Placido disk-based topography versus high-resolution rotating Scheimpflug camera for corneal power measurements in keratoconic and post-LASIK eyes: reliability and agreement.

Rachele R Penna1, Ugo de Sanctis1, Martina Catalano1, Luca Brusasco1, Federico M Grignolo1.   

Abstract

AIM: To compare the repeatability/reproducibility of measurement by high-resolution Placido disk-based topography with that of a high-resolution rotating Scheimpflug camera and assess the agreement between the two instruments in measuring corneal power in eyes with keratoconus and post-laser in situ keratomileusis (LASIK).
METHODS: One eye each of 36 keratoconic patients and 20 subjects who had undergone LASIK was included in this prospective observational study. Two independent examiners worked in a random order to take three measurements of each eye with both instruments. Four parameters were measured on the anterior cornea: steep keratometry (Ks), flat keratometry (Kf), mean keratometry (Km), and astigmatism (Ks-Kf). Intra-examiner repeatability and inter-examiner reproducibility were evaluated by calculating the within-subject standard deviation (Sw) the coefficient of repeatability (R), the coefficient of variation (CoV), and the intraclass correlation coefficient (ICC). Agreement between instruments was tested with the Bland-Altman method by calculating the 95% limits of agreement (95% LoA).
RESULTS: In keratoconic eyes, the intra-examiner and inter-examiner ICC were >0.95. As compared with measurement by high-resolution Placido disk-based topography, the intra-examiner R of the high-resolution rotating Scheimpflug camera was lower for Kf (0.32 vs 0.88), Ks (0.61 vs 0.88), and Km (0.32 vs 0.84) but higher for Ks-Kf (0.70 vs 0.57). Inter-examiner R values were lower for all parameters measured using the high-resolution rotating Scheimpflug camera. The 95% LoA were -1.28 to +0.55 for Kf, -1.36 to +0.99 for Ks, -1.08 to +0.50 for Km, and -1.11 to +1.48 for Ks-Kf. In the post-LASIK eyes, the intra-examiner and inter-examiner ICC were >0.87 for all parameters. The intra-examiner and inter-examiner R were lower for all parameters measured using the high-resolution rotating Scheimpflug camera. The intra-examiner R was 0.17 vs 0.88 for Kf, 0.21 vs 0.88 for Ks, 0.17 vs 0.86 for Km, and 0.28 vs 0.33 for Ks-Kf. The inter-examiner R was 0.09 vs 0.64 for Kf, 0.15 vs 0.56 for Ks, 0.09 vs 0.59 for Km, and 0.18 vs 0.23 for Ks-Kf. The 95% LoA were -0.54 to +0.58 for Kf, -0.51 to +0.53 for Ks and Km, and -0.28 to +0.27 for Ks-Kf.
CONCLUSION: As compared with Placido disk-based topography, the high-resolution rotating Scheimpflug camera provides more repeatable and reproducible measurements of Ks, Kf and Ks in keratoconic and post-LASIK eyes. Agreement between instruments is fair in keratoconus and very good in post-LASIK eyes.

Entities:  

Keywords:  Placido disk-based topography; Scheimpflug camera; corneal power measurement; keratoconus; post-LASIK

Year:  2017        PMID: 28393039      PMCID: PMC5360783          DOI: 10.18240/ijo.2017.03.20

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  20 in total

Review 1.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

2.  Differential precision of corneal Pentacam HR measurements in early and advanced keratoconus.

Authors:  Tom H Flynn; Daya P Sharma; Catey Bunce; Mark R Wilkins
Journal:  Br J Ophthalmol       Date:  2015-12-11       Impact factor: 4.638

3.  Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes.

Authors:  Giacomo Savini; Michele Carbonelli; Piero Barboni; Kenneth J Hoffer
Journal:  J Cataract Refract Surg       Date:  2011-02       Impact factor: 3.351

4.  Measurement error.

Authors:  J M Bland; D G Altman
Journal:  BMJ       Date:  1996-06-29

Review 5.  Biomicroscopic signs and disease severity in keratoconus. Collaborative Longitudinal Evaluation of Keratoconus (CLEK) Study Group.

Authors:  K Zadnik; J T Barr; M O Gordon; T B Edrington
Journal:  Cornea       Date:  1996-03       Impact factor: 2.651

6.  Effect of keratoconus grades on repeatability of keratometry readings: Comparison of 5 devices.

Authors:  Hassan Hashemi; Abbasali Yekta; Mehdi Khabazkhoob
Journal:  J Cataract Refract Surg       Date:  2015-05       Impact factor: 3.351

7.  Repeatability of placido-based corneal topography in keratoconus.

Authors:  Sara Ortiz-Toquero; Guadalupe Rodriguez; Victoria de Juan; Raul Martin
Journal:  Optom Vis Sci       Date:  2014-12       Impact factor: 1.973

8.  Overview of the repeatability, reproducibility, and agreement of the biometry values provided by various ophthalmic devices.

Authors:  Jos J Rozema; Kristien Wouters; Danny G P Mathysen; Marie-José Tassignon
Journal:  Am J Ophthalmol       Date:  2014-08-13       Impact factor: 5.258

9.  Variability in Oculus Pentacam tomographer measurements in patients with keratoconus.

Authors:  Lucas M M Vianna; Beatriz Muñoz; Frank S Hwang; Arusha Gupta; Albert S Jun
Journal:  Cornea       Date:  2015-03       Impact factor: 2.651

10.  Repeatability of corneal parameters with Pentacam after laser in situ keratomileusis.

Authors:  Rajeev Jain; Grewal Dilraj; Satinder Pal Singh Grewal
Journal:  Indian J Ophthalmol       Date:  2007 Sep-Oct       Impact factor: 1.848

View more
  5 in total

1.  Comparative clinical accuracy analysis of the newly developed ZZ IOL and four existing IOL formulas for post-corneal refractive surgery eyes.

Authors:  Jun Zhang; Jie Shao; Li Zheng; Ye Shen; Xia Zhao
Journal:  BMC Ophthalmol       Date:  2021-05-25       Impact factor: 2.209

2.  One-Year Efficacy and Safety of Combined Photorefractive Keratectomy and Accelerated Corneal Collagen Cross-Linking after Intacs SK Intracorneal Ring Segment Implantation in Moderate Keratoconus.

Authors:  Il Hwan Koh; Kyoung Yul Seo; Seong Bae Park; Hun Yang; InSik Kim; Jin Sun Kim; David G Hwang; Sang Min Nam
Journal:  Biomed Res Int       Date:  2019-07-08       Impact factor: 3.411

3.  Association between Corneal Stiffness Parameter at the First Applanation and Keratoconus Severity.

Authors:  Kaili Yang; Liyan Xu; Qi Fan; Shengwei Ren
Journal:  J Ophthalmol       Date:  2020-12-02       Impact factor: 1.909

4.  Trends in Corneal Topography and Tomography Imaging for Keratoconus Management.

Authors:  Alan W Kong; Tessnim R Ahmad; Marcus L Turner; Jackson Barnett; Gurbani Kaur; Neel D Pasricha; Maanasa Indaram
Journal:  Clin Ophthalmol       Date:  2022-04-28

5.  Inter-Ocular Asymmetry in Anterior Corneal Aberrations Using Placido Disk-Based Topography.

Authors:  Zeyad Alzaben; Yazan Gammoh; Marta Freixas; Ahmad Zaben; Miguel A Zapata; Dana N Koff
Journal:  Clin Ophthalmol       Date:  2020-05-27
  5 in total