Fast kilovoltage/megavoltage (kVMV) breathhold cone-beam CT for image-guided radiotherapy of lung cancer.

Hansjoerg Wertz1, Dzmitry Stsepankou, Manuel Blessing, Michael Rossi, Chris Knox, Kevin Brown, Uwe Gros, Judit Boda-Heggemann, Cornelia Walter, Juergen Hesser, Frank Lohr, Frederik Wenz.   

Abstract

Long image acquisition times of 60-120 s for cone-beam CT (CBCT) limit the number of patients with lung cancer who can undergo volume image guidance under breathhold. We developed a low-dose dual-energy kilovoltage-megavoltage-cone-beam CT (kVMV-CBCT) based on a clinical treatment unit reducing imaging time to < or =15 s. Simultaneous kVMV-imaging was achieved by dedicated synchronization hardware controlling the output of the linear accelerator (linac) based on detector panel readout signals, preventing imaging artifacts from interference of the linac's MV-irradiation and panel readouts. Optimization was performed to minimize the imaging dose. Single MV-projections, reconstructed MV-CBCT images and images of simultaneous 90 degrees kV- and 90 degrees MV-CBCT (180 degrees kVMV-CBCT) were acquired with different parameters. Image quality and imaging dose were evaluated and compared to kV-imaging. Hardware-based kVMV synchronization resulted in artifact-free projections. A combined 180 degrees kVMV-CBCT scan with a total MV-dose of 5 monitor units was acquired in 15 s and with sufficient image quality. The resolution was 5-6 line pairs cm(-1) (Catphan phantom). The combined kVMV-scan dose was equivalent to a kV-radiation scan dose of approximately 33 mGy. kVMV-CBCT based on a standard linac is promising and can provide ultra-fast online volume image guidance with low imaging dose and sufficient image quality for fast and accurate patient positioning for patients with lung cancer under breathhold.

Entities:  

Mesh:

Year:  2010        PMID: 20616405     DOI: 10.1088/0031-9155/55/15/001

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  14 in total

1.  [Image-guided radiation therapy. Paradigm change in radiation therapy].

Authors:  F Wenz; C Belka; M Reiser; S O Schönberg
Journal:  Radiologe       Date:  2012-03       Impact factor: 0.635

2.  [Image-guided radiation therapy].

Authors:  J Boda-Heggemann; M Guckenberger; U Ganswindt; C Belka; H Wertz; M Blessing; F Wenz; M Fuss; F Lohr
Journal:  Radiologe       Date:  2012-03       Impact factor: 0.635

Review 3.  kV cone-beam CT-based IGRT: a clinical review.

Authors:  Judit Boda-Heggemann; Frank Lohr; Frederik Wenz; Michael Flentje; Matthias Guckenberger
Journal:  Strahlenther Onkol       Date:  2011-04-26       Impact factor: 3.621

Review 4.  Are three doses of stereotactic ablative radiotherapy (SABR) more effective than 30 doses of conventional radiotherapy?

Authors:  Anna O Simeonova; Katharina Fleckenstein; Hansjörg Wertz; Anian Frauenfeld; Judit Boda-Heggemann; Frank Lohr; Frederik Wenz
Journal:  Transl Lung Cancer Res       Date:  2012-03

5.  Towards clinical implementation of ultrafast combined kV-MV CBCT for IGRT of lung cancer : Evaluation of registration accuracy based on phantom study.

Authors:  Anna Arns; Manuel Blessing; Jens Fleckenstein; Dzmitry Stsepankou; Judit Boda-Heggemann; Anna Simeonova-Chergou; Jürgen Hesser; Frank Lohr; Frederik Wenz; Hansjörg Wertz
Journal:  Strahlenther Onkol       Date:  2016-02-10       Impact factor: 3.621

6.  Physical density estimations of single- and dual-energy CT using material-based forward projection algorithm: a simulation study.

Authors:  Kai-Wen Li; Daiyu Fujiwara; Akihiro Haga; Huisheng Liu; Li-Sheng Geng
Journal:  Br J Radiol       Date:  2021-09-29       Impact factor: 3.039

7.  Concurrent kilovoltage CBCT imaging and megavoltage beam delivery: suppression of cross-scatter with 2D antiscatter grids and grid-based scatter sampling.

Authors:  Farhang Bayat; Mohamed Elsayed Eldib; Brian Kavanagh; Moyed Miften; Cem Altunbas
Journal:  Phys Med Biol       Date:  2022-08-09       Impact factor: 4.174

8.  Megavoltage cross-scatter rejection and correction using 2D antiscatter grids in kilovoltage CBCT imaging.

Authors:  Farhang Bayat; Mohamed Elsayed Eldib; Cem Altunbas
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2022-04-04

9.  Megavoltage cone beam computed tomography: Commissioning and evaluation of patient dose.

Authors:  Hassan S Abou-Elenein; Ehab M Attalla; H Ammar; Ismail Eldesoky; Mohamed Farouk; Mohamed S Zaghloul
Journal:  J Med Phys       Date:  2011-10

Review 10.  Applications of linac-mounted kilovoltage Cone-beam Computed Tomography in modern radiation therapy: A review.

Authors:  Kavitha Srinivasan; Mohammad Mohammadi; Justin Shepherd
Journal:  Pol J Radiol       Date:  2014-07-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.