Accuracy of non-invasive continuous total hemoglobin measurement by Pulse CO-Oximetry in severe traumatized and surgical bleeding patients.

Werner Baulig1, Burkhardt Seifert2, Donat R Spahn3, Oliver M Theusinger3.   

Abstract

The Masimo Radical-7 Pulse CO-Oximeter (Masimo Corp., USA) non-invasively computes hemoglobin concentration (SpHb). SpHb was compared to Co-Oximeter readings (CoOxHb) of arterial samples in surgery patients of the emergency department. Forty-six patients were enrolled. The Masimo R1 25L (revision F and G) adult adhesive sensor was attached to the ring finger of the arterially cannulated hand. Before start, every 30 min during surgery and in the case of severe bleeding SpHb and CoOxHb values were documented. SpHb and post hoc adjusted SpHb (AdSpHb) values were analyzed. Linear regression analysis and Bland-Altman plot for agreement were performed. The detection failure rate of SpHb was 24.5 %. CoOxHb and SpHb showed a strong correlation (r = +0.81), but agreement was moderate [bias (LOA) of -0.6 (-3.0; +1.9)] g/dl. Positive and negative predicted value was 0.49 and 0.69. Exclusion of changes of CoOxHb values ≤1 g/dl resulted in a positive and negative predictive value of 0.66 and 1.00. Post hoc adjustment of the SpHb (AdSpHb) improved linear correlation of CoOxHb and AdSpHb [r = +0.90 (p < 0.001)] but less the agreement [bias (LOA) of CoOxHb and AdSpHb = -0.1 (-2.1/+1.9) g/dl]. SpHb agreed only moderately with CoOxHb values and predicted decreases of CoOxHb only if changes of SpHb ≤ 1.0 g/dl were excluded. The detection failure rate of SpHb was high. At present, additional refinements of the current technology are necessary to further improve performance of non-invasive hemoglobin measurement in the clinical setting.

Entities:  

Keywords:  Bleeding; Continuous; Measurement; Non-invasive; Patients; SpHb; Traumatized

Mesh:

Substances:

Year:  2015        PMID: 26686689     DOI: 10.1007/s10877-015-9816-2

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  32 in total

1.  Comparison of blood counts in venous, fingertip and arterial blood and their measurement variation.

Authors:  Z W Yang; S H Yang; L Chen; J Qu; J Zhu; Z Tang
Journal:  Clin Lab Haematol       Date:  2001-06

Review 2.  The current status of continuous noninvasive measurement of total, carboxy, and methemoglobin concentration.

Authors:  Micha Y Shamir; Aharon Avramovich; Todd Smaka
Journal:  Anesth Analg       Date:  2011-09-29       Impact factor: 5.108

3.  Hemoximetry as the "gold standard"? Error assessment based on differences among identical blood gas analyzer devices of five manufacturers.

Authors:  Hartmut Gehring; Lutz Duembgen; Mareike Peterlein; Söhnke Hagelberg; Leif Dibbelt
Journal:  Anesth Analg       Date:  2007-12       Impact factor: 5.108

4.  Error in noninvasive spectrophotometric measurement of blood hemoglobin concentration under conditions of blood loss.

Authors:  Rotem Naftalovich; Daniel Naftalovich
Journal:  Med Hypotheses       Date:  2011-08-05       Impact factor: 1.538

5.  Ability of the Masimo pulse CO-Oximeter to detect changes in hemoglobin.

Authors:  Douglas A Colquhoun; Katherine T Forkin; Marcel E Durieux; Robert H Thiele
Journal:  J Clin Monit Comput       Date:  2012-01-17       Impact factor: 2.502

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Continuous noninvasive hemoglobin monitoring during complex spine surgery.

Authors:  Lauren Berkow; Stephanie Rotolo; Erin Mirski
Journal:  Anesth Analg       Date:  2011-09-29       Impact factor: 5.108

8.  Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement.

Authors:  Lionel Lamhaut; Roxana Apriotesei; Xavier Combes; Marc Lejay; Pierre Carli; Benoît Vivien
Journal:  Anesthesiology       Date:  2011-09       Impact factor: 7.892

9.  Evaluation of point-of-care haemoglobin measuring devices: a comparison of Radical-7™ pulse co-oximetry, HemoCue(®) and laboratory haemoglobin measurements in obstetric patients*.

Authors:  V A Skelton; N Wijayasinghe; S Sharafudeen; A Sange; N S Parry; C Junghans
Journal:  Anaesthesia       Date:  2012-10-22       Impact factor: 6.955

10.  Accuracy of point-of-care-testing (POCT) for determining hemoglobin concentrations.

Authors:  H Gehring; C Hornberger; L Dibbelt; A Rothsigkeit; K Gerlach; J Schumacher; P Schmucker
Journal:  Acta Anaesthesiol Scand       Date:  2002-09       Impact factor: 2.105

View more
  6 in total

1.  Comparison of invasive and noninvasive blood hemoglobin measurement in the operating room: a systematic review and meta-analysis.

Authors:  Hosein Shabaninejad; Nashmil Ghadimi; Kourosh Sayehmiri; Hossein Hosseinifard; Rasoul Azarfarin; Hasan Abolghasem Gorji
Journal:  J Anesth       Date:  2019-03-20       Impact factor: 2.078

Review 2.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

3.  Validity of accuracy and trending ability of non-invasive continuous total hemoglobin measurement in complex spine surgery: a prospective cohort study.

Authors:  Feng-Cheng Chang; Jr-Rung Lin; Fu-Chao Liu
Journal:  BMC Anesthesiol       Date:  2019-07-04       Impact factor: 2.217

4.  Intraoperative continuous noninvasive hemoglobin monitoring in patients with placenta previa undergoing cesarean section: a prospective observational study.

Authors:  Hyunsung Kim; Sang-Hwan Do; Jung-Won Hwang; Hyo-Seok Na
Journal:  Anesth Pain Med (Seoul)       Date:  2019-10-31

5.  Accuracy of Continuous and Noninvasive Hemoglobin Monitoring in the Presence of CO2 Insufflation: An Observational Pilot Study.

Authors:  Hye Sun Lee; Ji Won Yoo; Ha Yeon Kim; Na Young Kim; Ji Eun Kim
Journal:  Med Sci Monit       Date:  2021-10-27

6.  Accuracy of non-invasive haemoglobin measurements in patients undergoing transurethral resection of prostate surgery.

Authors:  Naveen Selvaraj; Gaurav Jain; Debendra Kumar Tripathy; Ankur Mittal; Haritha Indulekha
Journal:  Indian J Anaesth       Date:  2021-05-10
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.