Daniel E Hyer1, Christopher F Serago, Siyong Kim, Jonathan G Li, David E Hintenlang.
Abstract
The main purpose of this work was to quantify patient organ doses from the two kilovoltage cone beam computed tomography (CBCT) systems currently available on medical linear accelerators, namely the X-ray Volumetric Imager (XVI, Elekta Oncology Systems) and the On-Board Imager (OBI, Varian Medical Systems). Organ dose measurements were performed using a fiber-optic coupled (FOC) dosimetry system along with an adult male anthropomorphic phantom for three different clinically relevant scan sites: head, chest, and pelvis. The FOC dosimeter was previously characterized at diagnostic energies by Hyer et al. [Med Phys 2009;36(5):1711-16] and a total uncertainty of approximately 4% was found for in-phantom dose measurements. All scans were performed using current manufacturer-installed clinical protocols and appropriate bow-tie filters. A comparison of image quality between these manufacturer-installed protocols was also performed using a Catphan 440 image quality phantom. Results indicated that for the XVI, the dose to the lens of the eye (1.07 mGy) was highest in a head scan, thyroid dose (19.24 mGy) was highest in a chest scan, and gonad dose (29 mGy) was highest in a pelvis scan. For the OBI, brain dose (3.01 mGy) was highest in a head scan, breast dose (5.34 mGy) was highest in a chest scan, and gonad dose (34.61 mGy) was highest in a pelvis scan. Image quality measurements demonstrated that the OBI provided superior image quality for all protocols, with both better spatial resolution and low-contrast detectability. The measured organ doses were also used to calculate a reference male effective dose to allow further comparison of the two machines and imaging protocols. The head, chest, and pelvis scans yielded effective doses of 0.04, 7.15, and 3.73 mSv for the XVI, and 0.12, 1.82, and 4.34mSv for the OBI, respectively.Entities:
Mesh:
Year: 2010 PMID: 20592702 PMCID: PMC5719945 DOI: 10.1120/jacmp.v11i2.3183
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Scanning configuration for each imaging protocol used. As seen, the collimation was changed and the detector panel was shifted orthogonal to the kV beam axis when a larger FOV was required. XVI head scan, no bow‐tie filter, imaging panel centered, 27 cm FOV; (b): XVI chest and pelvis scan, bow‐tie filter, imaging panel shifted, 41 cm FOV; (c): OBI head scan, bow‐tie filter, imaging panel centered, 25 cm FOV; (d): OBI chest and pelvis scan, half bow‐tie filter, imaging panel shifted, 45 cm FOV.
Nominal technical settings and measured HVLs for each imaging protocol investigated.
|
|
|
| ||||
|---|---|---|---|---|---|---|
|
|
|
|
|
|
| |
| kV Collimator | S20 | M20 | M10 | ‐ | ‐ | |
| kV Filter | F0 | F1 | F1 | Full bow‐tie | Half bow‐tie | Half bow‐tie |
| kVp | 100 | 120 | 120 | 100 | 110 | 125 |
| mA | 10 | 40 | 64 | 20 | 20 | 80 |
| ms/projection | 10 | 40 | 40 | 20 | 20 | 13 |
| # of Projections | 361 | 643 | 643 | 360 | 655 | 655 |
| Total mAs | 36.1 | 1028.8 | 1646.1 | 145 | 262 | 680 |
| Meaured HVL (mm Al) | 5.9 | 8.9 | 8.9 | 5.4 | 5.7 | 6.4 |
| Acquisition Angle | 350°‐190° | 273°‐269° | 273°‐269° | 88°‐292° | 88°‐92° | 88°‐92° |
| cw | cw | cw | cw/ccw | cw/ccw | cw/ccw | |
| Acquisition Time |
|
|
|
|
|
|
| Axial Field of View (cm) | 27 | 41 | 41 | 25 | 45 | 45 |
| Long. Field of View (cm) | 26 | 26 | 12.5 | 18 | 16 | 16 |
On the XVI, F0 is the designation for no filter and F1 is the designation for a full bow‐tie filter.
For acquisition angle, cw means clockwise and ccw means counter‐clockwise, as viewed from the patient table.
Figure 2Completed torso of the 50th percentile adult male anthropomorphic phantom.
Figure 3Schematic detailing the components of the dosimetry system. (The exterior heat shrink tubing extends past the scintillator to restrict any ambient light from entering the assembly, but is not shown to aid in visualization.)
Figure 4Axial slice of the physical phantom with an FOC dosimeter installed in the right kidney for dose measurements (displayed in prone position but phantom was oriented in supine position for measurements).
Organs investigated and number of measurement locations
|
|
|
|---|---|
| Brain | 4 |
| Salivary glands | 6 |
| Thyroid | 1 |
| Esophagus | 6 |
| Lung | 8 |
| Breast | 2 |
| Liver | 4 |
| Stomach | 4 |
| Colon | 10 |
| Bladder | 1 |
| Gonads (testes) | 2 |
| Bone marrow | 19 |
| Bone surface | 19 |
| Skin | 2 |
| Remainder Organs: | |
| Extrathoracic region | 4 |
| Oral mucosa | 1 |
| Thymus | 1 |
| Heart | 1 |
| Spleen | 2 |
| Adrenals | 2 |
| Gall bladder | 1 |
| Kidneys | 2 |
| Pancreas | 1 |
| Small intestine | 6 |
| Prostate | 1 |
| Other Organs: | |
| Lens | 2 |
Weight fractions of red bone marrow, , and endosteum, , for various locations of interest of the 50th percentile adult male hybrid phantom.
|
|
| |
|---|---|---|
| Cranium | 0.049 | 0.118 |
| Cervical vertebrae | 0.032 | 0.021 |
| Humerus‐proximal | 0.031 | 0.027 |
| Clavicles | 0.011 | 0.007 |
| Scapula | 0.091 | 0.070 |
| Ribs | 0.102 | 0.042 |
| Sternum | 0.026 | 0.010 |
| Thoracic vertebrae | 0.130 | 0.048 |
| Lumbar vertebrae | 0.130 | 0.054 |
| Os coxae | 0.265 | 0.173 |
| Sacrum | 0.081 | 0.037 |
| Femur‐proximal | 0.045 | 0.065 |
Organ and effective doses from the Elekta XVI CBCT system.
|
|
| ||
|---|---|---|---|
|
|
|
| |
| Brain | 0.70 | 0.49 | ‐ |
| Salivary glands | 0.78 | 1.86 | ‐ |
| Thyroid | 0.05 | 19.24 | ‐ |
| Esophagus | 0.02 | 13.56 | ‐ |
| Lung | 0.01 | 14.29 | 0.02 |
| Breast | ‐ | 16.80 | ‐ |
| Liver | ‐ | 6.58 | 0.19 |
| Stomach | ‐ | 4.68 | 0.23 |
| Colon | ‐ | 0.40 | 2.04 |
| Bladder | ‐ | 0.03 | 15.67 |
| Gonads (testes) | ‐ | ‐ | 29.00 |
| Bone marrow (whole body) | 0.07 | 5.14 | 1.05 |
| Bone marrow (irradiated site) | 0.80 | 12.42 | 5.50 |
| Bone surface (whole body) | 0.11 | 2.59 | 1.17 |
| Bone surface (irradiated site) | 0.80 | 12.42 | 5.50 |
| Skin (whole body) | 0.09 | 2.62 | 3.07 |
| Skin (irradiated site) | 1.34 | 14.92 | 27.88 |
| Remainder Organs: | |||
| Extrathoracic region | 0.60 | 5.21 | ‐ |
| Oral mucosa | 0.69 | 1.34 | ‐ |
| Thymus | 0.01 | 14.29 | ‐ |
| Heart | ‐ | 13.87 | 0.10 |
| Spleen | ‐ | 7.17 | 0.20 |
| Adrenals | ‐ | 3.76 | 0.23 |
| Gall bladder | ‐ | 1.83 | 0.28 |
| Kidneys | ‐ | 1.20 | 0.31 |
| Pancreas | ‐ | 1.21 | 0.33 |
| Small intestine | ‐ | 0.28 | 1.06 |
| Prostate | ‐ | ‐ | 27.63 |
| Other Organs: | |||
| Lens | 1.07 | 0.52 | ‐ |
| Effective dose (msV) | 0.04 | 7.15 | 3.73 |
Organ and effective doses from the Varian OBI CBCT system.
|
|
| ||
|---|---|---|---|
|
|
|
| |
| Brain | 3.01 | 0.14 | ‐ |
| Salivary glands | 2.42 | 0.30 | ‐ |
| Thyroid | ‐ | 2.38 | ‐ |
| Esophagus | 0.01 | 3.23 | ‐ |
| Lung | ‐ | 4.31 | 0.01 |
| Breast | ‐ | 5.34 | ‐ |
| Liver | ‐ | 0.97 | 0.28 |
| Stomach | ‐ | 0.74 | 0.30 |
| Colon | ‐ | ‐ | 3.26 |
| Bladder | ‐ | ‐ | 15.30 |
| Gonads (testes) | ‐ | ‐ | 34.61 |
| Bone marrow (whole body) | 0.28 | 1.29 | 1.14 |
| Bone marrow (irradiated site) | 3.45 | 3.27 | 5.77 |
| Bone surface (whole body) | 0.47 | 0.63 | 1.14 |
| Bone surface (irradiated site) | 3.45 | 3.27 | 5.77 |
| Skin (whole body) | 0.16 | 1.03 | 3.05 |
| Skin (irradiated site) | 2.39 | 5.85 | 27.77 |
| Remainder Organs: | |||
| Extrathoracic region | 1.06 | 0.85 | ‐ |
| Oral mucosa | 1.39 | 0.38 | ‐ |
| Thymus | ‐ | 4.83 | ‐ |
| Heart | ‐ | 4.50 | 0.08 |
| Spleen | ‐ | 0.93 | 0.28 |
| Adrenals | ‐ | 0.65 | 0.34 |
| Gall bladder | ‐ | 0.14 | 0.52 |
| Kidneys | ‐ | 0.08 | 0.59 |
| Pancreas | ‐ | 0.06 | 0.52 |
| Small intestine | ‐ | ‐ | 1.72 |
| Prostate | ‐ | ‐ | 27.25 |
| Other Organs: | |||
| Lens | 0.59 | 0.15 | ‐ |
| Effective dose (msV) | 0.12 | 1.82 | 4.34 |
Results of image quality tests for manufacturer installed protocols. Resolution was evaluated by the smallest pattern that could be visually resolved in the CTP592 module, and detectability was evaluated by the smallest acrylic sphere that could be visually resolved in the CTP401 module.
|
|
|
| ||||
|---|---|---|---|---|---|---|
|
|
|
|
|
|
| |
| Resolution |
|
|
|
|
|
|
| Detectability |
| 10 mm | 6 mm | 4 mm | 6 mm | 4 mm |
Figure 5Example of reconstructed images of CTP592 resolution phantom: XVI (left), OBI (right).