Optical biometry intraocular lens power calculation using different formulas in patients with different axial lengths.

Jia-Kang Wang1, Shu-Wen Chang.   

Abstract

AIM: : To investigate the predictability of intraocular lens (IOL) power calculation using the IOLMaster and different IOL power calculation formulas in eyes with various axial length (AL).
METHODS: : Patients were included who underwent uneventful phacoemulsification with IOL implantation in the Department of Ophthalmology, Far Eastern Memorial Hospital, Taipei, Taiwan, China from February 2007 to January 2009. Preoperative AL and keratometric values (Ks) were measured by IOLMaster optical biometry. Patients were divided into 3 groups based on AL less than 22mm (Group 1), 22-26mm (Group 2), and more than 26mm (Group 3). The power of the implanted IOL was used to calculate the predicted postoperative spherical equivalence (SE) by various formulas: the Haigis, Hoffer Q, Holladay 1, and SRK/T. The predictive accuracy of each formula was analyzed by comparing the difference between the actual and predicted postoperative SE (MedAE, median absolute error). All the patients had follow-up periods exceeding 3 months.
RESULTS: : Totally, there were 200 eyes (33 eyes in Group 1, 92 eyes in Group 2, 75 eyes in Group 3). In all patients, the Haigis had the significantly lower MedAE generated by the other formulas (P<0.05). In Group 1 to 3, the MedAE calculated by the Haigis was either significantly lower or comparable to those calculated by the other formulas.
CONCLUSION: : Compared with other formulas using IOLMaster biometric data, the Haigis formula yields superior refractive results in eyes with various AL.

Entities:  

Keywords:  calculation formula; intraocular lens; optical biometry

Year:  2013        PMID: 23638414      PMCID: PMC3633751          DOI: 10.3980/j.issn.2222-3959.2013.02.08

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  37 in total

1.  Comparative estimation of anterior chamber depth by ultrasonography, Orbscan II, and IOLMaster.

Authors:  Aravind R Reddy; Milind V Pande; Paul Finn; Hazem El-Gogary
Journal:  J Cataract Refract Surg       Date:  2004-06       Impact factor: 3.351

2.  Sources of error in intraocular lens power calculation.

Authors:  T Olsen
Journal:  J Cataract Refract Surg       Date:  1992-03       Impact factor: 3.351

3.  Intraocular lens power calculation in short eyes.

Authors:  E A Gavin; C J Hammond
Journal:  Eye (Lond)       Date:  2007-03-16       Impact factor: 3.775

4.  Intraocular lens power calculation using the IOLMaster and various formulas in eyes with long axial length.

Authors:  Jia-Kang Wang; Chao-Yu Hu; Shu-Wen Chang
Journal:  J Cataract Refract Surg       Date:  2008-02       Impact factor: 3.351

5.  Development of the SRK/T intraocular lens implant power calculation formula.

Authors:  J A Retzlaff; D R Sanders; M C Kraff
Journal:  J Cataract Refract Surg       Date:  1990-05       Impact factor: 3.351

6.  Comparison of the SRK/T formula and other theoretical and regression formulas.

Authors:  D R Sanders; J A Retzlaff; M C Kraff; H V Gimbel; M G Raanan
Journal:  J Cataract Refract Surg       Date:  1990-05       Impact factor: 3.351

7.  Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation.

Authors:  János Németh; Orsolya Fekete; Norbert Pesztenlehrer
Journal:  J Cataract Refract Surg       Date:  2003-01       Impact factor: 3.351

8.  Partial coherence interferometry: a novel approach to biometry in cataract surgery.

Authors:  W Drexler; O Findl; R Menapace; G Rainer; C Vass; C K Hitzenberger; A F Fercher
Journal:  Am J Ophthalmol       Date:  1998-10       Impact factor: 5.258

9.  Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis.

Authors:  W Haigis; B Lege; N Miller; B Schneider
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2000-09       Impact factor: 3.117

10.  Intraocular lens power calculation using IOLMaster and various formulas in short eyes.

Authors:  Young Rae Roh; Sang Mok Lee; Young Keun Han; Mee Kum Kim; Won Ryang Wee; Jin Hak Lee
Journal:  Korean J Ophthalmol       Date:  2011-05-24
View more
  20 in total

1.  Comparison of refractive outcomes using Scheimpflug Holladay equivalent keratometry or IOLMaster 700 keratometry for IOL power calculation.

Authors:  Mustafa Aksoy; Leyla Asena; Sirel Gür Güngör; Ali Küçüködük; Ahmet Akman
Journal:  Int Ophthalmol       Date:  2021-03-17       Impact factor: 2.031

2.  Refractive results of phacoemulsification in vitrectomized patients.

Authors:  Jia-Kang Wang; Shu-Wen Chang
Journal:  Int Ophthalmol       Date:  2016-08-25       Impact factor: 2.031

3.  Scheimpflug camera combined with placido-disk corneal topography and optical biometry for intraocular lens power calculation.

Authors:  Ahmet Kirgiz; Kurşat Atalay; Havva Kaldirim; Kubra Serefoglu Cabuk; Mehmet Orcun Akdemir; Muhittin Taskapili
Journal:  Int Ophthalmol       Date:  2016-09-01       Impact factor: 2.031

4.  A Comparative Study to Assess the Predictability of Different IOL Power Calculation Formulas in Eyes of Short and Long Axial Length.

Authors:  Dharmil Doshi; Purvi Limdi; Nilesh Parekh; Neepa Gohil
Journal:  J Clin Diagn Res       Date:  2017-01-01

5.  Effectiveness, Sensitivity, and Specificity of Intraocular Lens Power Calculation Formulas for Short Eyes.

Authors:  Wiktor Stopyra
Journal:  Turk J Ophthalmol       Date:  2022-06-29

6.  Performance of the SRK/T formula using A-Scan ultrasound biometry after phacoemulsification in eyes with short and long axial lengths.

Authors:  Yunus Karabela; Mustafa Eliacik; Faruk Kaya
Journal:  BMC Ophthalmol       Date:  2016-07-08       Impact factor: 2.209

7.  Outcomes of Sutureless Iris-Claw Lens Implantation.

Authors:  Tomasz Choragiewicz; Robert Rejdak; Andrzej Grzybowski; Katarzyna Nowomiejska; Joanna Moneta-Wielgoś; Małgorzata Ozimek; Anselm G M Jünemann
Journal:  J Ophthalmol       Date:  2016-08-24       Impact factor: 1.909

8.  Effect of pupillary dilation on Haigis formula-calculated intraocular lens power measurement by using optical biometry.

Authors:  Bharkbhum Khambhiphant; Suganlaya Sasiwilasagorn; Nattida Chatbunchachai; Krit Pongpirul
Journal:  Clin Ophthalmol       Date:  2016-07-27

Review 9.  Refractive lens exchange in modern practice: when and when not to do it?

Authors:  Jorge L Alió; Andrzej Grzybowski; Dorota Romaniuk
Journal:  Eye Vis (Lond)       Date:  2014-12-10

10.  Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500.

Authors:  Ahmet Akman; Leyla Asena; Sirel Gür Güngör
Journal:  Br J Ophthalmol       Date:  2015-12-16       Impact factor: 4.638

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.