Literature DB >> 11821203

Immersion A-scan compared with partial coherence interferometry: outcomes analysis.

Mark Packer1, I Howard Fine, Richard S Hoffman, Peggy G Coffman, Laurie K Brown.   

Abstract

PURPOSE: To compare 2 methods of axial length measurement, immersion ultrasonography and partial coherence interferometry, and to elucidate surgical outcomes based on immersion measurements.
SETTING: Oregon Eye Institute, Eugene, Oregon, USA.
METHODS: Axial length measurements in 50 cataractous eyes were obtained by optical biometry (IOLMaster, Zeiss Humphrey Systems) and immersion ultrasound (Axis II, Quantel Medical), and the results were compared. Intraocular lens (IOL) power calculations in the same eyes after cataract extraction and posterior chamber IOL implantation were evaluated retrospectively based on the postoperative spherical equivalent prediction error.
RESULTS: Immersion ultrasonography and partial coherence interferometry measurements correlated in a highly positive manner (correlation coefficient = 0.996). Outcomes analysis demonstrated 92.0% of eyes were within +/-0.5 diopter of emmetropia based on immersion axial length measurements.
CONCLUSION: Immersion ultrasonography provided highly accurate axial length measurements and permitted highly accurate IOL power calculations.

Entities:  

Mesh:

Year:  2002        PMID: 11821203     DOI: 10.1016/s0886-3350(01)01259-7

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  29 in total

1.  IOLMaster biometry: refractive results of 100 consecutive cases.

Authors:  H Eleftheriadis
Journal:  Br J Ophthalmol       Date:  2003-08       Impact factor: 4.638

2.  Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation.

Authors:  Mirjana Bjeloš Rončević; Mladen Bušić; Ivan Cima; Biljana Kuzmanović Elabjer; Damir Bosnar; Daliborka Miletić
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-09-18       Impact factor: 3.117

3.  Intraobserver and interobserver repeatability of ocular components measurement in cataract eyes using a new optical low coherence reflectometer.

Authors:  Mirjana Bjeloš Rončević; Mladen Bušić; Ivan Cima; Biljana Kuzmanović Elabjer; Damir Bosnar; Daliborka Miletić
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-10-28       Impact factor: 3.117

4.  The effect of head tilt on keratometric measurement using the IOLMaster.

Authors:  C Y Park; J R Do; S H Kim; C-Y Lim; R S Chuck
Journal:  Eye (Lond)       Date:  2013-09-20       Impact factor: 3.775

5.  Agreement analysis of LENSTAR with other techniques of biometry.

Authors:  S Jasvinder; T F Khang; K K S Sarinder; V P Loo; V Subrayan
Journal:  Eye (Lond)       Date:  2011-03-11       Impact factor: 3.775

Review 6.  Optimising biometry for best outcomes in cataract surgery.

Authors:  R Sheard
Journal:  Eye (Lond)       Date:  2013-12-06       Impact factor: 3.775

7.  Partial coherence interferometry versus immersion ultrasonography for axial length measurement in children.

Authors:  Phoebe D Lenhart; Amy K Hutchinson; Michael J Lynn; Scott R Lambert
Journal:  J Cataract Refract Surg       Date:  2010-12       Impact factor: 3.351

8.  Comparison of intraocular lens power prediction using immersion ultrasound and optical biometry with and without formula optimization.

Authors:  Gabor Nemeth; Attila Nagy; Andras Berta; Laszlo Modis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-04-13       Impact factor: 3.117

9.  Agreement of IOL power and axial length obtained by IOLMaster 500 vs IOLMaster 500 with Sonolink connection.

Authors:  Sabong Srivannaboon; Chareenun Chirapapaisan; Patchara Nantasri; Mathinee Chongchareon; Pratuangsri Chonpimai
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-01-12       Impact factor: 3.117

10.  Reproducibility of a long-range swept-source optical coherence tomography ocular biometry system and comparison with clinical biometers.

Authors:  Ireneusz Grulkowski; Jonathan J Liu; Jason Y Zhang; Benjamin Potsaid; Vijaysekhar Jayaraman; Alex E Cable; Jay S Duker; James G Fujimoto
Journal:  Ophthalmology       Date:  2013-06-04       Impact factor: 12.079

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.