| Literature DB >> 22821447 |
Wiebke Semper-Hogg1, Silvan Kraft, Sebastian Stiller, Juergen Mehrhof, Katja Nelson.
Abstract
OBJECTIVES: Position stability of the abutment should be investigated in four implant systems with a conical implant-abutment connection.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22821447 PMCID: PMC3607726 DOI: 10.1007/s00784-012-0786-1
Source DB: PubMed Journal: Clin Oral Investig ISSN: 1432-6981 Impact factor: 3.573
Specifications of the implants and abutments used
| System | Ident. | Article no. | Lot no. | Ident. | Mat | Article no. | Lot no. | Manufacturer |
|---|---|---|---|---|---|---|---|---|
| Nobel Active (S1) | Internal RP, 4.3 × 11.5 mm | 34132 | 707722 | Esthetic abutment | Ti | 34199 | 708540 | Nobel Biocare AB, SE-40226 Göteborg, Sweden |
| 696896 | 708540 | |||||||
| 696896 | 708540 | |||||||
| 707023 | 708540 | |||||||
| 707023 | 708540 | |||||||
| 707023 | 708540 | |||||||
| 707023 | 708540 | |||||||
| Bone Level (S2) | Bone Level, 4.1 RC SLActive, 12 mm | 021.4112 | K6756 | Universal abutment | Ti | 022.2104 | X3223 | Institut Straumann AG, CH-4002 Basel, Switzerland |
| F7007 | R8701 | |||||||
| F7007 | R8701 | |||||||
| F7007 | R8701 | |||||||
| F7007 | R8701 | |||||||
| F7007 | R8701 | |||||||
| F7007 | R8701 | |||||||
| Ankylos (S3) | C/X Implant, B14∅4.5/L14 | 170556/31010440 | 0020039200 | Regular/X abutment GH3.0/A0 | Ti | 31024130 | 20035835 | FRIADENT GmbH, 68229 Mannheim, Germany |
| 0020034919 | 20035835 | |||||||
| 0020034919 | 20035835 | |||||||
| 0020039201 | 20035835 | |||||||
| 0020039201 | 20035835 | |||||||
| 0020038909 | 20035835 | |||||||
| 0020038909 | 20035835 | |||||||
| Conelog (S4) | Screw-Line ConeLog Implant Promote plus ∅4.3 L11 | E-C1062.4311 | 0000029157 | Universal abutment | Ti | C2211.4300 | 0000028679 | Camlog Biotechnologies AG, CH-4053 Basel, Switzerland |
| 0010024599 | 0010027056 | |||||||
| 0010024599 | 0010027056 | |||||||
| 0010024599 | 0010027056 | |||||||
| 0010024599 | 0010027056 | |||||||
| 0010024599 | 0010027056 | |||||||
| 0010024599 | 0010027056 |
Fig. 1Abutments showing the positional index designs. a S1: Nobel Active. Conical implant–abutment connection with a cone angle of 12° and an internal hexagonal positional index. b S2: Bone Level. Conical implant–abutment connection with a cone angle of 15° and cams and grooves. c S3: Ankylos C/X. Conical implant–abutment connection with a cone angle of 5.7° and six cams and grooves. d S4: Conelog. Conical implant–abutment connection with a cone angle of 7.5°and three cams and grooves
Values inserted into the previously developed formulas to calculate the rotational freedom of the positional index for the implant systems used
| Polygon |
|
|
|
|
|---|---|---|---|---|
| S1 | 1.52 | 6 | 0.02 | 3.1 |
| Cam–groove |
|
|
|
|
| S2 | 1.35 | 2.0 | 0.02 | 1.7 |
| S3 | 1.00 | 29.5 | 0.02 | 2.6 |
| S4 | 1.15 | 18.1 | 0.02 | 2.1 |
Minimal (min) and maximal (max) experimental rotational and vertical displacement measurement values for each implant system and test person
| System/implant | Person 1 | Person 2 | ||
|---|---|---|---|---|
| Rotation | ||||
| Min (°) | Max (°) | Min (°) | Max (°) | |
| S1 | 0.07 | 3.95 | <0.01 | 3.23 |
| 1 | 0.14 | 3.41 | 0.02 | 2.59 |
| 2 | 0.07 | 3.37 | <0.01 | 1.31 |
| 3 | 0.86 | 3.88 | 0.01 | 1.03 |
| 4 | 0.13 | 2.29 | 0.01 | 3.23 |
| 5 | 1.15 | 3.95 | 0.04 | 0.69 |
| 6 | 0.60 | 3.28 | 0.04 | 0.63 |
| S2 | 0.38 | 2.58 | 0.04 | 2.50 |
| 1 | 0.55 | 1.30 | 0.22 | 1.40 |
| 2 | 0.38 | 1.22 | 0.25 | 1.24 |
| 3 | 0.44 | 2.58 | 0.25 | 2.50 |
| 4 | 1.00 | 1.89 | 0.14 | 1.89 |
| 5 | 1.25 | 1.48 | 0.41 | 1.62 |
| 6 | 0.61 | 2.00 | 0.04 | 1.39 |
| S3 | 0.13 | 3.84 | 0.02 | 4.20 |
| 1 | 0.87 | 3.56 | 0.02 | 1.05 |
| 2 | 1.24 | 3.42 | 0.04 | 4.10 |
| 3 | 0.51 | 3.84 | 0.29 | 4.20 |
| 4 | 1.53 | 3.23 | 0.05 | 0.76 |
| 5 | 0.14 | 1.73 | 0.14 | 1.55 |
| 6 | 0.13 | 3.15 | 0.20 | 1.49 |
| S4 | <0.01 | 1.30 | <0.01 | 1.72 |
| 1 | 0.07 | 0.88 | <0.01 | 0.88 |
| 2 | <0.01 | 1.28 | 0.05 | 0.93 |
| 3 | <0.01 | 1.21 | <0.01 | 0.77 |
| 4 | 0.01 | 1.17 | <0.01 | 0.39 |
| 5 | 0.02 | 0.89 | <0.01 | 0.28 |
| 6 | 0.04 | 1.30 | <0.01 | 1.72 |
| Vertical displacement | ||||
| Min (μm) | Max (μm) | Min (μm) | Max (μm) | |
| S1 | <1 | 21 | <1 | 14 |
| 1 | <1 | 13 | <1 | 12 |
| 2 | <1 | 12 | <1 | 14 |
| 3 | <1 | 19 | <1 | 8 |
| 4 | <1 | 21 | <1 | 10 |
| 5 | <1 | 11 | <1 | 7 |
| 6 | <1 | 7 | <1 | 9 |
| S2 | <1 | 8 | <1 | 8 |
| 1 | <1 | 6 | <1 | 5 |
| 2 | <1 | 5 | <1 | 5 |
| 3 | <1 | 8 | <1 | 7 |
| 4 | <1 | 8 | <1 | 8 |
| 5 | <1 | 6 | <1 | 6 |
| 6 | <1 | 6 | <1 | 6 |
| S3 | <1 | 11 | <1 | 7 |
| 1 | <1 | 7 | <1 | 7 |
| 2 | <1 | 9 | <1 | 7 |
| 3 | <1 | 5 | <1 | 4 |
| 4 | <1 | 4 | <1 | 2 |
| 5 | <1 | 11 | <1 | 5 |
| 6 | <1 | 6 | <1 | 3 |
| S4 | <1 | 25 | <1 | 17 |
| 1 | <1 | 10 | <1 | 9 |
| 2 | <1 | 9 | <1 | 7 |
| 3 | <1 | 6 | <1 | 14 |
| 4 | <1 | 25 | <1 | 17 |
| 5 | <1 | 13 | <1 | 10 |
| 6 | <1 | 10 | <1 | 11 |