Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients.

Denis Frasca1, Claire Dahyot-Fizelier, Karen Catherine, Quentin Levrat, Bertrand Debaene, Olivier Mimoz.   

Abstract

OBJECTIVE: To determine whether noninvasive hemoglobin measurement by Pulse CO-Oximetry could provide clinically acceptable absolute and trend accuracy in critically ill patients, compared to other invasive methods of hemoglobin assessment available at bedside and the gold standard, the laboratory analyzer.
DESIGN: Prospective study.
SETTING: Surgical intensive care unit of a university teaching hospital. PATIENTS: Sixty-two patients continuously monitored with Pulse CO-Oximetry (Masimo Radical-7).
INTERVENTIONS: None. MEASUREMENTS AND
RESULTS: Four hundred seventy-one blood samples were analyzed by a point-of-care device (HemoCue 301), a satellite lab CO-Oximeter (Siemens RapidPoint 405), and a laboratory hematology analyzer (Sysmex XT-2000i), which was considered the reference device. Hemoglobin values reported from the invasive methods were compared to the values reported by the Pulse CO-Oximeter at the time of blood draw. When the case-to-case variation was assessed, the bias and limits of agreement were 0.0±1.0 g/dL for the Pulse CO-Oximeter, 0.3±1.3g/dL for the point-of-care device, and 0.9±0.6 g/dL for the satellite lab CO-Oximeter compared to the reference method. Pulse CO-Oximetry showed similar trend accuracy as satellite lab CO-Oximetry, whereas the point-of-care device did not appear to follow the trend of the laboratory analyzer as well as the other test devices.
CONCLUSION: When compared to laboratory reference values, hemoglobin measurement with Pulse CO-Oximetry has absolute accuracy and trending accuracy similar to widely used, invasive methods of hemoglobin measurement at bedside. Hemoglobin measurement with pulse CO-Oximetry has the additional advantages of providing continuous measurements, noninvasively, which may facilitate hemoglobin monitoring in the intensive care unit.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21666449     DOI: 10.1097/CCM.0b013e3182227e2d

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  35 in total

1.  Non-invasive haemoglobin estimation in patients with thalassaemia major.

Authors:  Murtadha K Al Khabori; Arwa Z Al-Riyami; Khalil Al-Farsi; Mohammed Al-Huneini; Abdulhakeem Al-Hashim; Nasser Al-Kemyani; Issa Al-Qarshoubi; Hammad Khan; Khalfan Al-Amrani; Shahina Daar
Journal:  Sultan Qaboos Univ Med J       Date:  2014-10-14

2.  Comparison of the accuracy of noninvasive hemoglobin monitoring for preoperative evaluation between adult and pediatric patients: a retrospective study.

Authors:  Yong-Hee Park; Seho Lim; Hyun Kang; Hwa-Yong Shin; Chong Wha Baek; Young Cheol Woo
Journal:  J Clin Monit Comput       Date:  2018-01-04       Impact factor: 2.502

3.  Effects of indigo carmine intravenous injection on noninvasive and continuous total hemoglobin measurement.

Authors:  Tsuyoshi Isosu; Tomohiko Satoh; Rieko Oishi; Tsuyoshi Imaizumi; Takahiro Hakozaki; Shinju Obara; Yukihiro Ikegami; Shin Kurosawa; Masahiro Murakawa
Journal:  J Clin Monit Comput       Date:  2015-06-17       Impact factor: 2.502

4.  Does a non-invasive hemoglobin monitor correlate with a venous blood sample in the acutely ill?

Authors:  Benjamin A von Schweinitz; Robert A De Lorenzo; Peter J Cuenca; Richard L Anschutz; Paul B Allen
Journal:  Intern Emerg Med       Date:  2014-10-17       Impact factor: 3.397

5.  Validation of continuous and noninvasive hemoglobin monitoring by pulse CO-oximetry in Japanese surgical patients.

Authors:  Tsuyoshi Isosu; Shinju Obara; Atsuyuki Hosono; Satoshi Ohashi; Yuko Nakano; Tsuyoshi Imaizumi; Midori Mogami; Masahiro Murakawa
Journal:  J Clin Monit Comput       Date:  2012-09-18       Impact factor: 2.502

Review 6.  Factors affecting hemoglobin measurement.

Authors:  Lauren Berkow
Journal:  J Clin Monit Comput       Date:  2013-03-26       Impact factor: 2.502

7.  The relative trending accuracy of noninvasive continuous hemoglobin monitoring during hemodialysis in critically ill patients.

Authors:  Hiroshi Yamada; Minako Saeki; Junko Ito; Kazuhiro Kawada; Aya Higurashi; Hiromi Funakoshi; Kohji Takeda
Journal:  J Clin Monit Comput       Date:  2014-05-03       Impact factor: 2.502

8.  Effects of indigo carmine intravenous injection on noninvasive and continuous total hemoglobin measurement with using the Revision L sensor.

Authors:  Tsuyoshi Isosu; Shinju Obara; Takahiro Hakozaki; Tsuyoshi Imaizumi; Yuzo Iseki; Midori Mogami; Satoshi Ohashi; Yukihiro Ikegami; Shin Kurosawa; Masahiro Murakawa
Journal:  J Clin Monit Comput       Date:  2016-02-22       Impact factor: 2.502

9.  Ability of the Masimo pulse CO-Oximeter to detect changes in hemoglobin.

Authors:  Douglas A Colquhoun; Katherine T Forkin; Marcel E Durieux; Robert H Thiele
Journal:  J Clin Monit Comput       Date:  2012-01-17       Impact factor: 2.502

10.  The use of a noninvasive hemoglobin monitor for determining fluid distribution and elimination in pediatric patients undergoing minor surgery.

Authors:  Qingquan Lian; Husong Li; Ruifeng Zeng; Junhui Lang; Wangning Shangguan; Huacheng Liu; Benfu Wang; Peter M Rodhe; Christer H Svensen
Journal:  J Clin Monit Comput       Date:  2014-02-19       Impact factor: 2.502

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.