| Literature DB >> 26599111 |
Hesong Shen1,2, Guochao Dai3, Mingyue Luo1, Chaijie Duan4, Wenli Cai5, Dan Liang1, Xinhua Wang1, Dongyun Zhu1, Wenru Li1, Jianping Qiu1.
Abstract
PURPOSE: To investigate image quality and radiation dose of CT coronary angiography (CTCA) scanned using automatic tube current modulation (ATCM) and reconstructed by strong adaptive iterative dose reduction three-dimensional (AIDR3D).Entities:
Mesh:
Substances:
Year: 2015 PMID: 26599111 PMCID: PMC4657884 DOI: 10.1371/journal.pone.0142185
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Patient demographics.
|
| 57: 27 |
|
| 58.0 ± 17.3 (30–80) |
|
| 60.3 ± 13.7 (51–78) |
|
| 21.1 ± 3.2 (17–27) |
|
| 58.4 ± 6.2 (52–65) |
|
| 84 |
|
| 52 |
|
| 13 |
|
| 9 |
|
| 6 |
|
| 4 |
Quantitative image quality of CTCA with ATCM and strong AIDR3D, standard AIDR3D and FBP.
| Item | Strong AIDR3D | Standard AIDR3D | FBP |
|
|
|---|---|---|---|---|---|
|
| 27.2±4.4 | 33.3±5.2 | 60.0±12.4 | <0.05 | <0.001 |
|
| 21.1±5.1 | 17.1±3.1 | 11.4±2.8 | <0.05 | <0.05 |
|
| 24.7±5.1 | 19.1±2.9 | 13.5±3.2 | <0.05 | <0.001 |
Noise = Image noise; SNR = Signal-to-noise ratio; CNR = Contrast-to-noise ratio. P1: Strong AIDR3D compared with standard AIDR3D; P2: Strong AIDR3D compared with FBP.
Segments of qualitative image quality of three parts CTCA with ATCM and strong AIDR3D, standard AIDR3D and FBP.
| Proximal CTCA | Middle CTCA | Distal CTCA | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Score | Strong | Standard | FBP | Strong | Standard | FBP | Strong | Standard | FBP |
|
| 0 | 1 | 12 | 0 | 3 | 47 | 2 | 4 | 75 |
|
| 0 | 3 | 107 | 18 | 29 | 284 | 24 | 38 | 231 |
|
| 39 | 38 | 213 | 102 | 99 | 173 | 99 | 95 | 112 |
|
| 297 | 294 | 4 | 384 | 373 | 0 | 295 | 283 | 2 |
|
| 336 | 336 | 336 | 504 | 504 | 504 | 420 | 420 | 420 |
Strong = Strong AIDR3D; Standard = Standard AIDR3D.
Fig 1CTCA images of left anterior descending artery.
A: Image with a qualitative quality score of 4 with ATCM and strong AIDR3D, excellent attenuation of vessel lumen and clear vessel wall definition with barely perceived image noise. B: Image with a qualitative quality score of 3 with ATCM and FBP, good attenuation of vessel lumen and well-maintained vessel wall definition with minimal image noise.
Fig 4CTCA images of right coronary artery.
A: Image with a qualitative quality score of 2 with ATCM and strong AIDR3D, some limitations in vessel wall definition and contrast resolution with moderate image noise that would affect diagnosis. B: Image with a qualitative quality score of 1 with ATCM and standard AIDR3D, impaired image quality with severe image noise.