Effects of reflector and crystal surface on the performance of a depth-encoding PET detector with dual-ended readout.

Silin Ren1, Yongfeng Yang1, Simon R Cherry1.   

Abstract

PURPOSE: Depth encoding detectors are required to improve the spatial resolution and spatial resolution uniformity of small animal positron emission tomography (PET) scanners, as well as dedicated breast and brain scanners. Depth of interaction (DOI) can be measured by using dual-ended readout of lutetium oxyorthosilicate (LSO) scintillator arrays with position-sensitive avalanche photodiodes. Inter-crystal reflectors and crystal surface treatments play important roles in determining the performance of dual-ended detectors. In this paper, the authors evaluated five LSO arrays made with three different intercrystal reflectors and with either polished or unpolished crystal surfaces.
METHODS: The crystal size in all arrays was 1.5 mm, which is typical of the detector size used in small animal and dedicated breast scanners. The LSO arrays were measured with dual-ended readout and were compared in terms of flood histogram, energy resolution, and DOI resolution performance.
RESULTS: The four arrays using enhanced specular reflector (ESR) and Toray reflector provided similar quality flood histograms and the array using Crystal Wrap reflector gave the worst flood histogram. The two arrays using ESR reflector provided the best energy resolution and the array using Crystal Wrap reflector yielded the worst energy resolution. All arrays except the polished ESR array provided good DOI resolution ranging from 1.9 mm to 2.9 mm. DOI resolution improved as the gradient in light collection efficiency with depth (GLCED) increased. The geometric mean energies were also calculated for these dual-ended readout detectors as an alternative to the conventional summed total energy. It was shown that the geometric mean energy is advantageous in that it provides more uniform photopeak amplitude at different depths for arrays with high GLCED, and is beneficial in event selection by allowing a fixed energy window independent of depth. A new method of DOI calculation that improved the linearity of DOI ratio vs depth and simplifies the DOI calibration procedure also was developed and tested.
CONCLUSIONS: The results of these studies provide useful guidance in selecting the proper reflectors and crystal surface treatments when LSO arrays are used for high-resolution PET applications in small animal scanners or dedicated breast and brain scanners.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24989406      PMCID: PMC4187348          DOI: 10.1118/1.4881097

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  14 in total

1.  Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging.

Authors:  Yuan-Chuan Tai; Ananya Ruangma; Douglas Rowland; Stefan Siegel; Danny F Newport; Patrick L Chow; Richard Laforest
Journal:  J Nucl Med       Date:  2005-03       Impact factor: 10.057

2.  Evaluation of high performance data acquisition boards for simultaneous sampling of fast signals from PET detectors.

Authors:  Martin S Judenhofer; Bernd J Pichler; Simon R Cherry
Journal:  Phys Med Biol       Date:  2005-01-07       Impact factor: 3.609

3.  Performance measurements of a depth-encoding PET detector module based on position-sensitive avalanche photodiode read-out.

Authors:  P A Dokhale; R W Silverman; K S Shah; R Grazioso; R Farrell; J Glodo; M A McClish; G Entine; V H Tran; S R Cherry
Journal:  Phys Med Biol       Date:  2004-09-21       Impact factor: 3.609

4.  Depth of interaction resolution measurements for a high resolution PET detector using position sensitive avalanche photodiodes.

Authors:  Yongfeng Yang; Purushottam A Dokhale; Robert W Silverman; Kanai S Shah; Mickel A McClish; Richard Farrell; Gerald Entine; Simon R Cherry
Journal:  Phys Med Biol       Date:  2006-04-10       Impact factor: 3.609

5.  Depth of interaction calibration for PET detectors with dual-ended readout by PSAPDs.

Authors:  Yongfeng Yang; Jinyi Qi; Yibao Wu; Sara St James; Richard Farrell; Purushottam A Dokhale; Kanai S Shah; Simon R Cherry
Journal:  Phys Med Biol       Date:  2008-12-19       Impact factor: 3.609

6.  Simulation of light transport in scintillators based on 3D characterization of crystal surfaces.

Authors:  Emilie Roncali; Simon R Cherry
Journal:  Phys Med Biol       Date:  2013-03-11       Impact factor: 3.609

7.  Tapered LSO arrays for small animal PET.

Authors:  Yongfeng Yang; Sara St James; Yibao Wu; Huini Du; Jinyi Qi; Richard Farrell; Purushottam A Dokhale; Kanai S Shah; Keith Vaigneur; Simon R Cherry
Journal:  Phys Med Biol       Date:  2010-11-30       Impact factor: 3.609

8.  Performance evaluation of the GE healthcare eXplore VISTA dual-ring small-animal PET scanner.

Authors:  Yuchuan Wang; Jurgen Seidel; Benjamin M W Tsui; Juan J Vaquero; Martin G Pomper
Journal:  J Nucl Med       Date:  2006-11       Impact factor: 10.057

9.  Performance evaluation of the inveon dedicated PET preclinical tomograph based on the NEMA NU-4 standards.

Authors:  Qinan Bao; Danny Newport; Mu Chen; David B Stout; Arion F Chatziioannou
Journal:  J Nucl Med       Date:  2009-02-17       Impact factor: 10.057

10.  Characterization of a high-resolution hybrid DOI detector for a dedicated breast PET/CT scanner.

Authors:  Felipe Godinez; Abhijit J Chaudhari; Yongfeng Yang; Richard Farrell; Ramsey D Badawi
Journal:  Phys Med Biol       Date:  2012-05-11       Impact factor: 3.609

View more
  17 in total

1.  A depth-of-interaction encoding PET detector module with dual-ended readout using large-area silicon photomultiplier arrays.

Authors:  Junwei Du; Xiaowei Bai; Simon R Cherry
Journal:  Phys Med Biol       Date:  2018-12-14       Impact factor: 3.609

2.  Improving timing performance of double-ended readout in TOF-PET detectors.

Authors:  L Guo; J Tian; P Chen; S E Derenzo; W-S Choong
Journal:  J Instrum       Date:  2020-01-02       Impact factor: 1.415

3.  An integrated model of scintillator-reflector properties for advanced simulations of optical transport.

Authors:  Emilie Roncali; Mariele Stockhoff; Simon R Cherry
Journal:  Phys Med Biol       Date:  2017-04-11       Impact factor: 3.609

4.  A Prototype High-Resolution Small-Animal PET Scanner Dedicated to Mouse Brain Imaging.

Authors:  Yongfeng Yang; Julien Bec; Jian Zhou; Mengxi Zhang; Martin S Judenhofer; Xiaowei Bai; Kun Di; Yibao Wu; Mercedes Rodriguez; Purushottam Dokhale; Kanai S Shah; Richard Farrell; Jinyi Qi; Simon R Cherry
Journal:  J Nucl Med       Date:  2016-03-24       Impact factor: 10.057

5.  Performance of a high-resolution depth-encoding PET detector module using linearly-graded SiPM arrays.

Authors:  Junwei Du; Xiaowei Bai; Alberto Gola; Fabio Acerbi; Alessandro Ferri; Claudio Piemonte; Yongfeng Yang; Simon R Cherry
Journal:  Phys Med Biol       Date:  2018-02-05       Impact factor: 3.609

6.  Optimization of a depth of interaction encoding PET block detector for a PET/MRI insert.

Authors:  Aaron R Selfridge; Simon R Cherry; Martin S Judenhofer
Journal:  Phys Med Biol       Date:  2018-12-06       Impact factor: 3.609

7.  Monte Carlo simulations of time-of-flight PET with double-ended readout: calibration, coincidence resolving times and statistical lower bounds.

Authors:  Stephen E Derenzo
Journal:  Phys Med Biol       Date:  2017-03-22       Impact factor: 3.609

Review 8.  Positron Emission Tomography: Current Challenges and Opportunities for Technological Advances in Clinical and Preclinical Imaging Systems.

Authors:  Juan José Vaquero; Paul Kinahan
Journal:  Annu Rev Biomed Eng       Date:  2015       Impact factor: 9.590

9.  Shared-photodetector readout to improve the sensitivity of positron emission tomography.

Authors:  Junwei Du; Peng Peng; Xiaowei Bai; Simon R Cherry
Journal:  Phys Med Biol       Date:  2018-10-09       Impact factor: 3.609

10.  A depth-encoding PET detector for high resolution PET using 1 mm SiPMs.

Authors:  Junwei Du; Xiaowei Bai; Simon R Cherry
Journal:  Phys Med Biol       Date:  2020-08-19       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.