Breath-hold target localization with simultaneous kilovoltage/megavoltage cone-beam computed tomography and fast reconstruction.

Manuel Blessing1, Dzmitry Stsepankou, Hansjoerg Wertz, Anna Arns, Frank Lohr, Jürgen Hesser, Frederik Wenz.   

Abstract

PURPOSE: Hypofractionated high-dose radiotherapy for small lung tumors has typically been based on stereotaxy. Cone-beam computed tomography and breath-hold techniques have provided a noninvasive basis for precise cranial and extracranial patient positioning. The cone-beam computed tomography acquisition time of 60 s, however, is beyond the breath-hold capacity of patients, resulting in respiratory motion artifacts. By combining megavoltage (MV) and kilovoltage (kV) photon sources (mounted perpendicularly on the linear accelerator) and accelerating the gantry rotation to the allowed limit, the data acquisition time could be reduced to 15 s. METHODS AND MATERIALS: An Elekta Synergy 6-MV linear accelerator, with iViewGT as the MV- and XVI as the kV-imaging device, was used with a Catphan phantom and an anthropomorphic thorax phantom. Both image sources performed continuous image acquisition, passing an angle interval of 90° within 15 s. For reconstruction, filtered back projection on a graphics processor unit was used. It reconstructed 100 projections acquired to a 512 × 512 × 512 volume within 6 s.
RESULTS: The resolution in the Catphan phantom (CTP528 high-resolution module) was 3 lines/cm. The spatial accuracy was within 2-3 mm. The diameters of different tumor shapes in the thorax phantom were determined within an accuracy of 1.6 mm. The signal-to-noise ratio was 68% less than that with a 180°-kV scan. The dose generated to acquire the MV frames accumulated to 82.5 mGy, and the kV contribution was <6 mGy.
CONCLUSION: The present results have shown that fast breath-hold, on-line volume imaging with a linear accelerator using simultaneous kV-MV cone-beam computed tomography is promising and can potentially be used for image-guided radiotherapy for lung cancer patients in the near future.
Copyright © 2010 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20554124     DOI: 10.1016/j.ijrobp.2010.01.030

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  11 in total

1.  [Image-guided radiation therapy. Paradigm change in radiation therapy].

Authors:  F Wenz; C Belka; M Reiser; S O Schönberg
Journal:  Radiologe       Date:  2012-03       Impact factor: 0.635

Review 2.  kV cone-beam CT-based IGRT: a clinical review.

Authors:  Judit Boda-Heggemann; Frank Lohr; Frederik Wenz; Michael Flentje; Matthias Guckenberger
Journal:  Strahlenther Onkol       Date:  2011-04-26       Impact factor: 3.621

Review 3.  Are three doses of stereotactic ablative radiotherapy (SABR) more effective than 30 doses of conventional radiotherapy?

Authors:  Anna O Simeonova; Katharina Fleckenstein; Hansjörg Wertz; Anian Frauenfeld; Judit Boda-Heggemann; Frank Lohr; Frederik Wenz
Journal:  Transl Lung Cancer Res       Date:  2012-03

4.  Towards clinical implementation of ultrafast combined kV-MV CBCT for IGRT of lung cancer : Evaluation of registration accuracy based on phantom study.

Authors:  Anna Arns; Manuel Blessing; Jens Fleckenstein; Dzmitry Stsepankou; Judit Boda-Heggemann; Anna Simeonova-Chergou; Jürgen Hesser; Frank Lohr; Frederik Wenz; Hansjörg Wertz
Journal:  Strahlenther Onkol       Date:  2016-02-10       Impact factor: 3.621

5.  Concurrent kilovoltage CBCT imaging and megavoltage beam delivery: suppression of cross-scatter with 2D antiscatter grids and grid-based scatter sampling.

Authors:  Farhang Bayat; Mohamed Elsayed Eldib; Brian Kavanagh; Moyed Miften; Cem Altunbas
Journal:  Phys Med Biol       Date:  2022-08-09       Impact factor: 4.174

6.  Megavoltage cross-scatter rejection and correction using 2D antiscatter grids in kilovoltage CBCT imaging.

Authors:  Farhang Bayat; Mohamed Elsayed Eldib; Cem Altunbas
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2022-04-04

Review 7.  Recent advances in radiation oncology.

Authors:  Cristina Garibaldi; Barbara Alicja Jereczek-Fossa; Giulia Marvaso; Samantha Dicuonzo; Damaris Patricia Rojas; Federica Cattani; Anna Starzyńska; Delia Ciardo; Alessia Surgo; Maria Cristina Leonardi; Rosalinda Ricotti
Journal:  Ecancermedicalscience       Date:  2017-11-30

8.  Phantom-based evaluation of dose exposure of ultrafast combined kV-MV-CBCT towards clinical implementation for IGRT of lung cancer.

Authors:  Anna Arns; Manuel Blessing; Jens Fleckenstein; Dzmitry Stsepankou; Judit Boda-Heggemann; Juergen Hesser; Frank Lohr; Frederik Wenz; Hansjoerg Wertz
Journal:  PLoS One       Date:  2017-11-10       Impact factor: 3.240

Review 9.  Applications of linac-mounted kilovoltage Cone-beam Computed Tomography in modern radiation therapy: A review.

Authors:  Kavitha Srinivasan; Mohammad Mohammadi; Justin Shepherd
Journal:  Pol J Radiol       Date:  2014-07-03

10.  An investigation of gantry angle data accuracy for cine-mode EPID images acquired during arc IMRT.

Authors:  Peter M McCowan; Daniel W Rickey; Pejman Rowshanfarzad; Peter B Greer; William Ansbacher; Boyd M McCurdy
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.