Evaluation of the practicality of optical biometry and applanation ultrasound in 253 eyes.

Mana Tehrani1, Frank Krummenauer, Eric Blom, H Burkhard Dick.   

Abstract

PURPOSE: To evaluate the percentage of eyes that could not be measured using optical biometry and ultrasound applanation and the reasons.
SETTING: Department of Ophthalmology, Johannes Gutenberg-University Hospital, Mainz, Germany.
METHODS: Optical biometry (IOLMaster, Carl Zeiss Meditec AG) and A-scan ultrasound biometry were performed consecutively in 253 eyes scheduled for cataract surgery the next day. Lens opacities were evaluated with the Opacity Lensmeter (Interzeag), and a slitlamp examination and measurement of visual acuity were performed. The 2 techniques were compared in terms of the rate of and reasons for primary measurement failure.
RESULTS: Measurement with the IOLMaster was not possible in 44 eyes (17%). Failed measurements were the result of a combination of low visual acuity and lens opacity in 45% of eyes, posterior subcapsular opacity in 25%, and macular disease in 7%. Measurement with ultrasound biometry was not possible in 10 eyes (4%); 7 eyes were filled with silicone oil and in 3 cases, the patient refused biometry.
CONCLUSIONS: Optical biometry allowed comfortable, noncontact, high-precision measurement in the optical axis. Uncorrected visual acuity and lens opacity were predictors of successful measurements. Eyes with dense cataract or poor visual acuity are better evaluated using ultrasound applanation.

Entities:  

Mesh:

Year:  2003        PMID: 12686242     DOI: 10.1016/s0886-3350(02)01740-6

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  20 in total

1.  A novel index for predicting intraocular pressure reduction following cataract surgery.

Authors:  S A Issa; J Pacheco; U Mahmood; J Nolan; S Beatty
Journal:  Br J Ophthalmol       Date:  2005-05       Impact factor: 4.638

2.  Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes.

Authors:  P Kunavisarut; P Poopattanakul; C Intarated; K Pathanapitoon
Journal:  Eye (Lond)       Date:  2012-08-10       Impact factor: 3.775

3.  Agreement analysis of LENSTAR with other techniques of biometry.

Authors:  S Jasvinder; T F Khang; K K S Sarinder; V P Loo; V Subrayan
Journal:  Eye (Lond)       Date:  2011-03-11       Impact factor: 3.775

4.  Ocular biometry through fully refocused steady-state magnetic resonance imaging sequence: reliability and agreement with the IOLMaster® 500.

Authors:  Lorenzo Ismael Perez-Sanchez; Julia Gutierrez-Vazquez; Maria Satrustegui-Lapetra; Francisco Ferreira-Manuel; Juan Jose Arevalo-Manso; Juan Jesus Gomez-Herrera; Juan Jose Criado-Alvarez
Journal:  Int Ophthalmol       Date:  2021-02-23       Impact factor: 2.031

5.  Biometry and refractive outcome of eyes filled with silicone oil by standardized echography and partial coherence interferometry.

Authors:  Johannes Nepp; Katharina Krepler; Kerstin Jandrasits; Wolfgang Hauff; Georg Hanselmayer; Michaela Velikay-Parel; Karl C Ossoinig; Andreas Wedrich
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-10-20       Impact factor: 3.117

6.  The effect of pupil dilation on AL-Scan biometric parameters.

Authors:  Ertuğrul Can; Mustafa Duran; Tuğba Çetinkaya; Nurşen Arıtürk
Journal:  Int Ophthalmol       Date:  2015-06-27       Impact factor: 2.031

7.  Accuracy of axial length measurements from immersion B-scan ultrasonography in highly myopic eyes.

Authors:  Qing-Hua Yang; Bing Chen; Guang-Hua Peng; Zhao-Hui Li; Yi-Fei Huang
Journal:  Int J Ophthalmol       Date:  2014-06-18       Impact factor: 1.779

8.  Agreement of IOL power and axial length obtained by IOLMaster 500 vs IOLMaster 500 with Sonolink connection.

Authors:  Sabong Srivannaboon; Chareenun Chirapapaisan; Patchara Nantasri; Mathinee Chongchareon; Pratuangsri Chonpimai
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-01-12       Impact factor: 3.117

9.  Axial Length Measurement Failure Rates with the IOLMaster and Lenstar LS 900 in Eyes with Cataract.

Authors:  Colm McAlinden; Qinmei Wang; Konrad Pesudovs; Xin Yang; Fangjun Bao; Ayong Yu; Shishi Lin; Yifan Feng; Jinhai Huang
Journal:  PLoS One       Date:  2015-06-10       Impact factor: 3.240

10.  Refractive predictability of partial coherence interferometry and factors that can affect it.

Authors:  Seung Mo Kim; Joohyun Choi; Sangkyung Choi
Journal:  Korean J Ophthalmol       Date:  2009-03-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.