Validation of noninvasive hemoglobin measurement by pulse co-oximeter in newborn infants.

C Nicholas1, R George2, S Sardesai3, M Durand3, R Ramanathan1, R Cayabyab3.   

Abstract

OBJECTIVE: To describe the accuracy of noninvasive hemoglobin (Hb) obtained with pulse co-oximeter (SpHb) compared with total Hb (tHb) from laboratory co-oximeter in neonates. STUDY
DESIGN: Neonates with birth weight (BW) <3000 g admitted to LAC+USC Medical Center neonatal intensive care unit were included. SpHb was recorded using Masimo Radical-7 and compared with tHb. A total of three data sets were obtained for each patient. Regression analysis and Bland-Altman analysis were performed. RESULT: Sixty-one patients (mean±s.d., BW 1177±610 g and gestational age 28.7±3.9 weeks) were enrolled. The mean tHb value was 13.9±2.0 g dl(-1) and the mean SpHb was 14.0±2.0 g dl(-1). There was a moderately positive correlation between SpHb and tHb (r=0.66, P<0.001) with a bias and precision of -0.09±1.67 g dl(-1). Data from a subgroup of infants with gestational age ⩽32 weeks (52/61 patients) were analyzed, and the correlation coefficient was moderately positive (r=0.69, P<0.001) with a bias and precision of -0.23±1.60 g dl(-1).
CONCLUSION: Our results suggest that noninvasive SpHb may be considered as an adjunct to invasive tHb measurements in newborn infants <3000 g especially in preterm infants ⩽32 weeks of gestation.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25742288     DOI: 10.1038/jp.2015.12

Source DB:  PubMed          Journal:  J Perinatol        ISSN: 0743-8346            Impact factor:   2.521


  17 in total

Review 1.  Noninvasive hemoglobin monitoring: how accurate is enough?

Authors:  Mark J Rice; Nikolaus Gravenstein; Timothy E Morey
Journal:  Anesth Analg       Date:  2013-07-10       Impact factor: 5.108

2.  Performance evaluation of a noninvasive hemoglobin monitoring device.

Authors:  Etienne Gayat; Antoine Bodin; Caroline Sportiello; Mireille Boisson; Jean-François Dreyfus; Emmanuel Mathieu; Marc Fischler
Journal:  Ann Emerg Med       Date:  2011-01-19       Impact factor: 5.721

3.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

4.  The accuracy of noninvasive hemoglobin monitoring using the radical-7 pulse CO-Oximeter in children undergoing neurosurgery.

Authors:  Yong-Hee Park; Ji-Hyun Lee; Hyun-Gul Song; Hyo-Jin Byon; Hee-Soo Kim; Jin-Tae Kim
Journal:  Anesth Analg       Date:  2012-11-09       Impact factor: 5.108

5.  Spectroscopic noninvasive measurement of hemoglobin compared with capillary and venous values in neonates.

Authors:  H Rabe; R Fernandez Alvarez; T Whitfield; F Lawson; H Jungmann
Journal:  Neonatology       Date:  2009-11-25       Impact factor: 4.035

6.  Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement.

Authors:  Lionel Lamhaut; Roxana Apriotesei; Xavier Combes; Marc Lejay; Pierre Carli; Benoît Vivien
Journal:  Anesthesiology       Date:  2011-09       Impact factor: 7.892

7.  Pulse total-hemoglobinometer provides accurate noninvasive monitoring.

Authors:  Eisei Noiri; Naoki Kobayashi; Yoshiaki Takamura; Takehiko Iijima; Toshiyuki Takagi; Kent Doi; Akihide Nakao; Tokunori Yamamoto; Sunao Takeda; Toshiro Fujita
Journal:  Crit Care Med       Date:  2005-12       Impact factor: 7.598

8.  Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb Station.

Authors:  Marlin Wayne Causey; Seth Miller; Andrew Foster; Alec Beekley; David Zenger; Matthew Martin
Journal:  Am J Surg       Date:  2011-05       Impact factor: 2.565

9.  Accuracy of noninvasive and continuous hemoglobin measurement by pulse co-oximetry during preoperative phlebotomy.

Authors:  Elisabeth Dewhirst; Aymen Naguib; Peter Winch; Julie Rice; Mark Galantowicz; Patrick McConnell; Joseph D Tobias
Journal:  J Intensive Care Med       Date:  2013-04-22       Impact factor: 3.510

10.  The efficacy of noninvasive hemoglobin measurement by pulse CO-oximetry in neonates.

Authors:  Young Hwa Jung; Juyoung Lee; Han-Suk Kim; Seung Han Shin; Jin A Sohn; Ee-Kyung Kim; Jung-Hwan Choi
Journal:  Pediatr Crit Care Med       Date:  2013-01       Impact factor: 3.624

View more
  3 in total

1.  Validation of noninvasive hemoglobin measurements using co-oximetry in anesthetized dogs.

Authors:  Matt R Read; Jenna Rondeau; Grace P S Kwong
Journal:  Can Vet J       Date:  2016-11       Impact factor: 1.008

2.  Minimizing blood loss and the need for transfusions in very premature infants.

Authors:  Brigitte Lemyre; Megan Sample; Thierry Lacaze-Masmonteil
Journal:  Paediatr Child Health       Date:  2015 Nov-Dec       Impact factor: 2.253

3.  Validity of non-invasive point-of-care hemoglobin estimation in healthy and sick children-a method comparison study.

Authors:  Aditya Bhat; Amit Upadhyay; Vijay Jaiswal; Deepak Chawla; Dharamveer Singh; Mithilesh Kumar; C P Yadav
Journal:  Eur J Pediatr       Date:  2015-08-19       Impact factor: 3.183

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.