Axial displacement of abutments into implants and implant replicas, with the tapered cone-screw internal connection, as a function of tightening torque.

Bruno Dailey1, Laurence Jordan, Olivier Blind, Bruno Tavernier.   

Abstract

PURPOSE: The passive fit of a superstructure on implant abutments is essential to success. One source of error when using a tapered cone-screw internal connection may be the difference between the tightening torque level applied to the abutments by the laboratory technician compared to that applied by the treating clinician. The purpose of this study was to measure the axial displacement of tapered cone-screw abutments into implants and their replicas as a function of the tightening torque level.
MATERIALS AND METHODS: Twenty tapered cone-screw abutments were selected. Two groups were created: 10 abutments were secured into 10 implants, and 10 abutments were secured into 10 corresponding implant replicas. Each abutment was tightened in increasing increments of 5 Ncm, from 0 Ncm to 45 Ncm, with a torque controller. The length of each sample was measured repeatedly with an Electronic Digital Micrometer. The mean axial displacement for the implant group and the replica group was calculated. The data were analyzed by the Mann-Whitney and Spearman tests.
RESULTS: For both groups, there was always an axial displacement of the abutment upon each incremental application of torque. The mean axial displacement values varied between 7 and 12 microm for the implant group and between 6 and 21 microm for the replica group at each 5-Ncm increment. From 0 to 45 Ncm, the total mean axial displacement values were 89 microm for the implant group and 122 microm for the replica group.
CONCLUSION: There was a continuous axial displacement of the abutments into implants and implant replicas when the applied torque was raised from 0 to 45 Ncm. Torque applied above the level recommended by the manufacturer increased the difference in displacement between the two groups.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19492640

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  7 in total

1.  Sinking and fit of abutment of locking taper implant system.

Authors:  Seung-Jin Moon; Hee-Jung Kim; Mee-Kyoung Son; Chae-Heon Chung
Journal:  J Adv Prosthodont       Date:  2009-07-31       Impact factor: 1.904

2.  Analytical and experimental position stability of the abutment in different dental implant systems with a conical implant-abutment connection.

Authors:  Wiebke Semper-Hogg; Silvan Kraft; Sebastian Stiller; Juergen Mehrhof; Katja Nelson
Journal:  Clin Oral Investig       Date:  2012-07-22       Impact factor: 3.573

3.  Complication incidence of two implant systems up to six years: a comparison between internal and external connection implants.

Authors:  Sung-Wook Chae; Young-Sung Kim; Yong-Moo Lee; Won-Kyung Kim; Young-Kyoo Lee; Su-Hwan Kim
Journal:  J Periodontal Implant Sci       Date:  2015-02-25       Impact factor: 2.614

4.  The effect of repeated torque tightening on total lengths of implant abutments in different internal implant‒abutment connections.

Authors:  Fariba Saleh Saber; Nader Abolfazli; Soheil Jannatii Ataei; Mahsa Taghizade Motlagh; Vahede Gharekhani
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2017-06-21

5.  Rotational tolerances of a titanium abutment in the as-received condition and after screw tightening in a conical implant connection.

Authors:  Rosario Prisco; Giuseppe Troiano; Luigi Laino; Khrystyna Zhurakivska
Journal:  J Adv Prosthodont       Date:  2021-12-22       Impact factor: 1.904

6.  Effect of tightening torque on the marginal adaptation of cement-retained implant-supported fixed dental prostheses.

Authors:  Jalil Ghanbarzadeh; Hossin Dashti; Reza Karamad; Marzieh Alikhasi; Mohammadreza Nakhaei
Journal:  Dent Res J (Isfahan)       Date:  2015 Jul-Aug

7.  Digital evaluation of axial displacement by implant-abutment connection type: An in vitro study.

Authors:  Sung-Jun Kim; KeunBaDa Son; Kyu-Bok Lee
Journal:  J Adv Prosthodont       Date:  2018-10-22       Impact factor: 1.904

  7 in total