Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: the effects of oximeter probe type and gender.

John R Feiner1, John W Severinghaus, Philip E Bickler.   

Abstract

INTRODUCTION: Pulse oximetry may overestimate arterial oxyhemoglobin saturation (Sao2) at low Sao2 levels in individuals with darkly pigmented skin, but other factors, such as gender and oximeter probe type, remain less studied.
METHODS: We studied the relationship between skin pigment and oximeter accuracy in 36 subjects (19 males, 17 females) of a range of skin tones. Clip-on type sensors and adhesive/disposable finger probes for the Masimo Radical, Nellcor N-595, and Nonin 9700 were studied. Semisupine subjects breathed air-nitrogen-CO2 mixtures via a mouthpiece to rapidly achieve 2- to 3-min stable plateaus of Sao2. Comparisons of Sao2 measured by pulse oximetry (Spo2) with Sao2 (by Radiometer OSM-3) were used in a multivariate model to assess the source of errors.
RESULTS: The mean bias (Spo2 - Sao2) for the 70%-80% saturation range was 2.61% for the Masimo Radical with clip-on sensor, -1.58% for the Radical with disposable sensor, 2.59% for the Nellcor clip, 3.6% for the Nellcor disposable, -0.60% for the Nonin clip, and 2.43% for the Nonin disposable. Dark skin increased bias at low Sao2; greater bias was seen with adhesive/disposable sensors than with the clip-on types. Up to 10% differences in saturation estimates were found among different instruments in dark-skinned subjects at low Sao2.
CONCLUSIONS: Multivariate analysis indicated that Sao2 level, sensor type, skin color, and gender were predictive of errors in Spo2 estimates at low Sao2 levels. The data suggest that clinically important bias should be considered when monitoring patients with saturations below 80%, especially those with darkly pigmented skin; but further study is needed to confirm these observations in the relevant populations.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18048893     DOI: 10.1213/01.ane.0000285988.35174.d9

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  69 in total

1.  Evaluation of various methods of point-of-care testing of haemoglobin concentration in blood donors.

Authors:  Abhay Singh; Anju Dubey; Atul Sonker; Rajendra Chaudhary
Journal:  Blood Transfus       Date:  2014-10-29       Impact factor: 3.443

Review 2.  Teaching Pediatric Life Support in Limited-Resource Settings: Contextualized Management Guidelines.

Authors:  Mark E Ralston; Allan de Caen
Journal:  J Pediatr Intensive Care       Date:  2016-06-29

Review 3.  In vivo validation of cerebral near-infrared spectroscopy: a review.

Authors:  Amalie la Cour; Gorm Greisen; Simon Hyttel-Sorensen
Journal:  Neurophotonics       Date:  2018-11-27       Impact factor: 3.593

Review 4.  Sex and gender differences and biases in artificial intelligence for biomedicine and healthcare.

Authors:  Davide Cirillo; Silvina Catuara-Solarz; Czuee Morey; Emre Guney; Laia Subirats; Simona Mellino; Annalisa Gigante; Alfonso Valencia; María José Rementeria; Antonella Santuccione Chadha; Nikolaos Mavridis
Journal:  NPJ Digit Med       Date:  2020-06-01

5.  Validation of continuous and noninvasive hemoglobin monitoring by pulse CO-oximetry in Japanese surgical patients.

Authors:  Tsuyoshi Isosu; Shinju Obara; Atsuyuki Hosono; Satoshi Ohashi; Yuko Nakano; Tsuyoshi Imaizumi; Midori Mogami; Masahiro Murakawa
Journal:  J Clin Monit Comput       Date:  2012-09-18       Impact factor: 2.502

Review 6.  Is pulse oximetry an essential tool or just another distraction? The role of the pulse oximeter in modern anesthesia care.

Authors:  Amit Shah; Kirk H Shelley
Journal:  J Clin Monit Comput       Date:  2013-01-12       Impact factor: 2.502

7.  The accuracy of a near-infrared spectroscopy cerebral oximetry device and its potential value for estimating jugular venous oxygen saturation.

Authors:  Keita Ikeda; David B MacLeod; Hilary P Grocott; Eugene W Moretti; Warwick Ames; Charles Vacchiano
Journal:  Anesth Analg       Date:  2014-12       Impact factor: 5.108

8.  Wrist-Sensor Pulse Oximeter Enables Prolonged Patient Monitoring in Chronic Lung Diseases.

Authors:  Alexander Guber; Gali Epstein Shochet; Sarah Kohn; David Shitrit
Journal:  J Med Syst       Date:  2019-06-14       Impact factor: 4.460

9.  Carbon dioxide monitoring during long-term noninvasive respiratory support in children.

Authors:  Rebeca Paiva; Uros Krivec; Guillaume Aubertin; Emmanuelle Cohen; Annick Clément; Brigitte Fauroux
Journal:  Intensive Care Med       Date:  2009-01-27       Impact factor: 17.440

10.  Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia.

Authors:  John R Feiner; Mark D Rollins; Jeffrey W Sall; Helge Eilers; Paul Au; Philip E Bickler
Journal:  Anesth Analg       Date:  2013-03-11       Impact factor: 5.108

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.