Validation of continuous and noninvasive hemoglobin monitoring by pulse CO-oximetry in Japanese surgical patients.

Tsuyoshi Isosu1, Shinju Obara, Atsuyuki Hosono, Satoshi Ohashi, Yuko Nakano, Tsuyoshi Imaizumi, Midori Mogami, Masahiro Murakawa.   

Abstract

We evaluated the accuracy of noninvasive and continuous total hemoglobin (SpHb) monitoring with the Radical-7(®) Pulse CO-Oximeter in Japanese surgical patients before and after an in vivo adjustment of the first SpHb value to match the first reference value from a satellite laboratory CO-Oximeter. Twenty patients undergoing surgical procedures with general anesthesia were monitored with Pulse CO-Oximetry for SpHb. Laboratory CO-Oximeter values (tHb) were compared to SpHb at the time of the blood draws. Bias, precision, limits of agreement and correlation coefficient of SpHb compared to tHb were calculated before and after SpHb values were adjusted by subtracting the difference between the first SpHb and tHb value from all subsequent SpHb values. Trending of SpHb to tHb and the effect of perfusion index (PI) on the agreement of SpHb to tHb were also analyzed. Ninety-two tHb values were compared to the SpHb. Bias ± 1SD was 0.2 ± 1.5 g/dL before in vivo adjustment and -0.7 ± 1.0 g/dL after in vivo adjustment. Bland-Altman analysis showed limits of agreement of -2.8 to 3.1 g/dL before in vivo adjustment and -2.8 to 1.4 g/dL after in vivo adjustment. The correlation coefficient was 0.76 prior to in vivo adjustment and 0.87 after in vivo adjustment. In patients with adequate perfusion (PI ≥1.4) the correlation coefficient was 0.89. In vivo adjustment of SpHb significantly improved the accuracy in our cohort of Japanese surgical patients. The strongest correlation between SpHb and tHb values was observed in patients with adequate peripheral perfusion suggesting that low perfusion may affect the accuracy of SpHb monitoring.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22986804     DOI: 10.1007/s10877-012-9397-2

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  9 in total

Review 1.  A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output.

Authors:  Lester A Critchley; Anna Lee; Anthony M-H Ho
Journal:  Anesth Analg       Date:  2010-08-24       Impact factor: 5.108

2.  Agreement between methods of measurement with multiple observations per individual.

Authors:  J Martin Bland; Douglas G Altman
Journal:  J Biopharm Stat       Date:  2007       Impact factor: 1.051

3.  The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-Oximetry in human subjects undergoing hemodilution.

Authors:  Mark R Macknet; Martin Allard; Richard L Applegate; James Rook
Journal:  Anesth Analg       Date:  2010-11-03       Impact factor: 5.108

4.  Recommendations for reference method for haemoglobinometry in human blood (ICSH standard 1986) and specifications for international haemiglobincyanide reference preparation (3rd edition). International Committee for Standardization in Haematology; Expert Panel on Haemoglobinometry.

Authors: 
Journal:  Clin Lab Haematol       Date:  1987

5.  Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients.

Authors:  Denis Frasca; Claire Dahyot-Fizelier; Karen Catherine; Quentin Levrat; Bertrand Debaene; Olivier Mimoz
Journal:  Crit Care Med       Date:  2011-10       Impact factor: 7.598

6.  Continuous noninvasive hemoglobin monitoring during complex spine surgery.

Authors:  Lauren Berkow; Stephanie Rotolo; Erin Mirski
Journal:  Anesth Analg       Date:  2011-09-29       Impact factor: 5.108

7.  Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement.

Authors:  Lionel Lamhaut; Roxana Apriotesei; Xavier Combes; Marc Lejay; Pierre Carli; Benoît Vivien
Journal:  Anesthesiology       Date:  2011-09       Impact factor: 7.892

8.  A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery.

Authors:  Ronald D Miller; Theresa A Ward; Stephen C Shiboski; Neal H Cohen
Journal:  Anesth Analg       Date:  2011-03-08       Impact factor: 5.108

9.  Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: the effects of oximeter probe type and gender.

Authors:  John R Feiner; John W Severinghaus; Philip E Bickler
Journal:  Anesth Analg       Date:  2007-12       Impact factor: 5.108

  9 in total
  13 in total

1.  Evaluation of multiwave pulse total-hemoglobinometer during general anesthesia.

Authors:  Daisuke Toyoda; Rie Yasumura; Mitsue Fukuda; Ryoichi Ochiai; Yoshifumi Kotake
Journal:  J Anesth       Date:  2013-10-22       Impact factor: 2.078

2.  Effects of indigo carmine intravenous injection on noninvasive and continuous total hemoglobin measurement.

Authors:  Tsuyoshi Isosu; Tomohiko Satoh; Rieko Oishi; Tsuyoshi Imaizumi; Takahiro Hakozaki; Shinju Obara; Yukihiro Ikegami; Shin Kurosawa; Masahiro Murakawa
Journal:  J Clin Monit Comput       Date:  2015-06-17       Impact factor: 2.502

3.  Comparison of invasive and noninvasive blood hemoglobin measurement in the operating room: a systematic review and meta-analysis.

Authors:  Hosein Shabaninejad; Nashmil Ghadimi; Kourosh Sayehmiri; Hossein Hosseinifard; Rasoul Azarfarin; Hasan Abolghasem Gorji
Journal:  J Anesth       Date:  2019-03-20       Impact factor: 2.078

4.  Effects of indigo carmine intravenous injection on noninvasive and continuous total hemoglobin measurement with using the Revision L sensor.

Authors:  Tsuyoshi Isosu; Shinju Obara; Takahiro Hakozaki; Tsuyoshi Imaizumi; Yuzo Iseki; Midori Mogami; Satoshi Ohashi; Yukihiro Ikegami; Shin Kurosawa; Masahiro Murakawa
Journal:  J Clin Monit Comput       Date:  2016-02-22       Impact factor: 2.502

5.  Noninvasively Measured Hemoglobin Concentration Reflects Arterial Hemoglobin Concentration Before but Not After Cardiopulmonary Bypass in Patients Undergoing Coronary Artery or Valve Surgery.

Authors:  Matthias L Riess; Paul S Pagel
Journal:  J Cardiothorac Vasc Anesth       Date:  2016-03-24       Impact factor: 2.628

6.  Validation of noninvasive hemoglobin measurement by pulse co-oximeter in newborn infants.

Authors:  C Nicholas; R George; S Sardesai; M Durand; R Ramanathan; R Cayabyab
Journal:  J Perinatol       Date:  2015-03-05       Impact factor: 2.521

7.  Effects of indigo carmine intravenous injection on oxygen reserve index (ORi™) measurement.

Authors:  Tsuyoshi Isosu; Keisuke Yoshida; Rieko Oishi; Tsuyoshi Imaizumi; Yuzo Iseki; Norie Sanbe; Yukihiro Ikegami; Shinju Obara; Shin Kurosawa; Masahiro Murakawa
Journal:  J Clin Monit Comput       Date:  2017-10-03       Impact factor: 2.502

8.  Accuracy of non-invasive continuous total hemoglobin measurement by Pulse CO-Oximetry in severe traumatized and surgical bleeding patients.

Authors:  Werner Baulig; Burkhardt Seifert; Donat R Spahn; Oliver M Theusinger
Journal:  J Clin Monit Comput       Date:  2015-12-19       Impact factor: 2.502

9.  Accuracy and trending of non-invasive hemoglobin measurement during different volume and perfusion statuses.

Authors:  Abdelmoneim Adel; Wael Awada; Bassant Abdelhamid; Heba Omar; Omnia Abd El Dayem; Ahmed Hasanin; Ashraf Rady
Journal:  J Clin Monit Comput       Date:  2018-01-15       Impact factor: 2.502

10.  Assessment of noninvasive, percutaneous hemoglobin measurement in pregnant and early postpartum women.

Authors:  Atsushi Yoshida; Keiko Saito; Kenji Ishii; Isao Azuma; Hidenori Sasa; Kenichi Furuya
Journal:  Med Devices (Auckl)       Date:  2014-01-15
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.