The purpose of the investigation was to assess the clinical and radiographic outcomes in four bearing surfaces. Eighty patients (91 hips) undergoing total hip arthroplasty between 2004 and 2007 were randomized to one of four bearing surfaces: (1) cobalt-chrome (CoCr) and ultra-high molecular weight polyethylene (UHMWPE); (2) CoCr and XLPE; (3) Oxinium and UHMWPE; and (4) Oxinium and XLPE. The mean follow-up for this study was 6.8 years. There were no significant differences in clinical outcomes. The linear wear rates for the four groups were 0.241 mm/year, 0.076 mm/year, 0.238 mm/year and 0.061 mm/year respectively. HXLPE results in significantly less wear than UHMWPE. However, we found no significant reduction in wear rate by using Oxinium in place of CoCr femoral heads at early follow-up.
Copyright © 2014 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  clinical outcomes; cross-linked polyethylene; oxidized zirconium; polyethylene wear; primary hip arthroplasty

Mesh:

Substances:

Year:  2014        PMID: 24961892     DOI: 10.1016/j.arth.2014.04.046

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  10 in total

1.  Do oxidized zirconium femoral heads reduce polyethylene wear in cemented THAs? A blinded randomized clinical trial.

Authors:  Amine Zaoui; Samer El Hage; Jean Langlois; Caroline Scemama; Jean Pierre Courpied; Moussa Hamadouche
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

Review 2.  [Tribology in hip arthroplasty : Benefits of different materials].

Authors:  J Philippe Kretzer; Maximilian Uhler; Sebastian Jäger; Therese Bormann; Robert Sonntag; Mareike Schonhoff; Stefan Schröder
Journal:  Orthopade       Date:  2021-02-25       Impact factor: 1.087

3.  Oxidized zirconium head on crosslinked polyethylene liner in total hip arthroplasty: a 7- to 12-year in vivo comparative wear study.

Authors:  George K Karidakis; Theofilos Karachalios
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

4.  Editor's Spotlight/Take 5: Polyethylene Wear Increases in Liners Articulating With Scratched Oxidized Zirconium Femoral Heads.

Authors:  Seth S Leopold
Journal:  Clin Orthop Relat Res       Date:  2018-02       Impact factor: 4.176

5.  Low wear rates seen in THAs with highly crosslinked polyethylene at 9 to 14 years in patients younger than age 50 years.

Authors:  Kevin L Garvin; Tyler C White; Anand Dusad; Curtis W Hartman; John Martell
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

Review 6.  The future role of metal-on-metal hip resurfacing.

Authors:  Gulraj S Matharu; Hemant G Pandit; David W Murray; Ronan B C Treacy
Journal:  Int Orthop       Date:  2015-02-24       Impact factor: 3.075

7.  Wear Rates of XLPE Nearly 50% Lower Than Previously Thought After Adjusting for Initial Creep: An RCT Comparing 4 Bearing Combinations.

Authors:  Amir Khoshbin; James Wu; Sarah Ward; Luana T Melo; Emil H Schemitsch; James P Waddell; Amit Atrey
Journal:  JB JS Open Access       Date:  2020-06-05

8.  Is there any difference in survivorship of total hip arthroplasty with different bearing surfaces? A systematic review and network meta-analysis.

Authors:  Si Yin; Dangfeng Zhang; Hui Du; Heng Du; Zhanhai Yin; Yusheng Qiu
Journal:  Int J Clin Exp Med       Date:  2015-11-15

9.  Thirteen-year wear rate comparison of highly crosslinked and conventional polyethylene in total hip arthroplasty: long-term follow-up of a prospective randomized controlled trial.

Authors:  Matthew G Teeter; Xunhua Yuan; Lyndsay E Somerville; Steven J MacDonald; Richard W McCalden; Douglas D Naudie
Journal:  Can J Surg       Date:  2017-06       Impact factor: 2.089

10.  Long-term performance of oxidized zirconium on conventional and highly cross-linked polyethylene in total hip arthroplasty.

Authors:  Alberto V Carli; Anay R Patel; Michael B Cross; David J Mayman; Kaitlin M Carroll; Paul M Pellicci; Seth A Jerabek
Journal:  SICOT J       Date:  2020-05-07