Model-based iterative reconstruction for flat-panel cone-beam CT with focal spot blur, detector blur, and correlated noise.

Steven Tilley1, Jeffrey H Siewerdsen, J Webster Stayman.   

Abstract

While model-based reconstruction methods have been successfully applied to flat-panel cone-beam CT (FP-CBCT) systems, typical implementations ignore both spatial correlations in the projection data as well as system blurs due to the detector and focal spot in the x-ray source. In this work, we develop a forward model for flat-panel-based systems that includes blur and noise correlation associated with finite focal spot size and an indirect detector (e.g. scintillator). This forward model is used to develop a staged reconstruction framework where projection data are deconvolved and log-transformed, followed by a generalized least-squares reconstruction that utilizes a non-diagonal statistical weighting to account for the correlation that arises from the acquisition and data processing chain. We investigate the performance of this novel reconstruction approach in both simulated data and in CBCT test-bench data. In comparison to traditional filtered backprojection and model-based methods that ignore noise correlation, the proposed approach yields a superior noise-resolution tradeoff. For example, for a system with 0.34 mm FWHM scintillator blur and 0.70 FWHM focal spot blur, using the correlated noise model instead of an uncorrelated noise model increased resolution by 42% (with variance matched at 6.9  ×  10(-8) mm(-2)). While this advantage holds across a wide range of systems with differing blur characteristics, the improvements are greatest for systems where source blur is larger than detector blur.

Entities:  

Mesh:

Year:  2015        PMID: 26649783      PMCID: PMC5011041          DOI: 10.1088/0031-9155/61/1/296

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  29 in total

1.  Maximum-likelihood transmission image reconstruction for overlapping transmission beams.

Authors:  D F Yu; J A Fessler; E P Ficaro
Journal:  IEEE Trans Med Imaging       Date:  2000-11       Impact factor: 10.048

2.  Microcalcification detection using cone-beam CT mammography with a flat-panel imager.

Authors:  Xing Gong; Aruna A Vedula; Stephen J Glick
Journal:  Phys Med Biol       Date:  2004-06-07       Impact factor: 3.609

3.  Penalized-likelihood sinogram restoration for computed tomography.

Authors:  Patrick J La Rivière; Junguo Bian; Phillip A Vargas
Journal:  IEEE Trans Med Imaging       Date:  2006-08       Impact factor: 10.048

4.  A three-dimensional statistical approach to improved image quality for multislice helical CT.

Authors:  Jean-Baptiste Thibault; Ken D Sauer; Charles A Bouman; Jiang Hsieh
Journal:  Med Phys       Date:  2007-11       Impact factor: 4.071

5.  A method for measuring the presampled MTF of digital radiographic systems using an edge test device.

Authors:  E Samei; M J Flynn; D A Reimann
Journal:  Med Phys       Date:  1998-01       Impact factor: 4.071

6.  3D forward and back-projection for X-ray CT using separable footprints.

Authors:  Yong Long; Jeffrey A Fessler; James M Balter
Journal:  IEEE Trans Med Imaging       Date:  2010-06-07       Impact factor: 10.048

7.  Noise correlation in CBCT projection data and its application for noise reduction in low-dose CBCT.

Authors:  Hua Zhang; Luo Ouyang; Jianhua Ma; Jing Huang; Wufan Chen; Jing Wang
Journal:  Med Phys       Date:  2014-03       Impact factor: 4.071

8.  Proximal ADMM for multi-channel image reconstruction in spectral X-ray CT.

Authors:  Alex Sawatzky; Qiaofeng Xu; Carsten O Schirra; Mark A Anastasio
Journal:  IEEE Trans Med Imaging       Date:  2014-04-30       Impact factor: 10.048

9.  Generalized Least-Squares CT Reconstruction with Detector Blur and Correlated Noise Models.

Authors:  J Webster Stayman; Wojciech Zbijewski; Steven Tilley; Jeffrey Siewerdsen
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19

10.  Correction for collimator-detector response in SPECT using point spread function template.

Authors:  Se Young Chun; Jeffrey A Fessler; Yuni K Dewaraja
Journal:  IEEE Trans Med Imaging       Date:  2012-10-18       Impact factor: 10.048

View more
  18 in total

1.  Penalized-Likelihood Reconstruction With High-Fidelity Measurement Models for High-Resolution Cone-Beam Imaging.

Authors:  Steven Tilley; Matthew Jacobson; Qian Cao; Michael Brehler; Alejandro Sisniega; Wojciech Zbijewski; J Webster Stayman
Journal:  IEEE Trans Med Imaging       Date:  2018-04       Impact factor: 10.048

2.  Modeling Shift-Variant X-Ray Focal Spot Blur for High-Resolution Flat-Panel Cone-Beam CT.

Authors:  Steven Tilley; Wojciech Zbijewski; Jeffrey H Siewerdsen; J Webster Stayman
Journal:  Conf Proc Int Conf Image Form Xray Comput Tomogr       Date:  2016-07

3.  Propagation-based phase-contrast tomography of a guinea pig inner ear with cochlear implant using a model-based iterative reconstruction algorithm.

Authors:  Lorenz Hehn; Regine Gradl; Andrej Voss; Benedikt Günther; Martin Dierolf; Christoph Jud; Konstantin Willer; Sebastian Allner; Jörg U Hammel; Roland Hessler; Kaye S Morgan; Julia Herzen; Werner Hemmert; Franz Pfeiffer
Journal:  Biomed Opt Express       Date:  2018-10-10       Impact factor: 3.732

4.  Known-Component Model-Based Material Decomposition for Dual Energy Imaging of Bone Compositions in the Presence of Metal Implant.

Authors:  S Z Liu; S Tilley; Q Cao; J H Siewerdsen; J W Stayman; W Zbijewski
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2019-05-28

5.  Modeling and evaluation of a high-resolution CMOS detector for cone-beam CT of the extremities.

Authors:  Qian Cao; Alejandro Sisniega; Michael Brehler; J Webster Stayman; John Yorkston; Jeffrey H Siewerdsen; Wojciech Zbijewski
Journal:  Med Phys       Date:  2017-11-27       Impact factor: 4.071

6.  Investigation of the efficacy of a data-driven CT artifact correction scheme for sparse and truncated projection data for intracranial hemorrhage diagnosis.

Authors:  Alexander R Podgorsak; Mohammad Mahdi Shiraz Bhurwani; Ciprian N Ionita
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2021-02-15

7.  Multiresolution iterative reconstruction in high-resolution extremity cone-beam CT.

Authors:  Qian Cao; Wojciech Zbijewski; Alejandro Sisniega; John Yorkston; Jeffrey H Siewerdsen; J Webster Stayman
Journal:  Phys Med Biol       Date:  2016-10-03       Impact factor: 3.609

8.  Blind deconvolution in model-based iterative reconstruction for CT using a normalized sparsity measure.

Authors:  Lorenz Hehn; Steven Tilley; Franz Pfeiffer; J Webster Stayman
Journal:  Phys Med Biol       Date:  2019-10-31       Impact factor: 3.609

9.  Effect of source blur on digital breast tomosynthesis reconstruction.

Authors:  Jiabei Zheng; Jeffrey A Fessler; Heang-Ping Chan
Journal:  Med Phys       Date:  2019-10-20       Impact factor: 4.071

10.  High-resolution extremity cone-beam CT with a CMOS detector: Task-based optimization of scintillator thickness.

Authors:  Q Cao; M Brehler; A Sisniega; J W Stayman; J Yorkston; J H Siewerdsen; W Zbijewski
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2017-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.