial eye length measurements pre-and post-laser-assisted in situ keratomileusis using the IOL Master: a pilot study.

Authors:  Spyridon E Chalkiadakis; Georgios A Amariotakis; Efstratios A Parikakis; Vasileios G Peponis
Journal:  Clin Ophthalmol       Date:  2010-11-08

7.  [Measurement of accommodation using optical biometry].

Authors:  A Nurispahic; K Kotliar; I Lanzl
Journal:  Ophthalmologe       Date:  2008-04       Impact factor: 1.059

8.  Reduced macular thickness and macular vessel density in early-treated adult patients with PKU.

Authors:  Csilla Serfozo; Andras Gellert Barta; Endre Horvath; Csaba Sumanszki; Bela Csakany; Miklos Resch; Zoltan Zsolt Nagy; Peter Reismann
Journal:  Mol Genet Metab Rep       Date:  2021-05-05

9.  Intraocular lens power calculation using IOLMaster and various formulas in short eyes.

Authors:  Young Rae Roh; Sang Mok Lee; Young Keun Han; Mee Kum Kim; Won Ryang Wee; Jin Hak Lee
Journal:  Korean J Ophthalmol       Date:  2011-05-24

10.  Refractive predictability of partial coherence interferometry and factors that can affect it.

Authors:  Seung Mo Kim; Joohyun Choi; Sangkyung Choi
Journal:  Korean J Ophthalmol       Date:  2009-03-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.