| Literature DB >> 27029933 |
Yao Chen1, Xiaobo Xia2.
Abstract
BACKGROUND: The purpose of this study was to compare corneal power and horizontal corneal diameter (white-to-white [WTW] distance) readings obtained by the Orbscan II topographer and the iTrace aberrometer.Entities:
Keywords: Keratometry; Orbscan II; White-to-white distance; iTrace
Mesh:
Year: 2016 PMID: 27029933 PMCID: PMC4815140 DOI: 10.1186/s12886-016-0210-8
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Comparison of the corneal power and WTW distance values measured by the 2 devices
| Parameters | Orbscan II | iTrace |
|
|---|---|---|---|
| Kf (D) | 42.56 ± 1.34 | 42.06 ± 1.34 | <0.0001 |
| Ks (D) | 43.76 ± 1.58 | 43.21 ± 1.57 | <0.0001 |
| WTW (mm) | 11.57 ± 0.34 | 11.33 ± 0.36 | <0.0001 |
Data are expressed as mean ± SD; apaired t test <0.05 considered significant; Kf = flat axis power; Ks = steep axis power, D = diopter
Fig. 1Kf measurements by the Orbsan II versus the iTrace. a Bland-Altman plot showing the mean difference and the limits of agreement. b Scatter diagram and Pearson correlation analysis
Fig. 2Ks measurements by the Orbsan II versus the iTrace. a Bland-Altman plot. b Scatter diagram and Pearson correlation analysis
Fig. 3The WTW distance measurements by the Orbsan II versus the iTrace. a Bland-Altman plot. b Scatter diagram and Pearson correlation analysis
Comparison of the corneal power and WTW distance measurements in age <19 years old and age ≥19 years old groups
| groups | parameters | Orbscan II | iTrace |
| 95 % LoA |
|---|---|---|---|---|---|
| age <19 | Kf (D) | 42.29 ± 1.38 | 41.84 ± 1.45 | <0.0001 | −0.16,1.07 |
| years old | Ks (D) | 43.47 ± 1.65 | 42.95 ± 1.71 | <0.0001 | −0.08,1.11 |
| ( | WTW (mm) | 11.60 ± 0.29 | 11.37 ± 0.30 | <0.0001 | 0.00,0.45 |
| age ≥19 | Kf (D) | 42.74 ± 1.30 | 42.20 ± 1.26 | <0.0001 | −0.23,1.30 |
| years old | Ks (D) | 43.96 ± 1.51 | 43.39 ± 1.45 | <0.0001 | −0.19,1.34 |
| ( | WTW (mm) | 11.55 ± 0.36 | 11.30 ± 0.39 | <0.0001 | 0.01,0.48 |
Data are expressed as mean ± SD; apaired t test <0.05 considered significant; Kf = flat axis power; Ks = steep axis power, D = diopter, LoA = limits of agreement
Comparison of the corneal power and WTW distance measurements in male and female subjects
| gender | parameters | Orbscan II | iTrace |
| 95 % LoA |
|---|---|---|---|---|---|
| Male | Kf (D) | 42.12 ± 1.39 | 41.63 ± 1.37 | <0.0001 | −0.28,1.26 |
| ( | Ks (D) | 43.34 ± 1.63 | 42.79 ± 1.65 | <0.0001 | −0.17,1.26 |
| WTW (mm) | 11.64 ± 0.32 | 11.40 ± 0.33 | <0.0001 | 0.01,0.47 | |
| Female | Kf (D) | 43.02 ± 1.14 | 42.51 ± 1.16 | <0.0001 | −0.13,1.16 |
| ( | Ks (D) | 44.20 ± 1.41 | 43.65 ± 1.36 | <0.0001 | −0.14,1.24 |
| WTW (mm) | 11.49 ± 0.34 | 11.25 ± 0.37 | <0.0001 | −0.01,0.48 |
Data are expressed as mean ± SD; apaired t test <0.05 considered significant; Kf = flat axis power; Ks = steep axis power, D = diopter, LoA = limits of agreement