BACKGROUND: The ideal bearing combination for total hip arthroplasty (THA) remains debatable. Highly cross-linked polyethylene (XLPE) is widely used, but long-term wear rates are not fully known, nor is how much the initial "creep," if any, affects overall wear. Additionally, the use of oxidized zirconium (OxZir) is purported to lower polyethylene wear rates, but this has not been proven. We present the 10-year data of a cohort of patients who underwent THA. Patients were prospectively randomized to 1 of 4 bearing combinations: a conventional ultra-high molecular weight polyethylene (UHMWPE) or XLPE acetabular liner coupled with either a cobalt-chromium (CoCr) or OxZir femoral head. The aims of the study were to (1) assess the extent to which creep affected overall wear rates and (2) assess wear rates between OxZir and CoCr with polyethylene.Entities:
Year: 2020 PMID: 33123666 PMCID: PMC7418907 DOI: 10.2106/JBJS.OA.19.00066
Source DB: PubMed Journal: JB JS Open Access ISSN: 2472-7245
Fig. 1Flow diagram of enrolled patients. F/U = follow-up.
Fig. 2The annual linear rates of wear for the 4 bearing surfaces (mm/yr). UHMWPE = ultra-high molecular weight polyethylene, CoCr = cobalt-chromium, Zir = oxidized zirconium, and XLPE = highly cross-linked polyethylene.
Fig. 3Bar graph showing the comparative linear wear rates (mm/yr) without adjustment for creep and with adjustment (the first 3 months eliminated). UHMWPE = ultra-high molecular weight polyethylene, CoCr = cobalt-chromium, Zir = oxidized zirconium, and XLPE = highly cross-linked polyethylene.
Fig. 4The percentage influence of creep relative to total linear wear for pooled UHMPE (ultra-high molecular weight polyethylene) and XLPE (highly cross-linked polyethylene) (mean and standard deviation).
Baseline Demographic and Operative Characteristics and 10-Year Wear Rates of UHMWPE and XLPE Groups (Pooled Head Types)*
| UHMWPE | XLPE | P Value | |
| No. of hips available for analysis | 30 | 40 | |
| Sex | |||
| Male | 16 | 21 | |
| Female | 14 | 19 | |
| Age | 52.7 ± 7.9 | 51.6 ± 10.8 | 0.325 |
| Operative side | |||
| Left | 16 | 20 | |
| Right | 14 | 20 | |
| Body mass index
| 27.4 | 26.9 | 0.34 |
| Acetabular cup size | 55.3 ± 3.5 | 54.5 ± 3.1 | 0.179 |
| Cup abduction angle | 44.2 ± 7.7 | 46.4 ± 6.2 | 0.100 |
| Cup anteversion angle | 12.5 ± 9.0 | 13.1 ± 8.5 | 0.166 |
| Total linear wear at 10 yr | 2.44 ± 1.14 | 0.620 ± 0.523 | <0.00001 |
| Linear wear rate at 10 yr | 0.242 ± 0.12 | 0.062 ± 0.052 | <0.00001 |
| Adjusted linear wear rate at 10 yr | 0.177 ± 0.095 | 0.0315 ± 0.0371 | <0.00001 |
| Total volumetric wear at 10 yr
| 542.1 ± 310.3 | 185.9 ± 110.1 | <0.00001 |
| % of measured wear at 10 yr due to creep | 30.7 ± 18.6 | 55.0 ± 28.5 | 0.000133 |
The values are given as the mean and standard deviation, except where otherwise indicated.
Significant difference between groups.
Comparative Wear Rates Among the 4 Groups*
| UHMWPE + CoCr | UHMWPE + OxZir | XLPE + CoCr | XLPE + OxZir | P Value | |
| Total linear wear at 10 yr | 2.49 ± 1.13 | 2.38 ± 1.19 | 0.710 ± 0.552 | 0.520 ± 0.485 | <0.00001 |
| Linear wear rate at 10 yr | 0.250 ± 0.112 | 0.233 ± 0.112 | 0.071 ± 0.055 | 0.053 ± 0.050 | <0.00001 |
| Adjusted linear wear rate at 10 yr
| 0.181 ± 0.096 | 0.170 ± 0.099 | 0.0387 ± 0.0435 | 0.0227 ± 0.0262 | <0.00001 |
| Total volumetric wear at 10 yr
| 535.5 ± 260.3 | 550.6 ± 377.1 | 196.5 ± 118.6 | 174.2 ± 102.0 | <0.00001 |
| % of measured wear at 10 yr due to creep | 29.5 ± 18.2 | 32.3 ± 19.9 | 58.3 ± 26.0 | 50.9 ± 31.8 | 0.00941 |
The values are given as the mean and standard deviation.
Significant for overall comparisons.