Validation of corneal topographic and aberrometric measurements obtained by color light-emitting diode reflection topography in healthy eyes.

David P Piñero1,2, Ainhoa Molina-Martín3, Vicent J Camps3, Dolores de Fez3, María Teresa Caballero3.   

Abstract

PURPOSE: To evaluate the intrasession repeatability of anterior corneal topographic and aberrometric measurements provided by a color-LED topographer as well as their interchangeability with those provided by a Scheimpflug-based system in healthy eyes.
METHODS: Thirty-five healthy eyes of 35 patients (age, 16-66 years) were enrolled. A complete eye examination was performed in all cases including a complete corneal analysis with the Scheimpflug-based system Pentacam (Oculus Optikgeräte) (one measurement) and the Cassini system (i-Optics) (three consecutive measurements). Intrasession repeatability of the Cassini measurements was assessed with the within-subject standard deviation (Sw) and the intraclass correlation coefficient (ICC). The Bland-Altman analysis was used to evaluate the agreement between both devices.
RESULTS: Mean Sw for keratometric readings was 0.02 mm (ICC ≥ 0.992), ranging between 0.16 and 0.05 D (ICC 0.930-0.978) for anterior and total astigmatic measurements. Mean Sw for asphericity and corneal diameter were 0.06 (ICC 0.926) and 0.03 mm (IC 0.997), respectively. Aberrometric parameters showed ICCs ≥ 0.816, except for Z42 (ICC 0.741) and Z44 (ICC 0.544). When comparing devices, statistically significant differences were found for most of topographic and aberrometric data (p ≤ 0.044). Likewise, ranges of agreement between devices were clinically relevant (keratometry > 0.06 mm; total astigmatic components > 0.69 D; asphericity 0.35; second-, third-, and fourth-order Zernike terms, more than 0.20, 0.13, and 0.01 μm, respectively).
CONCLUSIONS: Consistent anterior corneal topographic, total corneal astigmatic, and aberrometric measurements are obtained with color-LED topography in healthy eyes, which are not interchangeable with those provided by the Scheimpflug-based topography.

Entities:  

Keywords:  Aberrometry; Anterior corneal curvature; Cassini; Color light-emitting diode reflection topography; Corneal topography; Pentacam; Posterior corneal curvature

Mesh:

Year:  2019        PMID: 31482276     DOI: 10.1007/s00417-019-04453-5

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  31 in total

1.  Intrasubject repeatability of corneal morphology measurements obtained with a new Scheimpflug photography-based system.

Authors:  Raúl Montalbán; David P Piñero; Jaime Javaloy; Jorge L Alió
Journal:  J Cataract Refract Surg       Date:  2012-04-26       Impact factor: 3.351

2.  Repeatability of automatic measurements by a new anterior segment optical coherence tomographer combined with Placido topography and agreement with 2 Scheimpflug cameras.

Authors:  Giacomo Savini; Domenico Schiano-Lomoriello; Kenneth J Hoffer
Journal:  J Cataract Refract Surg       Date:  2018-04-25       Impact factor: 3.351

3.  Intrasubject repeatability of corneal power, thickness, and wavefront aberrations with a new version of a dual rotating Scheimpflug-Placido system.

Authors:  Alejandro Cerviño; Alberto Dominguez-Vicent; Teresa Ferrer-Blasco; Santiago García-Lázaro; Cesar Albarrán-Diego
Journal:  J Cataract Refract Surg       Date:  2014-11-18       Impact factor: 3.351

4.  Accuracy of Total Corneal Astigmatism Measurements With a Scheimpflug Imager and a Color Light-Emitting Diode Corneal Topographer.

Authors:  Stijn Klijn; Nicolaas J Reus; Charlotte M van der Sommen; Victor Arni D P Sicam
Journal:  Am J Ophthalmol       Date:  2016-04-27       Impact factor: 5.258

5.  Repeatability of aberrometric measurements in normal and keratoconus eyes using a new Scheimpflug-Placido topographer.

Authors:  Hasan Ali Bayhan; Seray Aslan Bayhan; Ersin Muhafız; Izzet Can
Journal:  J Cataract Refract Surg       Date:  2013-12-22       Impact factor: 3.351

6.  Repeatability of corneal first-surface wavefront aberrations measured with Pentacam corneal topography.

Authors:  Hema Shankar; Deepa Taranath; Chandramalar T Santhirathelagan; Konrad Pesudovs
Journal:  J Cataract Refract Surg       Date:  2008-05       Impact factor: 3.351

7.  Clinical Correlation between Placido, Scheimpflug and LED Color Reflection Topographies in Imaging of a Scarred Cornea.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Case Rep Ophthalmol       Date:  2014-10-01

8.  Color light-emitting diode reflection topography: validation of keratometric repeatability in a large sample of wide cylindrical-range corneas.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2015-02-05

9.  Corneal Topographic and Aberrometric Measurements Obtained with a Multidiagnostic Device in Healthy Eyes: Intrasession Repeatability.

Authors:  David P Piñero; Alberto López-Navarro; Inmaculada Cabezos; Dolores de Fez; María T Caballero; Vicent J Camps
Journal:  J Ophthalmol       Date:  2017-01-04       Impact factor: 1.909

10.  Repeatability and agreement of five imaging systems for measuring anterior segment parameters in healthy eyes.

Authors:  Mukesh Kumar; Rohit Shetty; Chaitra Jayadev; Harsha L Rao; Debarun Dutta
Journal:  Indian J Ophthalmol       Date:  2017-04       Impact factor: 1.848

View more
  1 in total

1.  Daily Variations of Corneal White-to-White Diameter Measured with Different Methods.

Authors:  Stefan Bandlitz; Makram Nakhoul; Konstantin Kotliar
Journal:  Clin Optom (Auckl)       Date:  2022-09-20
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.