Noninvasively Measured Hemoglobin Concentration Reflects Arterial Hemoglobin Concentration Before but Not After Cardiopulmonary Bypass in Patients Undergoing Coronary Artery or Valve Surgery.

Matthias L Riess1, Paul S Pagel2.   

Abstract

OBJECTIVE: This study compared noninvasively measured hemoglobin and arterial hemoglobin before and after cardiopulmonary bypass in patients undergoing coronary artery or valve surgery.
DESIGN: Observational study with retrospective data analysis.
SETTING: Veterans Affairs hospital. PARTICIPANTS: Thirty-five men.
INTERVENTIONS: None.
MEASUREMENTS AND MAIN RESULTS: Hemoglobin values were measured noninvasively by co-oximetry to corresponding arterial hemoglobin concentrations taken at clinically relevant time points chosen at the discretion of the cardiac anesthesiologist. Thirty-five and 27 pooled pairs of data were obtained before and after cardiopulmonary bypass, respectively. Arterial hemoglobin concentration was analyzed using i-STAT CG8+test cartridges routinely used in the authors' operating rooms and those of other institutions. Linear regression and Bland-Altman analysis revealed a significant positive bias, wide limits of agreement, and low correlation coefficients between the noninvasive and arterial hemoglobin measurements. These findings were especially notable after compared with before cardiopulmonary bypass.
CONCLUSIONS: The results suggested that noninvasive measurement of hemoglobin overestimates arterial hemoglobin by almost 1 g/dL when compared to iSTAT. A lack of precision also was observed with noninvasive measurement of hemoglobin, especially after cardiopulmonary bypass. These findings supported the contention that sole reliance on noninvasive measurement of hemoglobin for transfusion decisions in cardiac surgery patients may be inappropriate. Published by Elsevier Inc.

Entities:  

Keywords:  Masimo Radical-7; cardiac surgery; cardiopulmonary bypass; i-STAT; instrumentation; noninvasive hemoglobin monitoring

Mesh:

Substances:

Year:  2016        PMID: 27475734      PMCID: PMC5028293          DOI: 10.1053/j.jvca.2016.03.148

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  45 in total

Review 1.  The current status of continuous noninvasive measurement of total, carboxy, and methemoglobin concentration.

Authors:  Micha Y Shamir; Aharon Avramovich; Todd Smaka
Journal:  Anesth Analg       Date:  2011-09-29       Impact factor: 5.108

2.  Error in noninvasive spectrophotometric measurement of blood hemoglobin concentration under conditions of blood loss.

Authors:  Rotem Naftalovich; Daniel Naftalovich
Journal:  Med Hypotheses       Date:  2011-08-05       Impact factor: 1.538

3.  Performance evaluation of a noninvasive hemoglobin monitoring device.

Authors:  Etienne Gayat; Antoine Bodin; Caroline Sportiello; Mireille Boisson; Jean-François Dreyfus; Emmanuel Mathieu; Marc Fischler
Journal:  Ann Emerg Med       Date:  2011-01-19       Impact factor: 5.721

4.  The use of the iSTAT portable analyzer in patients undergoing cardiopulmonary bypass.

Authors:  N R Connelly; M Magee; B Kiessling
Journal:  J Clin Monit       Date:  1996-07

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  The accuracy of noninvasive hemoglobin monitoring using the radical-7 pulse CO-Oximeter in children undergoing neurosurgery.

Authors:  Yong-Hee Park; Ji-Hyun Lee; Hyun-Gul Song; Hyo-Jin Byon; Hee-Soo Kim; Jin-Tae Kim
Journal:  Anesth Analg       Date:  2012-11-09       Impact factor: 5.108

7.  Evaluation of point-of-care haemoglobin measuring devices: a comparison of Radical-7™ pulse co-oximetry, HemoCue(®) and laboratory haemoglobin measurements in obstetric patients*.

Authors:  V A Skelton; N Wijayasinghe; S Sharafudeen; A Sange; N S Parry; C Junghans
Journal:  Anaesthesia       Date:  2012-10-22       Impact factor: 6.955

8.  Non-invasive measurement of hemoglobin: assessment of two different point-of-care technologies.

Authors:  Etienne Gayat; Jérôme Aulagnier; Emmanuel Matthieu; Mireille Boisson; Marc Fischler
Journal:  PLoS One       Date:  2012-01-06       Impact factor: 3.240

9.  Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer.

Authors:  N Shah; E A Osea; G J Martinez
Journal:  Int J Lab Hematol       Date:  2013-06-27       Impact factor: 2.877

10.  Validity of non-invasive point-of-care hemoglobin estimation in healthy and sick children-a method comparison study.

Authors:  Aditya Bhat; Amit Upadhyay; Vijay Jaiswal; Deepak Chawla; Dharamveer Singh; Mithilesh Kumar; C P Yadav
Journal:  Eur J Pediatr       Date:  2015-08-19       Impact factor: 3.183

View more
  4 in total

1.  Comparison of invasive and noninvasive blood hemoglobin measurement in the operating room: a systematic review and meta-analysis.

Authors:  Hosein Shabaninejad; Nashmil Ghadimi; Kourosh Sayehmiri; Hossein Hosseinifard; Rasoul Azarfarin; Hasan Abolghasem Gorji
Journal:  J Anesth       Date:  2019-03-20       Impact factor: 2.078

2.  Validity of accuracy and trending ability of non-invasive continuous total hemoglobin measurement in complex spine surgery: a prospective cohort study.

Authors:  Feng-Cheng Chang; Jr-Rung Lin; Fu-Chao Liu
Journal:  BMC Anesthesiol       Date:  2019-07-04       Impact factor: 2.217

3.  Impact of Increased End-Tidal Carbon Dioxide on Continuous Noninvasive Hemoglobin Monitoring during Laparoscopic Gastrectomy: A Randomized Controlled Study.

Authors:  Ha-Yeon Kim; Jong-Bum Choi; Sang-Uk Han; Hye-Sun Lee; Kyuhyeok Lee; Ji-Eun Kim
Journal:  J Pers Med       Date:  2022-01-26

4.  Noninvasive hematocrit assessment for cardiovascular magnetic resonance extracellular volume quantification using a point-of-care device and synthetic derivation.

Authors:  Sean Robison; Gauri Rani Karur; Rachel M Wald; Paaladinesh Thavendiranathan; Andrew M Crean; Kate Hanneman
Journal:  J Cardiovasc Magn Reson       Date:  2018-03-15       Impact factor: 5.364

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.