Comparison of keratometry measurements using the Pentacam HR, the Orbscan IIz, and the TMS-4 topographer.

Zahra Tajbakhsh1, Ramin Salouti, Mohammad H Nowroozzadeh, Mohammad Aghazadeh-Amiri, Seyedmehdi Tabatabaee, Mohammad Zamani.   

Abstract

PURPOSE: To compare and evaluate the agreement between keratometry readings using three devices: the Pentacam HR, the Orbscan IIz, and the TMS-4 topographer.
METHODS: In this prospective comparative study, 115 eyes from 115 healthy refractive surgery candidates were sequentially examined using three devices. Fourier analysis was used to evaluate corneal astigmatism in these subjects. The outcome measures comprised steep and flat keratometry readings, corneal astigmatism, the mean keratometry reading, and the J0 and J45 components. Repeated-measures analysis of variance (RM-anova) and Bland-Altman plots with 95% limits of agreement were used to evaluate the difference between measurements.
RESULTS: The mean keratometry readings were 43.89 ± 1.51, 43.52 ± 1.48, and 43.79 ± 1.50 D using the TMS-4 topographer, the Orbscan IIz, and the Pentacam HR, respectively (p < 0.001; RM-anova). The mean keratometric astigmatism was 1.27 ± 0.87, 1.20 ± 0.86, and 1.21 ± 0.82 D using the TMS-4 topographer, the Orbscan IIz, and the Pentacam HR, respectively (p = 0.007; RM-anova). The J0 measurements were higher using the TMS-4 topographer compared with the Orbscan IIz (p = 0.021; Bonferroni test). The mean, steep, and flat keratometry data were in better agreement using the TMS-4 topographer and the Pentacam HR. In general, the Orbscan IIz device measured lower values than either the TMS-4 topographer or the Pentacam HR. However, the corneal astigmatism measurements and the J0 component showed the best agreement using the TMS-4 topographer and the Orbscan IIz.
CONCLUSIONS: The TMS-4 topographer and the Pentacam HR produced similar readings and can be used interchangeably to measure simulated keratometry values in young, healthy eyes. To measure corneal astigmatism, the TMS-4 topographer and the Orbscan IIz produced values that were similar and could be used interchangeably. Ophthalmic & Physiological Optics
© 2012 The College of Optometrists.

Entities:  

Mesh:

Year:  2012        PMID: 23057566     DOI: 10.1111/j.1475-1313.2012.00942.x

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  13 in total

1.  Difference in normal corneal thickness and curvature between Mongolian and Han nationalities.

Authors:  Hai-Xia Zhao; Li Zhang; Wen-Ying Guan
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

2.  A novel Hartman Shack-based topography system: repeatability and agreement for corneal power with Scheimpflug+Placido topographer and rotating prism auto-keratorefractor.

Authors:  Gaurav Prakash; Dhruv Srivastava; Sounak Choudhuri
Journal:  Int Ophthalmol       Date:  2015-03-28       Impact factor: 2.031

3.  Characteristic of entire corneal topography and tomography for the detection of sub-clinical keratoconus with Zernike polynomials using Pentacam.

Authors:  Zhe Xu; Weibo Li; Jun Jiang; Xiran Zhuang; Wei Chen; Mei Peng; Jianhua Wang; Fan Lu; Meixiao Shen; Yuanyuan Wang
Journal:  Sci Rep       Date:  2017-11-28       Impact factor: 4.379

4.  Repeatability and Agreement of Central Corneal Thickness and Keratometry Measurements between Four Different Devices.

Authors:  Laszlo Kiraly; Jana Stange; Kathleen S Kunert; Saadettin Sel
Journal:  J Ophthalmol       Date:  2017-03-05       Impact factor: 1.909

5.  The distribution of orbscan indices in young population.

Authors:  Hassan Hashemi; Mehdi Khabazkhoob; Negareh Yazdani; Hadi Ostadimoghaddam; Payam Nabovati; Raheleh Moravej; AbbasAli Yekta
Journal:  J Curr Ophthalmol       Date:  2016-08-31

6.  Improving precision for detecting change in the shape of the cornea in patients with keratoconus.

Authors:  Matthias Brunner; Gabriela Czanner; Riccardo Vinciguerra; Vito Romano; Sajjad Ahmad; Mark Batterbury; Claire Britten; Colin E Willoughby; Stephen B Kaye
Journal:  Sci Rep       Date:  2018-08-17       Impact factor: 4.379

7.  Repeatability of curvature measurements in central and paracentral corneal areas of keratoconus patients using Orbscan and Pentacam.

Authors:  Ali Mirzajani; Amir Asharlous; Parvin Kianpoor; Ebrahim Jafarzadehpur; Abbasali Yekta; Mehdi Khabazkhoob; Hassan Hashemi
Journal:  J Curr Ophthalmol       Date:  2019-01-23

8.  Performance of Zernike polynomials in reconstructing raw-elevation data captured by Pentacam HR, Medmont E300 and Eye Surface Profiler.

Authors:  Yueying Wei; Bernardo T Lopes; Ashkan Eliasy; Richard Wu; Arwa Fathy; Ahmed Elsheikh; Ahmed Abass
Journal:  Heliyon       Date:  2021-12-18

9.  Comparison of the Orbscan II topographer and the iTrace aberrometer for the measurements of keratometry and corneal diameter in myopic patients.

Authors:  Yao Chen; Xiaobo Xia
Journal:  BMC Ophthalmol       Date:  2016-03-31       Impact factor: 2.209

10.  Repeatability and agreement of five imaging systems for measuring anterior segment parameters in healthy eyes.

Authors:  Mukesh Kumar; Rohit Shetty; Chaitra Jayadev; Harsha L Rao; Debarun Dutta
Journal:  Indian J Ophthalmol       Date:  2017-04       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.