Effects of repeated manual disassembly and reassembly on the positional stability of various implant-abutment complexes: an experimental study.

Wiebke Semper1, Susanne Heberer, Juergen Mehrhof, Tania Schink, Katja Nelson.   

Abstract

PURPOSE: The purpose of this study was to evaluate rotational, vertical, and canting changes in the position of the rotation-safe component in the implant-abutment assemblies of five different implant systems (ITI, Steri-Oss, Camlog, Astra Tech, and Replace Select) after manual removal and reassembly.
MATERIALS AND METHODS: Prefabricated stainless steel models were used for each implant system, into which six implants were fixated with polymethylmethacrylate resin. Rotation-safe abutments (components) were screwed into the implants according to the manufacturers' specifications. Three test persons with varying knowledge of the theory and practice of implant dentistry manually assembled and reassembled the implant-abutment joint using each system-specific screwdriver 20 times each. A coordinate reading machine was used to detect discrepancies in position after each reassembly in relation to a coordinate system. Rotational freedom, changes in vertical height, and deviations in angulation were assessed. Statistical analysis was performed based on the nonparametric analysis of variance of repeated measurements.
RESULTS: The tested complexes showed rotational freedom that ranged from 0.92 to 4.92 degrees, with significant differences between the systems. Camlog was significantly different from all other systems tested regarding rotational freedom, whereas Steri-Oss, Astra Tech, and Replace Select showed no significant difference between each other because of their nondiscrepant mean degree of rotational freedom. Vertical alterations in position ranged from 1 to 83 microm. A statistically significant difference was detected between butt-joint and beveled implant-abutment connections, with ITI and Astra Tech showing no significant difference when compared to each other, but displaying a significant difference versus all other systems tested. Canting discrepancies were not significant, with no influence of implant system or test person clearly detectable.
CONCLUSION: Three-dimensional changes in the location of the abutment in relation to the implant result after manual assembly and reassembly of the implant-abutment complex.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20209190

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  8 in total

1.  Evaluation of impression accuracy for a four-implant mandibular model--a digital approach.

Authors:  Michael Stimmelmayr; Kurt Erdelt; Jan-Frederik Güth; Arndt Happe; Florian Beuer
Journal:  Clin Oral Investig       Date:  2011-10-19       Impact factor: 3.573

2.  In vitro analysis of the microbiological sealing of tapered implants after mechanical cycling.

Authors:  Deceles Cristina Costa Alves; Paulo Sérgio Perri de Carvalho; Carlos Nelson Elias; Eduardo Vedovatto; Elizabeth Ferreira Martinez
Journal:  Clin Oral Investig       Date:  2016-02-20       Impact factor: 3.573

3.  Clinical study evaluating the discrepancy of two different impression techniques of four implants in an edentulous jaw.

Authors:  Michael Stimmelmayr; Jan-Frederik Güth; Kurt Erdelt; Arndt Happe; Markus Schlee; Florian Beuer
Journal:  Clin Oral Investig       Date:  2012-12-06       Impact factor: 3.573

4.  Analytical and experimental position stability of the abutment in different dental implant systems with a conical implant-abutment connection.

Authors:  Wiebke Semper-Hogg; Silvan Kraft; Sebastian Stiller; Juergen Mehrhof; Katja Nelson
Journal:  Clin Oral Investig       Date:  2012-07-22       Impact factor: 3.573

5.  A three-dimensional finite element analysis of a passive and friction fit implant abutment interface and the influence of occlusal table dimension on the stress distribution pattern on the implant and surrounding bone.

Authors:  Hasan Sarfaraz; Anoopa Paulose; K Kamalakanth Shenoy; Akhter Hussain
Journal:  J Indian Prosthodont Soc       Date:  2015 Jul-Sep

6.  An evaluation of superhydrophilic surfaces of dental implants - a systematic review and meta-analysis.

Authors:  Arkadiusz Makowiecki; Jakub Hadzik; Artur Błaszczyszyn; Tomasz Gedrange; Marzena Dominiak
Journal:  BMC Oral Health       Date:  2019-05-10       Impact factor: 2.757

7.  Rotational tolerances of a titanium abutment in the as-received condition and after screw tightening in a conical implant connection.

Authors:  Rosario Prisco; Giuseppe Troiano; Luigi Laino; Khrystyna Zhurakivska
Journal:  J Adv Prosthodont       Date:  2021-12-22       Impact factor: 1.904

8.  Effect of tightening torque on the marginal adaptation of cement-retained implant-supported fixed dental prostheses.

Authors:  Jalil Ghanbarzadeh; Hossin Dashti; Reza Karamad; Marzieh Alikhasi; Mohammadreza Nakhaei
Journal:  Dent Res J (Isfahan)       Date:  2015 Jul-Aug
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.