Sarah Feger1, Matthias Rief1, Elke Zimmermann1, Peter Martus2, Joanne Désirée Schuijf3, Jörg Blobel3, Felicitas Richter4, Marc Dewey1.
Abstract
PURPOSE: The aim of this study was the systematic image quality evaluation of coronary CT angiography (CTA), reconstructed with the 3 different levels of adaptive iterative dose reduction (AIDR 3D) and compared to filtered back projection (FBP) with quantum denoising software (QDS).Entities:
Mesh:
Substances:
Year: 2015 PMID: 25945924 PMCID: PMC4422621 DOI: 10.1371/journal.pone.0125943
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Characteristics of the 30 Patients.
| Feature | |||
|---|---|---|---|
| Age | 64.1 | ±9.1 years | |
| Sex | Female | 15 | (50%) |
| Male | 15 | (50%) | |
| Abdominal circumference | 103.2 | ±11.8 cm | |
| Height | 168.4 | ±8.7 cm | |
| Weight | 81.9 | ±13.8 kg | |
| BMI | 28.8 | ±4.0 kg/m² | |
| Systolic blood pressure | 132.0 | ±11.3 mmHg | |
| Diastolic blood pressure | 79.3 | ±8.1 mmHg | |
| Arterial hypertension | 28 | (93%) | |
| Hyperlipidaemia | 23 | (77%) | |
| Cardiac insufficiency | 5 | (17%) | |
| Myocardial infarction | 15 | (50%) | |
| Smoking | 6 | (20%) | |
| Heart rate during scan | 56.2 | ±7.3 bpm | |
| Oral beta-blockers | 72.5 | ±41.2 mg | |
| Intravenous beta-blockers | 60 | ±112.5 mg | |
| Tube current | 341.7 | ±60.0 mA | |
| Radiation dose | 3.2 | ±2.6 mSv |
Values are given as arithmetic mean ± standard deviation (SD) or number of patients (%)
a Myocardial infarction dated back more than 48 hours
Mean age was 64 years. The average HR was <60 bpm. 25 patients received oral administration of approximately 73 mg atenolol, and 8 of these patients received additional intravenous administration of 60 mg esmolol. Average tube current was 342 mA with mean radiation dose of 3 mSv for CTA.
Comparison of signal, noise, SNR and CNR between the reconstructions FBP/QDS, AIDR 3D mild, AIDR 3D standard and AIDR 3D strong.
| ANOVA | t-test | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| FBP/QDS | MILD | STD | STR | p | p1 | p2 | p3 | p4 | p5 | p6 | |||||
| Signal | 155.3 | (53.7) | 155.0 | (53.5) | 152.5 | (53.4) | 149.9 | (53.3) | <0.001 | 1.000 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 |
| Noise | 32.8 | (9.3) | 34.5 | (8.9) | 29.8 | (8.3) | 27.0 | (7.8) | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 |
| SNR | 10.4 | (3.0) | 9.4 | (3.0) | 11.1 | (3.7) | 12.6 | (4.4) | <0.001 | 0.001 | 0.001 | <0.001 | <0.001 | <0.001 | <0.001 |
| CNR | 12.5 | (4.3) | 11.4 | (3.1) | 13.5 | (3.9) | 15.3 | (4.7) | <0.001 | 0.002 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 |
Values are given in arithmetic mean (SD); reconstruction with filtered back projection/ quantum denoising filtering system (FBP/QDS), adaptive iterative dose reduction (AIDR) 3D mild (MILD), standard (STD) and strong (STR); The values of all measurement points are summarised; First, Repeated Measures ANOVA overall analysis including every measurement point as dependent variable showed p<0.001 (p ANOVA) for signal, noise, SNR and also CNR. Subsequently, t-test was used as single test for the 4 quantitative parameters. Bonferroni correction was automatically performed for the multiple testing with 6 possibilities: p1 (FBP/QDS-AIDR 3D mild), p2 (FBP/QDS-AIDR 3D standard), p3 (FBP/QDS-AIDR 3D strong), p4 (AIDR 3D mild-AIDR 3D standard), p5 (AIDR 3D mild-AIDR 3D strong), p6 (AIDR 3D standard-AIDR 3D strong); signal presents the density in Hounsfield Units; noise is SD of the signal; SNR = signal in the vessel measurement point/noise in the aorta ascendens measurement point; CNR = (signal in the surrounding tissue of the vessel measurement point—signal in the vessel measurement point)/noise in the aorta ascendens measurement point.