Jiaqing Zhang1, Zhenzhen Liu1, Xiaozhang Qiu1, Ling Jin1, Lanhua Wang1, Guangming Jin1, Wei Wang1, Xuhua Tan1, Lixia Luo1, Yizhi Liu1.
Abstract
Purpose: The purpose of this study was to investigate the change in axial length (AL) after cataract surgery measured by swept source optical coherence tomography (IOLMaster 700), and explore ways to eliminate this AL measurement error in pseudophakic eyes.Entities:
Mesh:
Year: 2021 PMID: 34029364 PMCID: PMC8161700 DOI: 10.1167/tvst.10.6.29
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Demographic and Clinical Characteristics of Participants
| Parameter | Operated Eye | Fellow Eye |
|---|---|---|
| No. of eyes | 305 | |
| Age, year | 65.97 ± 13.39 | |
| Male, | 110 (36.07) | |
| Diabetes, | 54 (17.70) | |
| Hypertension, | 119 (39.02) | |
| Pre-operative BCVA, logMAR | 1.00 ± 0.59 | 0.47 ± 0.40 |
| Axial length, mm | 24.27 ± 2.47 | 24.00 ± 1.85 |
| Corneal astigmatism, D | 0.46 ± 1.22 | 0.60 ± 0.93 |
| ACD, mm | 3.10 ± 0.49 | 3.08 ± 0.43 |
| LT, mm | 4.51 ± 0.52 | 4.55 ± 0.52 |
| CD, mm | 11.79 ± 0.49 | 11.85 ± 0.48 |
BCVA = best corrected vision acuity; logMAR = logarithm of the minimum angle resolution; mm = millimeter; D = diopter; ACD = anterior chamber depth, as measured from corneal epithelium to lens; LT = lens thickness; CD = cornea diameter.
Axial Length Change After Cataract Surgery Measured by IOLMaster 700
| Operated Eye | |||||||
|---|---|---|---|---|---|---|---|
| Parameter | Total | 509M | PCB00 | MX60 | SN60WF |
| Fellow Eye |
| No. of eyes | 305 | 51 | 76 | 83 | 95 | 305 | |
| Pre-operative AL, mm | 24.27 ± 2.47 | 24.57 ± 2.36 | 24.15 ± 2.35 | 24.69 ± 2.84 | 23.84 ± 2.22 | 24.00 ± 1.85 | |
| Postoperative AL, mm | |||||||
| Pseudophakic mode | 24.17 ± 2.47 | 24.47 ± 2.32 | 24.03 ± 2.36 | 24.59 ± 2.86 | 23.76 ± 2.22 | ||
| AL change | −0.10 ± 0.09 | −0.11 ± 0.08 | −0.12 ± 0.08 | −0.10 ± 0.12 | −0.08 ± 0.07 | 0.107 | |
|
| <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | ||
| Phakic mode | 24.06 ± 2.47 | 23.38 ± 2.36 | 23.92 ± 2.35 | 24.48 ± 2.84 | 23.65 ± 2.23 | 24.01 ± 1.85 | |
| AL change | −0.21 ± 0.11 | −0.20 ± 0.05 | −0.23 ± 0.11 | −0.22 ± 0.17 | −0.18 ± 0.07 | 0.110 | 0.002 ± 0.03 |
|
| <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | 0.305 | |
| Aphakic mode | 24.28 ± 2.47 | 24.57 ± 2.29 | 24.14 ± 2.36 | 24.70 ± 2.86 | 23.86 ± 2.22 | ||
| AL change | −0.01 ± 0.10 | −0.001 ± 0.13 | 0.01 ± 0.08 | 0.01 ± 0.11 | 0.03 ± 0.07 | 0.107 | |
|
| 0.264 | 0.993 | 0.123 | 0.541 | <0.001 | ||
AL = axial length; mm = millimeter.
Statistical significance (P < 0.05).
Paired t-test.
Analysis of variance (ANOVA).
AL change = postoperative AL – pre-operative AL.
Figure 1.Scatterplots indicating the correlation between preoperative and postoperative axial length (AL) of three measurement modes: (A) pseudophakic mode; (B) phakic mode; and (C) aphakic mode.
Figure 2.Bland-Altman plots indicating the agreement between preoperative and postoperative axial length (AL) of three measurement modes: (A) pseudophakic mode; (B) phakic mode; and (C) aphakic mode.
Correction Factor of Axial Length Measurement for Pseudophakic Eye in Subgroups
| IOL Type | Manufacturer | Material | Haptics | Refractive Index | Correction Factor Δ (mm) |
|---|---|---|---|---|---|
| 509M | Carl Zeiss Meditec | Hydrophilic acrylic (25 %) with hydrophobic surface properties | Plate | 1.46 | 0.11 |
| PCB00 | Abbott Medical Optics | Hydrophobic acrylic | C-loop | 1.47 | 0.12 |
| MX60 | Bausch & Lomb | Hydrophobic acrylic | C-loop | 1.53 | 0.10 |
| SN60WF | Alcon | Hydrophobic acrylic | C-loop | 1.55 | 0.08 |
IOL = intraocular lens.