Evaluation of heavily calcified vessels with coronary CT angiography: comparison of iterative and filtered back projection image reconstruction.

Matthias Renker1, John W Nance, U Joseph Schoepf, Terrence X O'Brien, Peter L Zwerner, Mathias Meyer, J Matthias Kerl, Ralf W Bauer, Christian Fink, Thomas J Vogl, Thomas Henzler.   

Abstract

PURPOSE: To prospectively compare traditional filtered back projection (FBP) and iterative image reconstruction for the evaluation of heavily calcified arteries with coronary computed tomography (CT) angiography.
MATERIALS AND METHODS: The study had institutional review board approval and was HIPAA compliant. Written informed consent was obtained from all patients. Fifty-five consecutive patients (35 men, 20 women; mean age, 58 years ± 12 [standard deviation]) with Agatston scores of at least 400 underwent coronary CT angiography and cardiac catheterization. Image data were reconstructed with both FBP and iterative reconstruction techniques with corresponding cardiac algorithms. Image noise and subjective image quality were compared. To objectively assess the effect of FBP and iterative reconstruction on blooming artifacts, volumes of circumscribed calcifications were measured with dedicated volume analysis software. FBP and iterative reconstruction series were independently evaluated for coronary artery stenosis greater than 50%, and their diagnostic accuracy was compared, with cardiac catheterization as the reference standard. Statistical analyses included paired t tests, Kruskal-Wallis analysis of variance, and a modified McNemar test.
RESULTS: Image noise measured significantly lower (P = .011-.035) with iterative reconstruction instead of FBP. Image quality was rated significantly higher (P = .031 and .042) with iterative reconstruction series than with FBP. Calcification volumes measured significantly lower (P = .019 and .026) with iterative reconstruction (44.3 mm(3) ± 64.7 and 46.2 mm(3) ± 68.8) than with FBP (54.5 mm(3) ± 69.5 and 56.3 mm(3) ± 72.5). Iterative reconstruction significantly improved some measures of per-segment diagnostic accuracy of coronary CT angiography for the detection of significant stenosis compared with FBP (accuracy: 95.9% vs 91.8%, P = .0001; specificity: 95.8% vs 91.2%, P = .0001; positive predictive value: 76.9% vs 61.1%, P = .0001).
CONCLUSION: Iterative reconstruction reduces image noise and blooming artifacts from calcifications, leading to improved diagnostic accuracy of coronary CT angiography in patients with heavily calcified coronary arteries. © RSNA, 2011.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21693660     DOI: 10.1148/radiol.11103574

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  50 in total

1.  Iterative reconstruction of dual-source coronary CT angiography: assessment of image quality and radiation dose.

Authors:  Eun-Ah Park; Whal Lee; Kwang Woo Kim; Kwang Gi Kim; Allmendinger Thomas; Jin Wook Chung; Jae Hyung Park
Journal:  Int J Cardiovasc Imaging       Date:  2011-12-21       Impact factor: 2.357

2.  CT evaluation of coronary artery stents with iterative image reconstruction: improvements in image quality and potential for radiation dose reduction.

Authors:  Ullrich Ebersberger; Francesco Tricarico; U Joseph Schoepf; Philipp Blanke; J Reid Spears; Garrett W Rowe; William T Halligan; Thomas Henzler; Fabian Bamberg; Alexander W Leber; Ellen Hoffmann; Paul Apfaltrer
Journal:  Eur Radiol       Date:  2012-07-10       Impact factor: 5.315

3.  Feasibility of low-concentration iodinated contrast medium with lower-tube-voltage dual-source CT aortography using iterative reconstruction: comparison with automatic exposure control CT aortography.

Authors:  Hee Jeong Shin; Song Soo Kim; Jae-Hwan Lee; Jae-Hyeong Park; Jin-Ok Jeong; Seon Ah Jin; Byung Seok Shin; Kyung-Sook Shin; Moonsang Ahn
Journal:  Int J Cardiovasc Imaging       Date:  2015-11-30       Impact factor: 2.357

Review 4.  MDCT evaluation of atherosclerotic coronary artery disease: what should radiologists know?

Authors:  Hye Rin Kim; Seung Min Yoo; Ji Young Rho; Hwa Yeon Lee; Charles S White
Journal:  Int J Cardiovasc Imaging       Date:  2014-04-01       Impact factor: 2.357

5.  The effect of iterative image reconstruction algorithms on the feasibility of automated plaque assessment in coronary CT angiography.

Authors:  Stefan B Puchner; Maros Ferencik; Mihaly Karolyi; Synho Do; Pal Maurovich-Horvat; Hans-Ulrich Kauczor; Udo Hoffmann; Christopher L Schlett
Journal:  Int J Cardiovasc Imaging       Date:  2013-08-30       Impact factor: 2.357

Review 6.  The challenge of coronary calcium on coronary computed tomographic angiography (CCTA) scans: effect on interpretation and possible solutions.

Authors:  Makmur Andrew; Hoe John
Journal:  Int J Cardiovasc Imaging       Date:  2015-09-25       Impact factor: 2.357

7.  Consistency of Renal Stone Volume Measurements Across CT Scanner Model and Reconstruction Algorithm Configurations.

Authors:  Alice E Huang; Juan C Montoya; Maria Shiung; Shuai Leng; Cynthia H McCollough
Journal:  AJR Am J Roentgenol       Date:  2017-04-12       Impact factor: 3.959

8.  Blooming Artifact Reduction in Coronary Artery Calcification by A New De-blooming Algorithm: Initial Study.

Authors:  Ping Li; Lei Xu; Lin Yang; Rui Wang; Jiang Hsieh; Zhonghua Sun; Zhanming Fan; Jonathon A Leipsic
Journal:  Sci Rep       Date:  2018-05-02       Impact factor: 4.379

9.  Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom study.

Authors:  Felix G Meinel; Christian Canstein; U Joseph Schoepf; Martin Sedlmaier; Bernhard Schmidt; Brett S Harris; Thomas G Flohr; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2014-05-10       Impact factor: 5.315

10.  Changes in measured size of atherosclerotic plaque calcifications in dual-energy CT of ex vivo carotid endarterectomy specimens: effect of monochromatic keV image reconstructions.

Authors:  Lorenzo Mannelli; Lee M Mitsumori; Marina Ferguson; Dongxiang Xu; Baocheng Chu; Kelley R Branch; William P Shuman; Chun Yuan
Journal:  Eur Radiol       Date:  2012-08-21       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.