Kilovoltage cone-beam CT: comparative dose and image quality evaluations in partial and full-angle scan protocols.

Sangroh Kim1, Sua Yoo, Fang-Fang Yin, Ehsan Samei, Terry Yoshizumi.   

Abstract

PURPOSE: To assess imaging dose of partial and full-angle kilovoltage CBCT scan protocols and to evaluate image quality for each protocol.
METHODS: The authors obtained the CT dose index (CTDI) of the kilovoltage CBCT protocols in an on-board imager by ion chamber (IC) measurements and Monte Carlo (MC) simulations. A total of six new CBCT scan protocols were evaluated: Standard-dose head (100 kVp, 151 mA s, partial-angle), low-dose head (100 kVp, 75 mA s, partial-angle), high-quality head (100 kVp, 754 mA s, partial-angle), pelvis (125 kVp, 706 mA s, full-angle), pelvis spotlight (125 kVp, 752 mA s, partial-angle), and low-dose thorax (110 kVp, 271 mA s, full-angle). Using the point dose method, various CTDI values were calculated by (1) the conventional weighted CTDI (CTDIw) calculation and (2) Bakalyar's method (CTDIwb). The MC simulations were performed to obtain the CTDIw and CTDIwb, as well as from (3) central slice averaging (CTDI(2D)) and (4) volume averaging (CTDI(3D)) techniques. The CTDI values of the new protocols were compared to those of the old protocols (full-angle CBCT protocols). Image quality of the new protocols was evaluated following the CBCT image quality assurance (QA) protocol [S. Yoo et al., "A quality assurance program for the on-board image, "Med. Phys. 33(11), 4431-4447 (2006)] testing Hounsfield unit (HU) linearity, spatial linearity/resolution, contrast resolution, and HU uniformity.
RESULTS: The CTDI, were found as 6.0, 3.2, 29.0, 25.4, 23.8, and 7.7 mGy for the new protocols, respectively. The CTDI, and CTDIwb differed within +3% between IC measurements and MC simulations. Method (2) results were within +/- 12% of method (1). In MC simulations, the CTDIw and CTDIwb were comparable to the CTDI(2D) and CTDI(3D) with the differences ranging from -4.3% to 20.6%. The CTDI(3D) were smallest among all the CTDI values. CTDIw of the new protocols were found as approximately 14 times lower for standard head scan and 1.8 times lower for standard body scan than the old protocols, respectively. In the image quality QA tests, all the protocols except low-dose head and low-dose thorax protocols were within the tolerance in the HU verification test. The HU value for the two protocols was always higher than the nominal value. All the protocols passed the spatial linearity/resolution and HU uniformity tests. In the contrast resolution test, only high-quality head and pelvis scan protocols were within the tolerance. In addition, crescent effect was found in the partial-angle scan protocols.
CONCLUSIONS: The authors found that CTDIw of the new CBCT protocols has been significantly reduced compared to the old protocols with acceptable image quality. The CTDIw values in the point dose method were close to the volume averaging method within 9%-21% for all the CBCT scan protocols. The Bakalyar's method produced more accurate dose estimation within 14%. The HU inaccuracy from low-dose head and low-dose thorax protocols can render incorrect dose results in the treatment planning system. When high soft-tissue contrast data are desired, high-quality head or pelvis scan protocol is recommended depending on the imaging area. The point dose method can be applicable to estimate CBCT dose with reasonable accuracy in the clinical environment.

Entities:  

Mesh:

Year:  2010        PMID: 20831072     DOI: 10.1118/1.3438478

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  21 in total

1.  Antiscatter grids in mobile C-arm cone-beam CT: effect on image quality and dose.

Authors:  S Schafer; J W Stayman; W Zbijewski; C Schmidgunst; G Kleinszig; J H Siewerdsen
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

2.  Practical dose point-based methods to characterize dose distribution in a stationary elliptical body phantom for a cone-beam C-arm CT system.

Authors:  Jang-Hwan Choi; Dragos Constantin; Arundhuti Ganguly; Erin Girard; Richard L Morin; Robert L Dixon; Rebecca Fahrig
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

3.  Development and validation of a measurement-based source model for kilovoltage cone-beam CT Monte Carlo dosimetry simulations.

Authors:  Kyle McMillan; Michael McNitt-Gray; Dan Ruan
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

4.  Bow-tie wobble artifact: effect of source assembly motion on cone-beam CT.

Authors:  Dandan Zheng; John C Ford; Jun Lu; Dimitrios Lazos; Geoffrey D Hugo; Damodar Pokhrel; Lisha Zhang; Jeffrey F Williamson
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

5.  Impact of prone, supine and oblique patient positioning on CBCT image quality, contrast-to-noise ratio and figure of merit value in the maxillofacial region.

Authors:  Juha Koivisto; Maureen van Eijnatten; Jorma Järnstedt; Kirsi Holli-Helenius; Prasun Dastidar; Jan Wolff
Journal:  Dentomaxillofac Radiol       Date:  2017-04-07       Impact factor: 2.419

6.  Respiratory triggered 4D cone-beam computed tomography: a novel method to reduce imaging dose.

Authors:  Benjamin J Cooper; Ricky T O'Brien; Salim Balik; Geoffrey D Hugo; Paul J Keall
Journal:  Med Phys       Date:  2013-04       Impact factor: 4.071

7.  Contrast-to-noise ratio with different settings in a CBCT machine in presence of different root-end filling materials: an in vitro study.

Authors:  Husniye Demirturk Kocasarac; Dilek Helvacioglu Yigit; Boulos Bechara; Alper Sinanoglu; Marcel Noujeim
Journal:  Dentomaxillofac Radiol       Date:  2016-03-08       Impact factor: 2.419

8.  Metal artefact reduction with cone beam CT: an in vitro study.

Authors:  B B Bechara; W S Moore; C A McMahan; M Noujeim
Journal:  Dentomaxillofac Radiol       Date:  2012-01-12       Impact factor: 2.419

9.  Characterization and comparison of imaging contrast enhancement with PEG-functionalized gold nanoparticles in kV cone beam computed tomography and computed tomography imaging.

Authors:  Tara Gray; Nema Bassiri; Shaquan David; Devanshi Yogeshkumar Patel; Mohammad Abdul-Moqueet; Neil Kirby; Kathryn M Mayer
Journal:  Biomed Phys Eng Express       Date:  2020-05-29

10.  Distribution and Detection of Radiopaque Beads after Hepatic Transarterial Embolization in Swine: Cone-Beam CT versus MicroCT.

Authors:  John G Thompson; William van der Sterren; Ivane Bakhutashvili; Imramsjah M van der Bom; Alessandro G Radaelli; John W Karanian; Juan Esparza-Trujillo; David L Woods; Andrew L Lewis; Bradford J Wood; William F Pritchard
Journal:  J Vasc Interv Radiol       Date:  2018-03-02       Impact factor: 3.464

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.