Performance evaluation of the inveon dedicated PET preclinical tomograph based on the NEMA NU-4 standards.

Qinan Bao1, Danny Newport, Mu Chen, David B Stout, Arion F Chatziioannou.   

Abstract

UNLABELLED: The Inveon dedicated PET (DPET) scanner is the latest generation of preclinical PET systems devoted to high-resolution and high-sensitivity murine model imaging. In this study, we report on its performance based on the National Electrical Manufacturers Association (NEMA) NU-4 standards.
METHODS: The Inveon DPET consists of 64 lutetium oxyorthosilicate block detectors arranged in 4 contiguous rings, with a 16.1-cm ring diameter and a 12.7-cm axial length. Each detector block consists of a 20 x 20 lutetium oxyorthosilicate crystal array of 1.51 x 1.51 x 10.0 mm elements. The scintillation light is transmitted to position-sensitive photomultiplier tubes via optical light guides. Energy resolution, spatial resolution, sensitivity, scatter fraction, and counting-rate performance were evaluated. The NEMA NU-4 image-quality phantom and a healthy mouse injected with (18)F-FDG and (18)F(-) were scanned to evaluate the imaging capability of the Inveon DPET.
RESULTS: The energy resolution at 511 keV was 14.6% on average for the entire system. In-plane radial and tangential resolutions reconstructed with Fourier rebinning and filtered backprojection algorithms were below 1.8-mm full width at half maximum (FWHM) at the center of the field of view. The radial and tangential resolution remained under 2.0 mm, and the axial resolution remained under 2.5-mm FWHM within the central 4-cm diameter of the field of view. The absolute sensitivity of the system was 9.3% for an energy window of 250-625 keV and a timing window of 3.432 ns. At a 350- to 625-keV energy window and a 3.432-ns timing window, the peak noise equivalent counting rate was 1,670 kcps at 130 MBq for the mouse-sized phantom and 590 kcps at 110 MBq for the rat-sized phantom. The scatter fractions at the same acquisition settings were 7.8% and 17.2% for the mouse- and rat-sized phantoms, respectively. The mouse image-quality phantom results demonstrate that for typical mouse acquisitions, the image quality correlates well with the measured performance parameters in terms of image uniformity, recovery coefficients, attenuation, and scatter corrections.
CONCLUSION: The Inveon system, compared with previous generations of preclinical PET systems from the same manufacturer, shows significantly improved energy resolution, sensitivity, axial coverage, and counting-rate capabilities. The performance of the Inveon is suitable for successful murine model imaging experiments.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19223424      PMCID: PMC2803022          DOI: 10.2967/jnumed.108.056374

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  16 in total

1.  Comparison of 3-D maximum a posteriori and filtered backprojection algorithms for high-resolution animal imaging with microPET.

Authors:  A Chatziioannou; J Qi; A Moore; A Annala; K Nguyen; R Leahy; S R Cherry
Journal:  IEEE Trans Med Imaging       Date:  2000-05       Impact factor: 10.048

Review 2.  Use of positron emission tomography in animal research.

Authors:  S R Cherry; S S Gambhir
Journal:  ILAR J       Date:  2001

3.  NEMA NU 2 performance tests for scanners with intrinsic radioactivity.

Authors:  Charles C Watson; Michael E Casey; Lars Eriksson; Tim Mulnix; Doug Adams; Bernard Bendriem
Journal:  J Nucl Med       Date:  2004-05       Impact factor: 10.057

4.  Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging.

Authors:  Yuan-Chuan Tai; Ananya Ruangma; Douglas Rowland; Stefan Siegel; Danny F Newport; Patrick L Chow; Richard Laforest
Journal:  J Nucl Med       Date:  2005-03       Impact factor: 10.057

5.  Attenuation correction for small animal PET tomographs.

Authors:  Patrick L Chow; Fernando R Rannou; Arion F Chatziioannou
Journal:  Phys Med Biol       Date:  2005-04-06       Impact factor: 3.609

6.  Accelerated image reconstruction using ordered subsets of projection data.

Authors:  H M Hudson; R S Larkin
Journal:  IEEE Trans Med Imaging       Date:  1994       Impact factor: 10.048

7.  High-resolution 3D Bayesian image reconstruction using the microPET small-animal scanner.

Authors:  J Qi; R M Leahy; S R Cherry; A Chatziioannou; T H Farquhar
Journal:  Phys Med Biol       Date:  1998-04       Impact factor: 3.609

8.  Exact and approximate rebinning algorithms for 3-D PET data.

Authors:  M Defrise; P E Kinahan; D W Townsend; C Michel; M Sibomana; D F Newport
Journal:  IEEE Trans Med Imaging       Date:  1997-04       Impact factor: 10.048

9.  Performance characteristics of the 3-D OSEM algorithm in the reconstruction of small animal PET images. Ordered-subsets expectation-maximixation.

Authors:  R Yao; J Seidel; C A Johnson; M E Daube-Witherspoon; M V Green; R E Carson
Journal:  IEEE Trans Med Imaging       Date:  2000-08       Impact factor: 10.048

10.  Imaging of weak-source distributions in LSO-based small-animal PET scanners.

Authors:  Andrew L Goertzen; Joon Young Suk; Christopher J Thompson
Journal:  J Nucl Med       Date:  2007-09-14       Impact factor: 10.057

View more
  97 in total

1.  Small-animal molecular imaging methods.

Authors:  Robert A de Kemp; Frederick H Epstein; Ciprian Catana; Benjamin M W Tsui; Erik L Ritman
Journal:  J Nucl Med       Date:  2010-05-01       Impact factor: 10.057

2.  Estimation of the minimum detectable activity of preclinical PET imaging systems with an analytical method.

Authors:  Qinan Bao; Arion F Chatziioannou
Journal:  Med Phys       Date:  2010-11       Impact factor: 4.071

3.  Performance characterization of the Inveon preclinical small-animal PET/SPECT/CT system for multimodality imaging.

Authors:  Keiichi Magota; Naoki Kubo; Yuji Kuge; Ken-Ichi Nishijima; Songji Zhao; Nagara Tamaki
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-12-10       Impact factor: 9.236

4.  Striatal and extrastriatal microPET imaging of D2/D3 dopamine receptors in rat brain with [¹⁸F]fallypride and [¹⁸F]desmethoxyfallypride.

Authors:  Cristian C Constantinescu; Robert A Coleman; Min-Liang Pan; Jogeshwar Mukherjee
Journal:  Synapse       Date:  2011-03-21       Impact factor: 2.562

5.  Scatter characterization and correction for simultaneous multiple small-animal PET imaging.

Authors:  Rameshwar Prasad; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2014-04       Impact factor: 3.488

6.  On the assessment of spatial resolution of PET systems with iterative image reconstruction.

Authors:  Kuang Gong; Simon R Cherry; Jinyi Qi
Journal:  Phys Med Biol       Date:  2016-02-11       Impact factor: 3.609

7.  A cone-shaped phantom for assessment of small animal PET scatter fraction and count rate performance.

Authors:  Rameshwar Prasad; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2012-10       Impact factor: 3.488

8.  Dexamethasone-induced insulin resistance: kinetic modeling using novel PET radiopharmaceutical 6-deoxy-6-[(18)F]fluoro-D-glucose.

Authors:  Kuan-Hao Su; Visvanathan Chandramouli; Faramarz Ismail-Beigi; Raymond F Muzic
Journal:  Mol Imaging Biol       Date:  2014-10       Impact factor: 3.488

9.  Prelimbic Cortical Injections of a GABA Agonist and Antagonist: In Vivo Quantification of the Effect in the Rat Brain Using [(18)F] FDG MicroPET.

Authors:  Joke Parthoens; Stijn Servaes; Jeroen Verhaeghe; Sigrid Stroobants; Steven Staelens
Journal:  Mol Imaging Biol       Date:  2015-12       Impact factor: 3.488

10.  Resolution Enhancement in PET Reconstruction Using Collimation.

Authors:  Scott D Metzler; Samuel Matej; Joel S Karp
Journal:  IEEE Trans Nucl Sci       Date:  2013-02       Impact factor: 1.679

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.