Anna Arns1, Manuel Blessing1, Jens Fleckenstein1, Dzmitry Stsepankou1, Judit Boda-Heggemann1, Juergen Hesser1, Frank Lohr2, Frederik Wenz1, Hansjoerg Wertz1.
Abstract
PURPOSE: Combined ultrafast 90°+90° kV-MV-CBCT within single breath-hold of 15s has high clinical potential for accelerating imaging for lung cancer patients treated with deep inspiration breath-hold (DIBH). For clinical feasibility of kV-MV-CBCT, dose exposure has to be small compared to prescribed dose. In this study, kV-MV dose output is evaluated and compared to clinically-established kV-CBCT.Entities:
Mesh:
Year: 2017 PMID: 29125857 PMCID: PMC5681289 DOI: 10.1371/journal.pone.0187710
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Technical setup of applied imaging presets.
| Preset | kV-Chest1-CBCT | kV-Chest2-CBCT | kV-MV-CBCT |
|---|---|---|---|
| 120kVp | 120kVp | ||
| 1mAs (25mA, 40ms) per frame | 0.4mAs (20mA, 20ms) per frame | 0.1mAs (10mA, 10ms) per frame | |
| 650 | 660 | ||
| off-centered (M) | off-centered (M) | centered (S) | |
| F0 (empty), | F1 (bow-tie), | ||
| M20 (27.67cm x 42.64cm) | M20 (27.67cm x 42.64cm) | S20 (27.67cm x 27.67cm) | |
| ½ rpm | ½ rpm | 1 rpm | |
| 360° | 360° | 90° + cone angle | |
| (here: 0° to 100°) | |||
| 2min | minimum 2min | 15s | |
| 16mGy | 5mGy | ||
Fig 1HVL measurement setup.
Fig 2Setup and layout of different measurement positions of (a) inhomogeneous thorax phantom for patient simulation (1: tumor in lower left lung, 2: left lung, 3: right lung, 4: upper periphery, 5: spinal cord, 6: body center), and (b) CT dose body phantom for QA purposes.
Absolute dose correction factors for kV- and MV-energy, based on AAPM TG61 report and IAEA TRS 398, respectively, whereas the values for HVLBS and Q were calculated from the beam quality measurements.
| kV-energy correction factors | MV-energy correction factors | ||||
|---|---|---|---|---|---|
| kV-Chest1-CBCT | kV-Chest2-CBCT | kV-MV-CBCT | 6MV linac | ||
| 7.01mm Al | 7.63mm Al | 5.57mm Al | 0.6858 | ||
| 1.0394 | 1.0417 | 1.0346 | 0,9894 | ||
| 1.022 | 1.023 | 1.019 | 1.0·108 Gy/C | ||
| 1 | 1 | 1 | |||
| 8.757·107 Gy/C | |||||
Fig 3(a) HVL measurement results for all kV-dependent CBCT-presets, (b) PDD for MV beam quality determination, and (c) simulation of 4MU distributed over a 100° gantry arc with the TPS.
Absorbed dose for different presets in inhomogeneous thorax phantom (column 8), ratio between the imaging methods and reference dose output kV-MV (D/Dref) (column 9), comparison of MV-contribution dose output between measured and calculated (TPS) absolute dose by percentage difference (column 3–5), and separate contributions of kV- and RBE-corrected MV absorbed dose to kV-MV-CBCT in columns 6+7 (GE: Gray-equivalent with RBE = 2 for kV-dose contribution).
| Position | CBCT | calculated (TPS): | measured: | final absorbed dose | ||||
|---|---|---|---|---|---|---|---|---|
| MV part | MV part | MV part (kV-equivalent) | kV part | |||||
| 1: tumor- lower left lung | kV-MV | 40.40 mGy | 40.26 mGy | 20.13 mGE | 0.38 mGy | |||
| kV-Chest1 | ||||||||
| kV-Chest2 | ||||||||
| 2: upper left lung | kV-MV | 44.90 mGy | 45.10 mGy | 22.55 mGE | 0.33 mGy | 22.88 mGE | ||
| kV-Chest1 | 32.00 mGy | |||||||
| kV-Chest2 | 9.13 mGy | |||||||
| 3: right lung | kV-MV | 13.40 mGy | 13.02 mGy | 6.51 mGE | 0.09 mGy | |||
| kV-Chest1 | ||||||||
| kV-Chest2 | ||||||||
| 4: upper center | kV-MV | 37.30 mGy | 37.95 mGy | 18.98 mGE | 0.13 mGy | 19.10 mGE | ||
| kV-Chest1 | 23.06 mGy | |||||||
| kV-Chest2 | 6.21 mGy | |||||||
| 5: spinal cord | kV-MV | 30.00 mGy | 29.82 mGy | 14.91 mGE | 0.20 mGy | 15.12 mGE | ||
| kV-Chest1 | 16.45 mGy | |||||||
| kV-Chest2 | 4.62 mGy | |||||||
| 6: body center | kV-MV | 34.30 mGy | 34.83 mGy | 17.41 mGE | 0.16 mGy | 17.57 mGE | ||
| kV-Chest1 | 21.54 mGy | |||||||
| kV-Chest2 | 5.96 mGy | |||||||
Determination of weighted CTDI for different CBCT-presets (GE: Gray-equivalent with RBE = 2 for kV-dose contribution) (columns 2, 3 and 5) and ratio between the imaging methods and reference dose output kV-MV (D/Dref) (columns 4 and 6).
| Position | kV-MV-CBCT | kV-Chest1-CBCT | kV-Chest2-CBCT | ||
|---|---|---|---|---|---|
| 14.70 mGE | 9.68 mGy | 3.42 mGy | |||
| 19.83 mGE | 20.52 mGy | 6.41 mGy | |||
| 16.72 mGE | 20.24 mGy | 6.00 mGy | |||
| 5.66 mGE | 20.61 mGy | 5.53 mGy | |||
| 5.74 mGE | 22.27 mGy | 6.74 mGy | |||