Keratometry with five different techniques: a study of device repeatability and inter-device agreement.

Shiva Mehravaran1, Soheila Asgari, Sara Bigdeli, Ashkan Shahnazi, Hassan Hashemi.   

Abstract

The purpose of this study was to determine and compare the repeatability of keratometry measurements with the Javal manual keratometer, Topcon automated kerato-refractometer, IOLMaster, EyeSys Corneal Analysis System, and Pentacam Scheimpflug Topography System, and determine the agreement of measurements of the latter four devices with the Javal keratometer as the gold standard. In this cross-sectional study, 21 people with no history of ocular disease or surgery were examined twice with each device. Minimum, maximum, and average keratometry readings were recorded in diopters (D) for each eye and used in the analyses. For statistical analysis, we determined correlation coefficients and used the Bland-Altman method and calculated the 95 % limits of agreement (LoA). All repeatability coefficients were satisfactorily high. Best repeatability for minimum keratometry and maximum keratometry readings was seen with IOLMaster (95 % LoA -0.23 to 0.19 D and -0.31 to 0.32 D, respectively). Agreement with Javal manual keratometry was best with Topcon and IOLMaster for minimum keratometry readings (95 % LoA -0.67 to 0.28 D and -0.57 to 0.38 D, respectively), and with Topcon for maximum keratometry readings (95 % LoA -0.85 to 0.63 D). In our series of normal eyes, keratometry readings with Topcon, IOLMaster, and Pentacam showed very good agreement with Javal measurements, and inter-device agreements were better than the repeatability of the manual gold standard. However, differences between Javal and EyeSys may not be clinically acceptable, and these two devices should not be used interchangeably. Further studies are needed to investigate these issues in non-virgin eyes.

Entities:  

Mesh:

Year:  2014        PMID: 24562593     DOI: 10.1007/s10792-013-9895-3

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  36 in total

Review 1.  Calculation of intraocular lens power: a review.

Authors:  Thomas Olsen
Journal:  Acta Ophthalmol Scand       Date:  2007-04-02

2.  Accuracy of a new partial coherence interferometry analyser for biometric measurements.

Authors:  M P Holzer; M Mamusa; G U Auffarth
Journal:  Br J Ophthalmol       Date:  2009-03-15       Impact factor: 4.638

3.  Evaluation of keratometric, pachymetric, and elevation parameters of keratoconic corneas with pentacam.

Authors:  Kata Miháltz; Illés Kovács; Agnes Takács; Zoltán Z Nagy
Journal:  Cornea       Date:  2009-10       Impact factor: 2.651

4.  Comparison of keratometric readings as obtained by Javal Ophthalmometer and Corneal Analysis System (EyeSys).

Authors:  M K Tsilimbaris; I G Vlachonikolis; D Siganos; G Makridakis; I G Pallikaris
Journal:  Refract Corneal Surg       Date:  1991 Sep-Oct

5.  Comparison of keratometric values of healthy and diseased eyes measured by Javal keratometer, EyeSys, and PAR.

Authors:  D Varssano; C J Rapuano; J I Luchs
Journal:  J Cataract Refract Surg       Date:  1997-04       Impact factor: 3.351

6.  Assessment of toroidal surfaces by the measurement of curvature in three fixed meridians.

Authors:  C W Fowler
Journal:  Ophthalmic Physiol Opt       Date:  1989-01       Impact factor: 3.117

7.  Comparison of three keratometry instruments.

Authors:  D G Tennen; R H Keates; C Montoya
Journal:  J Cataract Refract Surg       Date:  1995-07       Impact factor: 3.351

8.  Repeatability and reproducibility of corneal curvature measurements using the Pentacam and Keratron topography systems.

Authors:  Takushi Kawamorita; Nanami Nakayama; Hiroshi Uozato
Journal:  J Refract Surg       Date:  2009-06       Impact factor: 3.573

9.  Effect of keratometer and axial length measurement errors on primary implant power calculations.

Authors:  J R McEwan; R K Massengill; S D Friedel
Journal:  J Cataract Refract Surg       Date:  1990-01       Impact factor: 3.351

10.  Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation.

Authors:  János Németh; Orsolya Fekete; Norbert Pesztenlehrer
Journal:  J Cataract Refract Surg       Date:  2003-01       Impact factor: 3.351

View more
  9 in total

1.  Placido disk-based topography versus high-resolution rotating Scheimpflug camera for corneal power measurements in keratoconic and post-LASIK eyes: reliability and agreement.

Authors:  Rachele R Penna; Ugo de Sanctis; Martina Catalano; Luca Brusasco; Federico M Grignolo
Journal:  Int J Ophthalmol       Date:  2017-03-18       Impact factor: 1.779

2.  Keratoconus after 40 years of age: a longitudinal comparative population-based study.

Authors:  Hassan Hashemi; Soheila Asgari; Shiva Mehravaran; Mohammad Hassan Emamian; Akbar Fotouhi
Journal:  Int Ophthalmol       Date:  2019-11-07       Impact factor: 2.031

3.  The effect of the timing of the cessation of contact lens use on the results of biometry.

Authors:  Colin Goudie; Andrew Tatham; Rhys Davies; Alison Sifton; Mark Wright
Journal:  Eye (Lond)       Date:  2018-02-02       Impact factor: 3.775

4.  Validation of corneal topographic and aberrometric measurements obtained by color light-emitting diode reflection topography in healthy eyes.

Authors:  David P Piñero; Ainhoa Molina-Martín; Vicent J Camps; Dolores de Fez; María Teresa Caballero
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2019-09-03       Impact factor: 3.117

5.  OPD-Scan III: a repeatability and inter-device agreement study of a multifunctional device in emmetropia, ametropia, and keratoconus.

Authors:  Soheila Asgari; Hassan Hashemi; Ebrahim Jafarzadehpur; Alireza Mohamadi; Farhad Rezvan; Akbar Fotouhi
Journal:  Int Ophthalmol       Date:  2016-02-08       Impact factor: 2.031

6.  Precision and agreement of corneal power measurements obtained using a new corneal topographer OphthaTOP.

Authors:  Jinhai Huang; Giacomo Savini; Hao Chen; Fangjun Bao; Yuanguang Li; Haisi Chen; Weicong Lu; Ye Yu; Qinmei Wang
Journal:  PLoS One       Date:  2015-01-05       Impact factor: 3.240

7.  Effect of ageing on keratoconic corneas.

Authors:  Michel Millodot; Ilya Ortenberg; Karen Lahav-Yacouel; Shmuel Behrman
Journal:  J Optom       Date:  2015-07-02

8.  OPD scan III accuracy: Topographic and aberrometric indices after accelerated corneal cross-linking.

Authors:  Soheila Asgari; Hassan Hashemi
Journal:  J Curr Ophthalmol       Date:  2017-11-14

9.  Agreement between Pentacam and handheld Auto-Refractor/Keratometer for keratometry measurement.

Authors:  Hassan Hashemi; Samira Heydarian; Abbas Ali Yekta; Mohamadreza Aghamirsalim; Mahin Ahmadi-Pishkuhi; Mehrnaz Valadkhan; Hadi Ostadimoghaddam; Ahmad Ahmadzadeh Amiri; Mehdi Khabazkhoob
Journal:  J Optom       Date:  2019-07-10
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.