Modeling and evaluation of a high-resolution CMOS detector for cone-beam CT of the extremities.

Qian Cao1, Alejandro Sisniega1, Michael Brehler1, J Webster Stayman1, John Yorkston2, Jeffrey H Siewerdsen1,3, Wojciech Zbijewski1.   

Abstract

PURPOSE: Quantitative assessment of trabecular bone microarchitecture in extremity cone-beam CT (CBCT) would benefit from the high spatial resolution, low electronic noise, and fast scan time provided by complementary <span class="Chemical">metal-oxide semiconductor (CMOS) x-ray detectors. We investigate the performance of CMOS sensors in extremity CBCT, in particular with respect to potential advantages of thin (<0.7 mm) scintillators offering higher spatial resolution.
METHODS: A cascaded systems model of a CMOS x-ray detector incorporating the effects of CsI:Tl scintillator thickness was developed. Simulation studies were performed using nominal extremity CBCT acquisition protocols (90 kVp, 0.126 mAs/projection). A range of scintillator thickness (0.35-0.75 mm), pixel size (0.05-0.4 mm), focal spot size (0.05-0.7 mm), magnification (1.1-2.1), and dose (15-40 mGy) was considered. The detectability index was evaluated for both CMOS and a-Si:H flat-panel detector (FPD) configurations for a range of imaging tasks emphasizing spatial frequencies associated with feature size aobj. Experimental validation was performed on a CBCT test bench in the geometry of a compact orthopedic CBCT system (SAD = 43.1 cm, SDD = 56.0 cm, matching that of the Carestream OnSight 3D system). The test-bench studies involved a 0.3 mm focal spot x-ray source and two CMOS detectors (Dalsa Xineos-3030HR, 0.099 mm pixel pitch) - one with the standard CsI:Tl thickness of 0.7 mm (C700) and one with a custom 0.4 mm thick scintillator (C400). Measurements of modulation transfer function (MTF), detective quantum efficiency (DQE), and CBCT scans of a cadaveric knee (15 mGy) were obtained for each detector.
RESULTS: Optimal detectability for high-frequency tasks (feature size of ~0.06 mm, consistent with the size of trabeculae) was ~4× for the C700 CMOS detector compared to the a-Si:H FPD at nominal system geometry of extremity CBCT. This is due to ~5× lower electronic noise of a CMOS sensor, which enables input quantum-limited imaging at smaller pixel size. Optimal pixel size for high-frequency tasks was <0.1 mm for a CMOS, compared to ~0.14 mm for an a-Si:H FPD. For this fine pixel pitch, detectability of fine features could be improved by using a thinner scintillator to reduce light spread blur. A 22% increase in detectability of 0.06 mm features was found for the C400 configuration compared to C700. An improvement in the frequency at 50% modulation (f50 ) of MTF was measured, increasing from 1.8 lp/mm for C700 to 2.5 lp/mm for C400. The C400 configuration also achieved equivalent or better DQE as C700 for frequencies above ~2 mm-1 . Images of cadaver specimens confirmed improved visualization of trabeculae with the C400 sensor.
CONCLUSIONS: The small pixel size of CMOS detectors yields improved performance in high-resolution extremity CBCT compared to a-Si:H FPDs, particularly when coupled with a custom 0.4 mm thick scintillator. The results indicate that adoption of a CMOS detector in extremity CBCT can benefit applications in quantitative imaging of trabecular microstructure in humans.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  CMOS detector; bone microstructure; cascaded systems modeling; cone-beam CT; optimization; orthopedic imaging

Mesh:

Substances:

Year:  2017        PMID: 29095489      PMCID: PMC5774240          DOI: 10.1002/mp.12654

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  67 in total

Review 1.  Subchondral bone changes in hand and knee osteoarthritis detected by radiography.

Authors:  Christopher Buckland-Wright
Journal:  Osteoarthritis Cartilage       Date:  2004       Impact factor: 6.576

2.  Fixed-flexion radiography of the knee provides reproducible joint space width measurements in osteoarthritis.

Authors:  Manish Kothari; Ali Guermazi; Gabriele von Ingersleben; Yves Miaux; Martine Sieffert; Jon E Block; Randall Stevens; Charles G Peterfy
Journal:  Eur Radiol       Date:  2004-05-19       Impact factor: 5.315

3.  Screen optics effects on detective quantum efficiency in digital radiography: zero-frequency effects.

Authors:  A R Lubinsky; Wei Zhao; Goran Ristic; J A Rowlands
Journal:  Med Phys       Date:  2006-05       Impact factor: 4.071

4.  MANTIS: combined x-ray, electron and optical Monte Carlo simulations of indirect radiation imaging systems.

Authors:  Aldo Badano; Josep Sempau
Journal:  Phys Med Biol       Date:  2006-03-01       Impact factor: 3.609

5.  Evolution of spatial resolution in breast CT at UC Davis.

Authors:  Peymon M Gazi; Kai Yang; George W Burkett; Shadi Aminololama-Shakeri; J Anthony Seibert; John M Boone
Journal:  Med Phys       Date:  2015-04       Impact factor: 4.071

6.  Optimal short scan convolution reconstruction for fanbeam CT.

Authors:  D L Parker
Journal:  Med Phys       Date:  1982 Mar-Apr       Impact factor: 4.071

7.  Validation of CBCT for the computation of textural biomarkers.

Authors:  Beatriz Paniagua; Antonio Carlos Ruellas; Erika Benavides; Steve Marron; Larry Woldford; Lucia Cevidanes
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-03-17

8.  Application of micro-CT assessment of 3-D bone microstructure in preclinical and clinical studies.

Authors:  Yebin Jiang; Jenny Zhao; Er-Yuan Liao; Ru-Chun Dai; Xian-Ping Wu; Harry K Genant
Journal:  J Bone Miner Metab       Date:  2005       Impact factor: 2.626

9.  Characterization of 3D joint space morphology using an electrostatic model (with application to osteoarthritis).

Authors:  Qian Cao; Gaurav Thawait; Grace J Gang; Wojciech Zbijewski; Thomas Reigel; Tyler Brown; Brian Corner; Shadpour Demehri; Jeffrey H Siewerdsen
Journal:  Phys Med Biol       Date:  2015-01-09       Impact factor: 3.609

Review 10.  Trabecular bone score (TBS) as a new complementary approach for osteoporosis evaluation in clinical practice.

Authors:  N C Harvey; C C Glüer; N Binkley; E V McCloskey; M-L Brandi; C Cooper; D Kendler; O Lamy; A Laslop; B M Camargos; J-Y Reginster; R Rizzoli; J A Kanis
Journal:  Bone       Date:  2015-05-16       Impact factor: 4.398

View more
  7 in total

1.  Mobile C-Arm with a CMOS detector: Technical assessment of fluoroscopy and Cone-Beam CT imaging performance.

Authors:  Niral M Sheth; Wojciech Zbijewski; Matthew W Jacobson; Godwin Abiola; Gerhard Kleinszig; Sebastian Vogt; Stefan Soellradl; Jens Bialkowski; William S Anderson; Clifford R Weiss; Greg M Osgood; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2018-11-13       Impact factor: 4.071

2.  Predicting image properties in penalized-likelihood reconstructions of flat-panel CBCT.

Authors:  Wenying Wang; Grace J Gang; Jeffrey H Siewerdsen; J Webster Stayman
Journal:  Med Phys       Date:  2018-11-20       Impact factor: 4.071

3.  High-Resolution Extremity Cone-Beam CT with a CMOS Detector: Evaluation of a Clinical Prototype in Quantitative Assessment of Bone Microarchitecture.

Authors:  Q Cao; M Brehler; A Sisniega; S Tilley; M M Shiraz Bhruwani; J W Stayman; J Yorkston; J H Siewerdsen; W Zbijewski
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2018-03-09

4.  Quantitative Evaluation of Bone Microstructure using High-Resolution Extremity Cone-Beam CT with a CMOS Detector.

Authors:  S Subramanian; M Brehler; Q Cao; F J Quevedo Gonzalez; R E Breighner; J A Carrino; T Wright; J Yorkston; J H Siewerdsen; W Zbijewski
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2019-03-15

5.  A Prototype Intraoral Periapical Sensor with High Frame Rates for a 2.5D Periapical Radiography System.

Authors:  Che-Wei Liao; Ker-Jer Huang; Jyh-Cheng Chen; Chih-Wei Kuo; Yin-Yi Wu; Jui-Ting Hsu
Journal:  Appl Bionics Biomech       Date:  2019-04-24       Impact factor: 1.781

6.  Effect of Filtered Back-Projection Filters to Low-Contrast Object Imaging in Ultra-High-Resolution (UHR) Cone-Beam Computed Tomography (CBCT).

Authors:  Sunghoon Choi; Chang-Woo Seo; Bo Kyung Cha
Journal:  Sensors (Basel)       Date:  2020-11-10       Impact factor: 3.576

7.  Quantitative Cone-Beam CT of Bone Mineral Density Using Model-Based Reconstruction.

Authors:  Q Cao; A Sisniega; J W Stayman; J Yorkston; J H Siewerdsen; W Zbijewski
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2019-03-01
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.