Literature DB >> 21048100

The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-Oximetry in human subjects undergoing hemodilution.

Mark R Macknet1, Martin Allard, Richard L Applegate, James Rook.   

Abstract

BACKGROUND: Total hemoglobin (tHb) is one the most frequently ordered laboratory measurements. Pulse CO-Oximetry™ (Masimo Corp., Irvine, CA) is a multi-wavelength spectrophotometric method for noninvasive and continuous hemoglobin monitoring (SpHb). In this study, we evaluated the accuracy of SpHb compared with laboratory CO-Oximeter measurement of tHb from arterial blood samples in 20 healthy volunteer subjects undergoing hemodilution.
METHODS: After enrollment, approximately 500 mL of blood was drawn from subjects through an arterial or venous catheter. Each subject then rapidly received crystalloid i.v. fluid to compensate for the decrease in intravascular volume and reduce the hemoglobin concentration. Subjects received a maximum of 30 mL/kg i.v. fluid. SpHb was continuously monitored and recorded, and serial arterial blood samples were taken during the procedure. SpHb accuracy was analyzed by pairing SpHb and tHb measurements after the arterial blood draw with the resulting tHb test result. Bias, precision, and the average root-mean-square error were calculated.
RESULTS: One hundred sixty-five tHb measurements were collected. The average decrease in tHb during the blood removal and hemodilution procedure was 2.4 ± 0.8 g/dL (mean ± SD). The average difference between 335 paired measurements of SpHb and tHb was -0.15 g/dL, 1 SD of the difference was 0.92 g/dL, and the average root-mean-square difference was 0.94 g/dL. The difference between SpHb and tHb was <2.0 g/dL for 97% of the measurements. The difference was <1.5 g/dL for 97% of the measurements when tHb was <10 g/dL.
CONCLUSIONS: Pulse CO-Oximetry-based SpHb measurement is accurate within 1.0 g/dL (1 SD) compared with laboratory CO-Oximeter tHb measurement in subjects undergoing hemodilution.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21048100     DOI: 10.1213/ANE.0b013e3181fc74b9

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  34 in total

Review 1.  Anemia in critical illness: insights into etiology, consequences, and management.

Authors:  Shailaja J Hayden; Tyler J Albert; Timothy R Watkins; Erik R Swenson
Journal:  Am J Respir Crit Care Med       Date:  2012-01-26       Impact factor: 21.405

2.  Evaluation of multiwave pulse total-hemoglobinometer during general anesthesia.

Authors:  Daisuke Toyoda; Rie Yasumura; Mitsue Fukuda; Ryoichi Ochiai; Yoshifumi Kotake
Journal:  J Anesth       Date:  2013-10-22       Impact factor: 2.078

3.  Comparison of the accuracy of noninvasive hemoglobin monitoring for preoperative evaluation between adult and pediatric patients: a retrospective study.

Authors:  Yong-Hee Park; Seho Lim; Hyun Kang; Hwa-Yong Shin; Chong Wha Baek; Young Cheol Woo
Journal:  J Clin Monit Comput       Date:  2018-01-04       Impact factor: 2.502

4.  Effects of indigo carmine intravenous injection on noninvasive and continuous total hemoglobin measurement.

Authors:  Tsuyoshi Isosu; Tomohiko Satoh; Rieko Oishi; Tsuyoshi Imaizumi; Takahiro Hakozaki; Shinju Obara; Yukihiro Ikegami; Shin Kurosawa; Masahiro Murakawa
Journal:  J Clin Monit Comput       Date:  2015-06-17       Impact factor: 2.502

5.  Does a non-invasive hemoglobin monitor correlate with a venous blood sample in the acutely ill?

Authors:  Benjamin A von Schweinitz; Robert A De Lorenzo; Peter J Cuenca; Richard L Anschutz; Paul B Allen
Journal:  Intern Emerg Med       Date:  2014-10-17       Impact factor: 3.397

6.  Validation of continuous and noninvasive hemoglobin monitoring by pulse CO-oximetry in Japanese surgical patients.

Authors:  Tsuyoshi Isosu; Shinju Obara; Atsuyuki Hosono; Satoshi Ohashi; Yuko Nakano; Tsuyoshi Imaizumi; Midori Mogami; Masahiro Murakawa
Journal:  J Clin Monit Comput       Date:  2012-09-18       Impact factor: 2.502

Review 7.  Factors affecting hemoglobin measurement.

Authors:  Lauren Berkow
Journal:  J Clin Monit Comput       Date:  2013-03-26       Impact factor: 2.502

8.  The relative trending accuracy of noninvasive continuous hemoglobin monitoring during hemodialysis in critically ill patients.

Authors:  Hiroshi Yamada; Minako Saeki; Junko Ito; Kazuhiro Kawada; Aya Higurashi; Hiromi Funakoshi; Kohji Takeda
Journal:  J Clin Monit Comput       Date:  2014-05-03       Impact factor: 2.502

9.  Effects of indigo carmine intravenous injection on noninvasive and continuous total hemoglobin measurement with using the Revision L sensor.

Authors:  Tsuyoshi Isosu; Shinju Obara; Takahiro Hakozaki; Tsuyoshi Imaizumi; Yuzo Iseki; Midori Mogami; Satoshi Ohashi; Yukihiro Ikegami; Shin Kurosawa; Masahiro Murakawa
Journal:  J Clin Monit Comput       Date:  2016-02-22       Impact factor: 2.502

10.  The use of a noninvasive hemoglobin monitor for determining fluid distribution and elimination in pediatric patients undergoing minor surgery.

Authors:  Qingquan Lian; Husong Li; Ruifeng Zeng; Junhui Lang; Wangning Shangguan; Huacheng Liu; Benfu Wang; Peter M Rodhe; Christer H Svensen
Journal:  J Clin Monit Comput       Date:  2014-02-19       Impact factor: 2.502