The influence of abutment screw tightening on screw joint configuration.

Lisa A Lang1, Rui-Feng Wang, Kenneth B May.   

Abstract

STATEMENT OF PROBLEM: Limiting abutment-to-implant hexagonal discrepancies and rotational movement of the abutment around the implant to less than 5 degrees would result in a more stable screw joint. However, the exact relationship after abutment screw tightening is unknown, as is the effect of a counter-torque device in limiting abutment movement during screw tightening.
PURPOSE: This study examined the orientation of the abutment hexagon to the implant hexagon after tightening of the abutment screw for several abutment systems with and without the use of a counter-torque device.
MATERIAL AND METHODS: Thirty conical self-tapping implants (3.75 x 10.0 mm) and 10 wide-platform Brånemark System implants (5.0 x 10.0 mm), along with 10 abutment specimens from the CeraOne, Estheticone, Procera, and AuraAdapt systems, were selected for this investigation. The implants were placed in a holding device prior to tightening of the abutments. When the tightening torque recommended for each abutment system was reached with the use of a torque controller, each implant abutment specimen was removed from the holding device and embedded in a hard resin medium. The specimens were sectioned in a horizontal direction at the level of the hexagons and cleansed of debris prior to examination. The hexagon orientations were assessed as the degree and direction of rotation of the abutment hexagon around the implant hexagon.
RESULTS: The range of the maximum degrees of rotation for all 4 abutment groups tightened with or without the counter-torque device was slightly more than 3.53 degrees. The absolute degrees of rotation for all 4 abutment groups were less than 1.50 degrees with or without the use of the counter-torque device.
CONCLUSION: The hexagon-to-hexagon orientation measured as rotational fit on all abutment systems was below the 5 degrees suggested as optimal for screw joint stability. The absolute degrees of rotation for all 4 abutment groups were less than 1.50 degrees regardless of whether the counter-torque device was used.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 11807487     DOI: 10.1067/mpr.2002.121488

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  6 in total

1.  Teeth movement in denture and implant-supported prosthesis influenced by microwave flask systems.

Authors:  Rafael L X Consani; Leonardo G Sonehara; Marcelo F Mesquita; Valentim A R Barão; Conrado R Caetano
Journal:  J Oral Biol Craniofac Res       Date:  2015-07-15

2.  In vitro assessment of the accuracy of digital impressions prepared using a single system for full-arch restorations on implants.

Authors:  Leonardo Ciocca; Roberto Meneghello; Carlo Monaco; Gianpaolo Savio; Lorenzo Scheda; Maria Rosaria Gatto; Paolo Baldissara
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-03-02       Impact factor: 2.924

3.  Analytical and experimental position stability of the abutment in different dental implant systems with a conical implant-abutment connection.

Authors:  Wiebke Semper-Hogg; Silvan Kraft; Sebastian Stiller; Juergen Mehrhof; Katja Nelson
Journal:  Clin Oral Investig       Date:  2012-07-22       Impact factor: 3.573

4.  A rationale method for evaluating unscrewing torque values of prosthetic screws in dental implants.

Authors:  Felipe Miguel Saliba; Mayra Cardoso; Marcelo Ferreira Torres; Alexandre Carvalho Teixeira; Eduardo José Veras Lourenço; Daniel de Moraes Telles
Journal:  J Appl Oral Sci       Date:  2011 Jan-Feb       Impact factor: 2.698

5.  Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections.

Authors:  Hakimeh Siadat; Elaheh Beyabanaki; Niloufar Mousavi; Marzieh Alikhasi
Journal:  J Adv Prosthodont       Date:  2017-08-16       Impact factor: 1.904

6.  Rotational tolerances of a titanium abutment in the as-received condition and after screw tightening in a conical implant connection.

Authors:  Rosario Prisco; Giuseppe Troiano; Luigi Laino; Khrystyna Zhurakivska
Journal:  J Adv Prosthodont       Date:  2021-12-22       Impact factor: 1.904

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.