Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers.

Shi-Ming Cheng1,2,3, Jia-Sheng Zhang1,2, Xu Shao1,2, Ze-Tong Wu1,2, Tian-Tian Li1,2, Po Wang1,2, Jun-Hai Lin1,2, A-Yong Yu4,5.   

Abstract

PURPOSE: To investigate the repeatability of Anterion and compare the agreement of ocular biometric measurements and predicted intraocular lens (IOL) powers with other three optical biometers.
METHODS: Flat keratometry (Kf), steep keratometry (Ks), J0 and 45 vectors, central cornea thickness (CCT), anterior chamber depth (ACD), and axial length (AL) from the Anterion, IOLMaster 700, Lenstar LS 900, and OA-2000 were recorded. The IOL powers were calculated with the Hoffer Q, Holladay 1, SRK/T, and Haigis formulas. The repeatability was evaluated using the within-subject standard deviation (Sw), repeatability coefficient (RC), coefficient of variation (COV), and intraclass correlation coefficient (ICC). Inter-device agreement between the four biometers was assessed with the 95% limits of agreement.
RESULTS: In total, 101 right eyes of 101 participants were enrolled. The Anterion showed good repeatability for all the included biometric parameters with all the CoV ≤ 0.30% and ICC ≥ 0.930 except for J45 with moderate repeatability (ICC was 0.849). Good agreement was found among the four devices for Kf, Ks, J0, J45, ACD, and AL. Generally, wide 95% LoA was found for the predicted IOL powers with the four IOL calculation formulas between the four devices.
CONCLUSIONS: The Anterion showed good repeatability of biometric measurements for most parameters. Good agreement among the four optical biometers was achieved for all the parameters except for CCT and the predicted IOL power. The AL values exhibited the best repeatability with Anterion and the best agreement among the biometers in our study.
© 2022. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.

Entities:  

Keywords:  Agreement; Anterion; Anterior segment biometry; Axial length; Optical coherence tomographer; Repeatability

Mesh:

Year:  2022        PMID: 35171331     DOI: 10.1007/s00417-022-05579-9

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  27 in total

1.  Influence of operator experience on the performance of ultrasound biometry compared to optical biometry before cataract surgery.

Authors:  Oliver Findl; Katharina Kriechbaum; Stefan Sacu; Barbara Kiss; Kaija Polak; Johannes Nepp; Gebtraud Schild; Georg Rainer; Saskia Maca; Vanessa Petternel; Birgit Lackner; Wolfgang Drexler
Journal:  J Cataract Refract Surg       Date:  2003-10       Impact factor: 3.351

2.  Effect of Orthokeratology on Axial Length Elongation in Anisomyopic Children.

Authors:  Yu Zhang; Yueguo Chen
Journal:  Optom Vis Sci       Date:  2019-01       Impact factor: 1.973

3.  Central Corneal Thickness Increase Due to Stromal Thickening With Diabetic Peripheral Neuropathy Severity.

Authors:  Navasuja Kumar; Rodica Pop-Busui; David C Musch; David M Reed; Anna C Momont; Munira Hussain; Nilesh Raval; Sayoko E Moroi; Roni Shtein
Journal:  Cornea       Date:  2018-09       Impact factor: 2.651

4.  Central corneal thickness and glaucoma in East Asian people.

Authors:  Alexander C Day; David Machin; Tin Aung; Gus Gazzard; Rahat Husain; Paul T K Chew; Peng T Khaw; Steve K L Seah; Paul J Foster
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-10-28       Impact factor: 4.799

5.  Evaluation of 2 new optical biometry devices and comparison with the current gold standard biometer.

Authors:  Yen-An Chen; Nino Hirnschall; Oliver Findl
Journal:  J Cataract Refract Surg       Date:  2011-01-17       Impact factor: 3.351

6.  Evaluation of the Lenstar LS 900 non-contact biometer.

Authors:  L P J Cruysberg; M Doors; F Verbakel; T T J M Berendschot; J De Brabander; R M M A Nuijts
Journal:  Br J Ophthalmol       Date:  2009-08-18       Impact factor: 4.638

7.  Corneal thickness in keratoconus: comparing optical, ultrasound, and optical coherence tomography pachymetry.

Authors:  Debarun Dutta; Harsha L Rao; Uday K Addepalli; Pravin K Vaddavalli
Journal:  Ophthalmology       Date:  2012-11-21       Impact factor: 12.079

8.  Validity and repeatability of the Aladdin ocular biometer.

Authors:  Priyanka Mandal; Emma J Berrow; Shehzad A Naroo; James S Wolffsohn; Detlef Uthoff; Detlef Holland; Sunil Shah
Journal:  Br J Ophthalmol       Date:  2013-11-13       Impact factor: 4.638

9.  Axial Length Measurement Failure Rates with the IOLMaster and Lenstar LS 900 in Eyes with Cataract.

Authors:  Colm McAlinden; Qinmei Wang; Konrad Pesudovs; Xin Yang; Fangjun Bao; Ayong Yu; Shishi Lin; Yifan Feng; Jinhai Huang
Journal:  PLoS One       Date:  2015-06-10       Impact factor: 3.240

10.  Axial length growth and the risk of developing myopia in European children.

Authors:  Jan Willem Lodewijk Tideman; Jan Roelof Polling; Johannes R Vingerling; Vincent W V Jaddoe; Cathy Williams; Jeremy A Guggenheim; Caroline C W Klaver
Journal:  Acta Ophthalmol       Date:  2017-12-19       Impact factor: 3.761

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.